A metallurgical examination of fractured stainless-steel ASIF tibial plates. 1976

M H Richman, and J K Weltman, and A Cole

Between 1970 and 1973 99 tibial fractures were treated by rigid internal fixation with ASIF plates. The fractures were all regarded as sufficiently stable for exercise without weight bearing, thus needing no additional external support during the healing period. Four of the plates broke late in the healing period, after the onset of weight bearing. These fractures had some degree of delayed union with slight resorption of the bone ends, resulting in cyclical bending of the plate. Examination of 2 of the fractured plates by scanning electron microscopy, electron microprobe analysis and optical metallography revealed that the primary cause of plate fracture was fatigue. There was no evidence that corrosion fatigue or inclusion content were factors leading to plate fracture.

UI MeSH Term Description Entries
D008297 Male Males
D008855 Microscopy, Electron, Scanning Microscopy in which the object is examined directly by an electron beam scanning the specimen point-by-point. The image is constructed by detecting the products of specimen interactions that are projected above the plane of the sample, such as backscattered electrons. Although SCANNING TRANSMISSION ELECTRON MICROSCOPY also scans the specimen point by point with the electron beam, the image is constructed by detecting the electrons, or their interaction products that are transmitted through the sample plane, so that is a form of TRANSMISSION ELECTRON MICROSCOPY. Scanning Electron Microscopy,Electron Scanning Microscopy,Electron Microscopies, Scanning,Electron Microscopy, Scanning,Electron Scanning Microscopies,Microscopies, Electron Scanning,Microscopies, Scanning Electron,Microscopy, Electron Scanning,Microscopy, Scanning Electron,Scanning Electron Microscopies,Scanning Microscopies, Electron,Scanning Microscopy, Electron
D001860 Bone Plates Implantable fracture fixation devices attached to bone fragments with screws to bridge the fracture gap and shield the fracture site from stress as bone heals. (UMDNS, 1999) Bone Plate,Plate, Bone,Plates, Bone
D001863 Bone Screws Specialized devices used in ORTHOPEDIC SURGERY to repair bone fractures. Bone Screw,Screw, Bone,Screws, Bone
D003343 Corrosion The gradual destruction of a metal or alloy due to oxidation or action of a chemical agent. (From McGraw-Hill Dictionary of Scientific and Technical Terms, 6th ed) Corrosions
D004577 Electron Probe Microanalysis Identification and measurement of ELEMENTS and their location based on the fact that X-RAYS emitted by an element excited by an electron beam have a wavelength characteristic of that element and an intensity related to its concentration. It is performed with an electron microscope fitted with an x-ray spectrometer, in scanning or transmission mode. Microscopy, Electron, X-Ray Microanalysis,Spectrometry, X-Ray Emission, Electron Microscopic,Spectrometry, X-Ray Emission, Electron Probe,X-Ray Emission Spectrometry, Electron Microscopic,X-Ray Emission Spectrometry, Electron Probe,X-Ray Microanalysis, Electron Microscopic,X-Ray Microanalysis, Electron Probe,Microanalysis, Electron Probe,Spectrometry, X Ray Emission, Electron Microscopic,Spectrometry, X Ray Emission, Electron Probe,X Ray Emission Spectrometry, Electron Microscopic,X Ray Emission Spectrometry, Electron Probe,X-Ray Microanalysis,Electron Probe Microanalyses,Microanalyses, Electron Probe,Microanalysis, X-Ray,Probe Microanalyses, Electron,Probe Microanalysis, Electron,X Ray Microanalysis,X Ray Microanalysis, Electron Microscopic,X Ray Microanalysis, Electron Probe
D005593 Fracture Fixation, Internal The use of internal devices (metal plates, nails, rods, etc.) to hold the position of a fracture in proper alignment. Osteosynthesis, Fracture,Fixation, Internal Fracture,Fixations, Internal Fracture,Fracture Fixations, Internal,Fracture Osteosyntheses,Fracture Osteosynthesis,Internal Fracture Fixation,Internal Fracture Fixations,Osteosyntheses, Fracture
D005599 Fractures, Ununited A fracture in which union fails to occur, the ends of the bone becoming rounded and eburnated, and a false joint occurs. (Stedman, 25th ed) Fracture, Ununited,Ununited Fracture,Ununited Fractures
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D000328 Adult A person having attained full growth or maturity. Adults are of 19 through 44 years of age. For a person between 19 and 24 years of age, YOUNG ADULT is available. Adults

Related Publications

M H Richman, and J K Weltman, and A Cole
May 1967, The Dental practitioner and dental record,
M H Richman, and J K Weltman, and A Cole
October 1969, The Dental practitioner and dental record,
M H Richman, and J K Weltman, and A Cole
January 1977, Acta orthopaedica Scandinavica,
M H Richman, and J K Weltman, and A Cole
September 1975, Journal of biomedical materials research,
M H Richman, and J K Weltman, and A Cole
January 1992, Injury,
M H Richman, and J K Weltman, and A Cole
January 1949, Actualites odonto-stomatologiques,
M H Richman, and J K Weltman, and A Cole
January 2012, The journal of contemporary dental practice,
M H Richman, and J K Weltman, and A Cole
February 1963, Zhonghua wai ke za zhi [Chinese journal of surgery],
M H Richman, and J K Weltman, and A Cole
November 1981, Journal of biomedical materials research,
M H Richman, and J K Weltman, and A Cole
April 1959, Rocky Mountain medical journal,
Copied contents to your clipboard!