Commissioning and quality assurance of the Pinnacle(3) radiotherapy treatment planning system for external beam photons. 2003

J L Bedford, and P J Childs, and V Nordmark Hansen, and M A Mosleh-Shirazi, and F Verhaegen, and A P Warrington
Joint Department of Physics, The Institute of Cancer Research and the Royal Marsden NHS Trust, Downs Road, Sutton, Surrey SM2 5PT, UK.

The commissioning of a Pinnacle(3) treatment planning system is described. Four Elekta linear accelerators were commissioned for external beam photons. Measured data were used to derive parameter values for the Pinnacle(3) beam model by (1). fitting a Monte Carlo model of the accelerator head to measured data and then extracting the parameters for the Pinnacle(3) beam model, and by (2). using the auto-modelling facility within Pinnacle(3). Both of these methods yielded dose distributions in accord with published recommendations. A separate small-field beam model, customized for an in-house compact blocking system, was also created, which satisfied appropriate acceptance criteria for stereotactically guided conformal brain treatments. Inhomogeneous, oblique, asymmetrical and irregular fields were also assessed, with calculated and measured doses agreeing to within +/-3%. Dose-volume histogram calculation was found to be accurate to within +/-5% dose or volume for a grid size of 4 mm x 4 mm x 4 mm, with better accuracy being achieved for finer grids. Isocentric doses were compared between Pinnacle(3)'s collapsed cone convolution algorithm and the Bentley-Milan algorithm within the Target-2 treatment planning system. Dose differences were generally less than 3% in the dose prescribed, with larger values for breast plans, where the Pinnacle(3) algorithm calculated scatter more accurately. Pelvic and thoracic plans were also verified using an anthropomorphic phantom, with local dose differences between calculated and delivered dose of up to 8%, but mainly less than 3%, and with no systematic difference. Ionization chamber verifications using START and RT-01 trial procedures demonstrated differences between calculated and measured doses of less than 2%. Following satisfactory performance in the commissioning process, Pinnacle(3) has now been introduced into routine clinical use.

UI MeSH Term Description Entries
D009010 Monte Carlo Method In statistics, a technique for numerically approximating the solution of a mathematical problem by studying the distribution of some random variable, often generated by a computer. The name alludes to the randomness characteristic of the games of chance played at the gambling casinos in Monte Carlo. (From Random House Unabridged Dictionary, 2d ed, 1993) Method, Monte Carlo
D010315 Particle Accelerators Devices which accelerate electrically charged atomic or subatomic particles, such as electrons, protons or ions, to high velocities so they have high kinetic energy. Betatrons,Linear Accelerators,Accelerator, Linear,Accelerator, Particle,Accelerators, Linear,Accelerators, Particle,Betatron,Linear Accelerator,Particle Accelerator
D010388 Pelvis The space or compartment surrounded by the pelvic girdle (bony pelvis). It is subdivided into the greater pelvis and LESSER PELVIS. The pelvic girdle is formed by the PELVIC BONES and SACRUM. Pelvic Region,Region, Pelvic
D011786 Quality Control A system for verifying and maintaining a desired level of quality in a product or process by careful planning, use of proper equipment, continued inspection, and corrective action as required. (Random House Unabridged Dictionary, 2d ed) Control, Quality,Controls, Quality,Quality Controls
D011879 Radiotherapy Dosage The total amount of radiation absorbed by tissues as a result of radiotherapy. Dosage, Radiotherapy,Dosages, Radiotherapy,Radiotherapy Dosages
D011880 Radiotherapy Planning, Computer-Assisted Computer-assisted mathematical calculations of beam angles, intensities of radiation, and duration of irradiation in radiotherapy. Computer-Assisted Radiotherapy Planning,Dosimetry Calculations, Computer-Assisted,Planning, Computer-Assisted Radiotherapy,Calculation, Computer-Assisted Dosimetry,Calculations, Computer-Assisted Dosimetry,Computer Assisted Radiotherapy Planning,Computer-Assisted Dosimetry Calculation,Computer-Assisted Dosimetry Calculations,Dosimetry Calculation, Computer-Assisted,Dosimetry Calculations, Computer Assisted,Planning, Computer Assisted Radiotherapy,Radiotherapy Planning, Computer Assisted
D012015 Reference Standards A basis of value established for the measure of quantity, weight, extent or quality, e.g. weight standards, standard solutions, methods, techniques, and procedures used in diagnosis and therapy. Standard Preparations,Standards, Reference,Preparations, Standard,Standardization,Standards,Preparation, Standard,Reference Standard,Standard Preparation,Standard, Reference
D001940 Breast In humans, one of the paired regions in the anterior portion of the THORAX. The breasts consist of the MAMMARY GLANDS, the SKIN, the MUSCLES, the ADIPOSE TISSUE, and the CONNECTIVE TISSUES. Breasts
D003198 Computer Simulation Computer-based representation of physical systems and phenomena such as chemical processes. Computational Modeling,Computational Modelling,Computer Models,In silico Modeling,In silico Models,In silico Simulation,Models, Computer,Computerized Models,Computer Model,Computer Simulations,Computerized Model,In silico Model,Model, Computer,Model, Computerized,Model, In silico,Modeling, Computational,Modeling, In silico,Modelling, Computational,Simulation, Computer,Simulation, In silico,Simulations, Computer
D005260 Female Females

Related Publications

J L Bedford, and P J Childs, and V Nordmark Hansen, and M A Mosleh-Shirazi, and F Verhaegen, and A P Warrington
March 2011, Journal of applied clinical medical physics,
J L Bedford, and P J Childs, and V Nordmark Hansen, and M A Mosleh-Shirazi, and F Verhaegen, and A P Warrington
January 2023, Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology,
J L Bedford, and P J Childs, and V Nordmark Hansen, and M A Mosleh-Shirazi, and F Verhaegen, and A P Warrington
March 2016, Journal of medical radiation sciences,
J L Bedford, and P J Childs, and V Nordmark Hansen, and M A Mosleh-Shirazi, and F Verhaegen, and A P Warrington
July 2013, Magnetic resonance in medicine,
J L Bedford, and P J Childs, and V Nordmark Hansen, and M A Mosleh-Shirazi, and F Verhaegen, and A P Warrington
May 1993, International journal of radiation oncology, biology, physics,
J L Bedford, and P J Childs, and V Nordmark Hansen, and M A Mosleh-Shirazi, and F Verhaegen, and A P Warrington
October 2008, International journal of radiation oncology, biology, physics,
J L Bedford, and P J Childs, and V Nordmark Hansen, and M A Mosleh-Shirazi, and F Verhaegen, and A P Warrington
January 2006, Journal of applied clinical medical physics,
J L Bedford, and P J Childs, and V Nordmark Hansen, and M A Mosleh-Shirazi, and F Verhaegen, and A P Warrington
January 2019, Nihon Hoshasen Gijutsu Gakkai zasshi,
J L Bedford, and P J Childs, and V Nordmark Hansen, and M A Mosleh-Shirazi, and F Verhaegen, and A P Warrington
May 1993, International journal of radiation oncology, biology, physics,
J L Bedford, and P J Childs, and V Nordmark Hansen, and M A Mosleh-Shirazi, and F Verhaegen, and A P Warrington
January 2012, Medical physics,
Copied contents to your clipboard!