Opacification of artificial intraocular lens: an electron microscopic study. 2001

T W Shek, and A Wong, and B Yau, and A K Yu
Department of Pathology, Queen Mary Hospital, Hong Kong. whshek@hkucc.hku.hk

This paper describes the usefulness of electron microscopy in the investigation of the cause of opacification of an intraocular artificial lens (IOL), which affected both the anterior and posterior surfaces of the IOL. The explanted lenses were remarkably similar and were uniformly opaque, with "reticulated" surfaces under dissecting and ordinary light microscopes. TEM showed that the surfaces of the explanted lenses were irregular, and there was a layer of electron-dense granular deposits that extended to a depth of approximately 5 microm into the lens substance. SEM showed a "cerebriform" lens surface with elevated areas alternating with depressed crevices, which corresponded nicely to the TEM appearance. Energy-dispersive x-ray analysis showed that the deposit was composed of calcium, oxygen, and phosphorus, which was later shown to be calcium hydroxyapatite by x-ray diffraction study. Electron microscopy has proven to be an essential tool in the investigation of the cause of this mysterious outbreak of opacification of the surfaces of the artificial lenses. Apart from directly visualizing the lens surfaces in a 2- and 3-dimensional manner, it also provides information on the elemental composition of the deposit. Such findings enable the clinicians and manufacturer to search for the underlying pathogenesis of the abnormal calcium hydroxyapatite crystals deposit.

UI MeSH Term Description Entries
D007910 Lenses, Intraocular Artificial implanted lenses. Implantable Contact Lens,Lens, Intraocular,Contact Lens, Implantable,Intraocular Lens,Intraocular Lenses,Lens, Implantable Contact
D008854 Microscopy, Electron Microscopy using an electron beam, instead of light, to visualize the sample, thereby allowing much greater magnification. The interactions of ELECTRONS with specimens are used to provide information about the fine structure of that specimen. In TRANSMISSION ELECTRON MICROSCOPY the reactions of the electrons that are transmitted through the specimen are imaged. In SCANNING ELECTRON MICROSCOPY an electron beam falls at a non-normal angle on the specimen and the image is derived from the reactions occurring above the plane of the specimen. Electron Microscopy
D008855 Microscopy, Electron, Scanning Microscopy in which the object is examined directly by an electron beam scanning the specimen point-by-point. The image is constructed by detecting the products of specimen interactions that are projected above the plane of the sample, such as backscattered electrons. Although SCANNING TRANSMISSION ELECTRON MICROSCOPY also scans the specimen point by point with the electron beam, the image is constructed by detecting the electrons, or their interaction products that are transmitted through the sample plane, so that is a form of TRANSMISSION ELECTRON MICROSCOPY. Scanning Electron Microscopy,Electron Scanning Microscopy,Electron Microscopies, Scanning,Electron Microscopy, Scanning,Electron Scanning Microscopies,Microscopies, Electron Scanning,Microscopies, Scanning Electron,Microscopy, Electron Scanning,Microscopy, Scanning Electron,Scanning Electron Microscopies,Scanning Microscopies, Electron,Scanning Microscopy, Electron
D011183 Postoperative Complications Pathologic processes that affect patients after a surgical procedure. They may or may not be related to the disease for which the surgery was done, and they may or may not be direct results of the surgery. Complication, Postoperative,Complications, Postoperative,Postoperative Complication
D002114 Calcinosis Pathologic deposition of calcium salts in tissues. Calcification, Pathologic,Calcinosis, Tumoral,Microcalcification,Microcalcinosis,Pathologic Calcification,Calcinoses,Calcinoses, Tumoral,Microcalcifications,Microcalcinoses,Tumoral Calcinoses,Tumoral Calcinosis
D002386 Cataract Partial or complete opacity on or in the lens or capsule of one or both eyes, impairing vision or causing blindness. The many kinds of cataract are classified by their morphology (size, shape, location) or etiology (cause and time of occurrence). (Dorland, 27th ed) Cataract, Membranous,Lens Opacities,Pseudoaphakia,Cataracts,Cataracts, Membranous,Lens Opacity,Membranous Cataract,Membranous Cataracts,Opacities, Lens,Opacity, Lens,Pseudoaphakias
D004577 Electron Probe Microanalysis Identification and measurement of ELEMENTS and their location based on the fact that X-RAYS emitted by an element excited by an electron beam have a wavelength characteristic of that element and an intensity related to its concentration. It is performed with an electron microscope fitted with an x-ray spectrometer, in scanning or transmission mode. Microscopy, Electron, X-Ray Microanalysis,Spectrometry, X-Ray Emission, Electron Microscopic,Spectrometry, X-Ray Emission, Electron Probe,X-Ray Emission Spectrometry, Electron Microscopic,X-Ray Emission Spectrometry, Electron Probe,X-Ray Microanalysis, Electron Microscopic,X-Ray Microanalysis, Electron Probe,Microanalysis, Electron Probe,Spectrometry, X Ray Emission, Electron Microscopic,Spectrometry, X Ray Emission, Electron Probe,X Ray Emission Spectrometry, Electron Microscopic,X Ray Emission Spectrometry, Electron Probe,X-Ray Microanalysis,Electron Probe Microanalyses,Microanalyses, Electron Probe,Microanalysis, X-Ray,Probe Microanalyses, Electron,Probe Microanalysis, Electron,X Ray Microanalysis,X Ray Microanalysis, Electron Microscopic,X Ray Microanalysis, Electron Probe
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D017886 Durapatite The mineral component of bones and teeth; it has been used therapeutically as a prosthetic aid and in the prevention and treatment of osteoporosis. Calcium Hydroxyapatite,Hydroxyapatite,Hydroxylapatite,Alveograf,Calcitite,Interpore-200,Interpore-500,Osprovit,Ossein-Hydroxyapatite Compound,Ossopan,Osteogen,Periograf,Hydroxyapatite, Calcium,Interpore 200,Interpore 500,Interpore200,Interpore500,Ossein Hydroxyapatite Compound
D019591 Pseudophakia Presence of an intraocular lens after cataract extraction.

Related Publications

T W Shek, and A Wong, and B Yau, and A K Yu
January 2021, Journal francais d'ophtalmologie,
T W Shek, and A Wong, and B Yau, and A K Yu
January 1987, Ophthalmologica. Journal international d'ophtalmologie. International journal of ophthalmology. Zeitschrift fur Augenheilkunde,
T W Shek, and A Wong, and B Yau, and A K Yu
January 1989, Ophthalmologica. Journal international d'ophtalmologie. International journal of ophthalmology. Zeitschrift fur Augenheilkunde,
T W Shek, and A Wong, and B Yau, and A K Yu
January 2018, Taiwan journal of ophthalmology,
T W Shek, and A Wong, and B Yau, and A K Yu
October 2020, Canadian journal of ophthalmology. Journal canadien d'ophtalmologie,
T W Shek, and A Wong, and B Yau, and A K Yu
March 1984, Nippon Ganka Gakkai zasshi,
T W Shek, and A Wong, and B Yau, and A K Yu
May 1986, Archives of ophthalmology (Chicago, Ill. : 1960),
T W Shek, and A Wong, and B Yau, and A K Yu
January 2007, Ceska a slovenska oftalmologie : casopis Ceske oftalmologicke spolecnosti a Slovenske oftalmologicke spolecnosti,
T W Shek, and A Wong, and B Yau, and A K Yu
September 2000, American journal of ophthalmology,
Copied contents to your clipboard!