Determination of 5-fluorouracil in microvolumes of human plasma by solvent extraction and high-performance liquid chromatography. 1999

W J Loos, and P de Bruijn, and L van Zuylen, and J Verweij, and K Nooter, and G Stoter, and A Sparreboom
Department of Medical Oncology, Rotterdam Cancer Institute (Daniel den Hoed Kliniek) and University Hospital Rotterdam, The Netherlands. loos@pcnh.azr.nl

In the present study, a new reversed-phase HPLC method has been developed and validated for the quantitative determination of 5-fluorouracil (5-FU) in human plasma using only 100-microl samples. The sample extraction and clean-up procedure involved a simple liquid-liquid extraction after addition of 5-chlorouracil (5-CU), used as internal standard, with 5 ml ethyl acetate. Chromatographic separations were performed on an Inertsil ODS-3 column (250x4.6 mm ID; 5 microM particle size), eluted with a mobile phase composed of acidified water (pH 2.0). The column effluent was monitored by UV absorption measurement at a wavelength of 266 nm. The calibration curves were constructed over a range of 0.20-50.0 microM and were fitted by weighted (1/x) linear regression analysis using the ratio of peak heights of 5-FU and 5-CU versus concentrations of the nominal standards. Extraction recoveries over the total range averaged 92 and 93% for 5-FU and 5-CU, respectively. The lower limit of quantitation was established at 0.20 microM (approximately 26 ng/ml), with within-run and between-run precisions of 4.2 and 7.0%, respectively, and an average accuracy of 109.3%. The within-run and between-run precisions at four tested concentrations analyzed in quintuplicate over a time period of four days were < 1.4 and < 4.4%, respectively. The accuracy at the tested concentrations ranged from 98.4 to 102.3%. Compared to previously described validated analytical methods for 5-FU, our present assay provides equivalent to superior sensitivity, using only microvolumes of sample.

UI MeSH Term Description Entries
D002138 Calibration Determination, by measurement or comparison with a standard, of the correct value of each scale reading on a meter or other measuring instrument; or determination of the settings of a control device that correspond to particular values of voltage, current, frequency or other output. Calibrations
D002851 Chromatography, High Pressure Liquid Liquid chromatographic techniques which feature high inlet pressures, high sensitivity, and high speed. Chromatography, High Performance Liquid,Chromatography, High Speed Liquid,Chromatography, Liquid, High Pressure,HPLC,High Performance Liquid Chromatography,High-Performance Liquid Chromatography,UPLC,Ultra Performance Liquid Chromatography,Chromatography, High-Performance Liquid,High-Performance Liquid Chromatographies,Liquid Chromatography, High-Performance
D005472 Fluorouracil A pyrimidine analog that is an antineoplastic antimetabolite. It interferes with DNA synthesis by blocking the THYMIDYLATE SYNTHETASE conversion of deoxyuridylic acid to thymidylic acid. 5-FU,5-FU Lederle,5-FU Medac,5-Fluorouracil,5-Fluorouracil-Biosyn,5-HU Hexal,5FU,Adrucil,Carac,Efudex,Efudix,Fluoro-Uracile ICN,Fluoroplex,Fluorouracil Mononitrate,Fluorouracil Monopotassium Salt,Fluorouracil Monosodium Salt,Fluorouracil Potassium Salt,Fluorouracil-GRY,Fluorouracile Dakota,Fluorouracilo Ferrer Far,Fluoruracil,Fluracedyl,Flurodex,Haemato-FU,Neofluor,Onkofluor,Ribofluor,5 FU Lederle,5 FU Medac,5 Fluorouracil,5 Fluorouracil Biosyn,5 HU Hexal,Dakota, Fluorouracile,Fluoro Uracile ICN,Fluorouracil GRY,Haemato FU
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D000964 Antimetabolites, Antineoplastic Antimetabolites that are useful in cancer chemotherapy. Antineoplastic Antimetabolites
D012997 Solvents Liquids that dissolve other substances (solutes), generally solids, without any change in chemical composition, as, water containing sugar. (Grant & Hackh's Chemical Dictionary, 5th ed) Solvent
D013056 Spectrophotometry, Ultraviolet Determination of the spectra of ultraviolet absorption by specific molecules in gases or liquids, for example Cl2, SO2, NO2, CS2, ozone, mercury vapor, and various unsaturated compounds. (McGraw-Hill Dictionary of Scientific and Technical Terms, 4th ed) Ultraviolet Spectrophotometry
D015203 Reproducibility of Results The statistical reproducibility of measurements (often in a clinical context), including the testing of instrumentation or techniques to obtain reproducible results. The concept includes reproducibility of physiological measurements, which may be used to develop rules to assess probability or prognosis, or response to a stimulus; reproducibility of occurrence of a condition; and reproducibility of experimental results. Reliability and Validity,Reliability of Result,Reproducibility Of Result,Reproducibility of Finding,Validity of Result,Validity of Results,Face Validity,Reliability (Epidemiology),Reliability of Results,Reproducibility of Findings,Test-Retest Reliability,Validity (Epidemiology),Finding Reproducibilities,Finding Reproducibility,Of Result, Reproducibility,Of Results, Reproducibility,Reliabilities, Test-Retest,Reliability, Test-Retest,Result Reliabilities,Result Reliability,Result Validities,Result Validity,Result, Reproducibility Of,Results, Reproducibility Of,Test Retest Reliability,Validity and Reliability,Validity, Face

Related Publications

W J Loos, and P de Bruijn, and L van Zuylen, and J Verweij, and K Nooter, and G Stoter, and A Sparreboom
February 2001, Journal of chromatography. B, Biomedical sciences and applications,
W J Loos, and P de Bruijn, and L van Zuylen, and J Verweij, and K Nooter, and G Stoter, and A Sparreboom
January 1993, Journal of chromatography,
W J Loos, and P de Bruijn, and L van Zuylen, and J Verweij, and K Nooter, and G Stoter, and A Sparreboom
March 1997, Se pu = Chinese journal of chromatography,
W J Loos, and P de Bruijn, and L van Zuylen, and J Verweij, and K Nooter, and G Stoter, and A Sparreboom
March 1997, Analytical biochemistry,
W J Loos, and P de Bruijn, and L van Zuylen, and J Verweij, and K Nooter, and G Stoter, and A Sparreboom
June 2006, Journal of chromatography. B, Analytical technologies in the biomedical and life sciences,
W J Loos, and P de Bruijn, and L van Zuylen, and J Verweij, and K Nooter, and G Stoter, and A Sparreboom
November 1983, Journal of chromatography,
W J Loos, and P de Bruijn, and L van Zuylen, and J Verweij, and K Nooter, and G Stoter, and A Sparreboom
October 1982, Journal of chromatography,
W J Loos, and P de Bruijn, and L van Zuylen, and J Verweij, and K Nooter, and G Stoter, and A Sparreboom
February 2000, Journal of chromatography. B, Biomedical sciences and applications,
W J Loos, and P de Bruijn, and L van Zuylen, and J Verweij, and K Nooter, and G Stoter, and A Sparreboom
January 1992, Journal of chromatography,
W J Loos, and P de Bruijn, and L van Zuylen, and J Verweij, and K Nooter, and G Stoter, and A Sparreboom
April 2001, Biomedical chromatography : BMC,
Copied contents to your clipboard!