Optimization and evaluation of atomic emission gas chromatographic detection for nitrogen using the 388 nm molecular emission spectral band. 2000

A M González, and P C Uden
Department of Chemistry, University of Massachusetts, Amherst 01003-4510, USA.

Nitrogen determination by gas chromatography with atomic emission (microwave-induced plasma) detection (GC-AED) has been accomplished using the 174 nm atomic emission line, but with very limited selectivity and sensitivity. Nitrogen can also be detected using the cyanogen (CN) molecular band at 388 nm. A commercial GC-AED system was modified to allow the use of the 388 nm line for nitrogen detection, giving an improvement of 100-fold in sensitivity and selectivity, when compared with the 174 nm mode. Limits of detection of 10 pg/s were common, representing a 10-fold improvement. Compound-independent behavior was found for the system, working with optimum operating conditions, while instrumental problems were clearly reflected by a drastic compound dependent behavior. Response factors showed an important dependency upon the concentration of the element present. This dependency affected the accuracy of the determination of empirical formulae using GC-AED.

UI MeSH Term Description Entries
D009584 Nitrogen An element with the atomic symbol N, atomic number 7, and atomic weight [14.00643; 14.00728]. Nitrogen exists as a diatomic gas and makes up about 78% of the earth's atmosphere by volume. It is a constituent of proteins and nucleic acids and found in all living cells.
D002138 Calibration Determination, by measurement or comparison with a standard, of the correct value of each scale reading on a meter or other measuring instrument; or determination of the settings of a control device that correspond to particular values of voltage, current, frequency or other output. Calibrations
D002849 Chromatography, Gas Fractionation of a vaporized sample as a consequence of partition between a mobile gaseous phase and a stationary phase held in a column. Two types are gas-solid chromatography, where the fixed phase is a solid, and gas-liquid, in which the stationary phase is a nonvolatile liquid supported on an inert solid matrix. Chromatography, Gas-Liquid,Gas Chromatography,Chromatographies, Gas,Chromatographies, Gas-Liquid,Chromatography, Gas Liquid,Gas Chromatographies,Gas-Liquid Chromatographies,Gas-Liquid Chromatography
D012680 Sensitivity and Specificity Binary classification measures to assess test results. Sensitivity or recall rate is the proportion of true positives. Specificity is the probability of correctly determining the absence of a condition. (From Last, Dictionary of Epidemiology, 2d ed) Specificity,Sensitivity,Specificity and Sensitivity

Related Publications

A M González, and P C Uden
January 1995, Journal of AOAC International,
A M González, and P C Uden
November 1991, Journal of chromatography,
A M González, and P C Uden
January 1980, Journal of analytical toxicology,
A M González, and P C Uden
July 1997, Analytical chemistry,
A M González, and P C Uden
May 1983, Journal of chromatography,
A M González, and P C Uden
November 2002, Journal of chromatography. A,
Copied contents to your clipboard!