Continuous intra-arterial oxygen monitoring: accuracy and reliability in the surgical intensive care unit. 1992

J F Lemus, and T Kearney, and D R Margulies, and D J Mackenzie, and B J Leyerle, and M M Shabot
Department of Surgery, Cedars-Sinai Medical Center, Los Angeles, California 90048.

The accuracy and reliability of an invasive intra-arterial oxygen sensor catheter was evaluated in 20 critically ill surgical intensive care unit (SICU) patients. All patients required continuous arterial blood pressure monitoring, at least 72 hours of ventilator support, and intermittent arterial blood gas sampling for clinical management. The intra-arterial sensor provided continuous PO2 (PsO2) values on a bedside electronic monitor. PsO2 values were sampled every 60 seconds and automatically stored on a bedside personal computer. Arterial blood gas (ABG) PaO2 values were collected and matched by collection time with corresponding PsO2 values. During 1,238 hours of continuous intra-arterial monitoring, 74,280 PsO2 values and 246 ABG PaO2 values were collected. Of the 246 PaO2 results, 175 (71.3%) had a matching PsO2. Regression of matched PsO2 and PaO2 values yielded a correlation coefficient of 0.58 and standard error of the estimate (SEE) of 33.1 (P < 0.0005). Even though matched PsO2 and PaO2 measurements demonstrated a linear relationship, only 34 per cent of the variation in PsO2 could be attributed to changes in PaO2. Technical sensor or instrument problems affected PsO2 monitoring in 17 of 20 patients and 28 of the 33 sensors tested. The authors conclude that continuous intra-arterial monitoring of PsO2 is a novel idea, but technical issues limit its use in acutely ill, conscious SICU patients.

UI MeSH Term Description Entries
D007362 Intensive Care Units Hospital units providing continuous surveillance and care to acutely ill patients. ICU Intensive Care Units,Intensive Care Unit,Unit, Intensive Care
D008297 Male Males
D008875 Middle Aged An adult aged 45 - 64 years. Middle Age
D008991 Monitoring, Physiologic The continuous measurement of physiological processes, blood pressure, heart rate, renal output, reflexes, respiration, etc., in a patient or experimental animal; includes pharmacologic monitoring, the measurement of administered drugs or their metabolites in the blood, tissues, or urine. Patient Monitoring,Monitoring, Physiological,Physiologic Monitoring,Monitoring, Patient,Physiological Monitoring
D010092 Oximetry The determination of oxygen-hemoglobin saturation of blood either by withdrawing a sample and passing it through a classical photoelectric oximeter or by electrodes attached to some translucent part of the body like finger, earlobe, or skin fold. It includes non-invasive oxygen monitoring by pulse oximetry. Pulse Oximetry,Oximetry, Pulse,Oximetries,Oximetries, Pulse,Pulse Oximetries
D010100 Oxygen An element with atomic symbol O, atomic number 8, and atomic weight [15.99903; 15.99977]. It is the most abundant element on earth and essential for respiration. Dioxygen,Oxygen-16,Oxygen 16
D011446 Prospective Studies Observation of a population for a sufficient number of persons over a sufficient number of years to generate incidence or mortality rates subsequent to the selection of the study group. Prospective Study,Studies, Prospective,Study, Prospective
D001784 Blood Gas Analysis Measurement of oxygen and carbon dioxide in the blood. Analysis, Blood Gas,Analyses, Blood Gas,Blood Gas Analyses,Gas Analyses, Blood,Gas Analysis, Blood
D001794 Blood Pressure PRESSURE of the BLOOD on the ARTERIES and other BLOOD VESSELS. Systolic Pressure,Diastolic Pressure,Pulse Pressure,Pressure, Blood,Pressure, Diastolic,Pressure, Pulse,Pressure, Systolic,Pressures, Systolic
D004868 Equipment Failure Failure of equipment to perform to standard. The failure may be due to defects or improper use. Defects, Equipment,Device Failure,Failure, Equipment,Malfunction, Equipment,Medical Device Failure,Misuse, Equipment,Device Failure, Medical,Device Failures, Medical,Failure, Medical Device,Failures, Medical Device,Defect, Equipment,Device Failures,Equipment Defect,Equipment Defects,Equipment Failures,Equipment Malfunction,Equipment Malfunctions,Equipment Misuse,Equipment Misuses,Failure, Device,Failures, Device,Failures, Equipment,Malfunctions, Equipment,Misuses, Equipment

Related Publications

J F Lemus, and T Kearney, and D R Margulies, and D J Mackenzie, and B J Leyerle, and M M Shabot
January 1977, Critical care medicine,
J F Lemus, and T Kearney, and D R Margulies, and D J Mackenzie, and B J Leyerle, and M M Shabot
November 2022, Journal of diabetes science and technology,
J F Lemus, and T Kearney, and D R Margulies, and D J Mackenzie, and B J Leyerle, and M M Shabot
August 2009, Anesthesia and analgesia,
J F Lemus, and T Kearney, and D R Margulies, and D J Mackenzie, and B J Leyerle, and M M Shabot
January 2002, Epilepsia,
J F Lemus, and T Kearney, and D R Margulies, and D J Mackenzie, and B J Leyerle, and M M Shabot
January 1999, Electroencephalography and clinical neurophysiology. Supplement,
J F Lemus, and T Kearney, and D R Margulies, and D J Mackenzie, and B J Leyerle, and M M Shabot
September 2016, Anaesthesia and intensive care,
J F Lemus, and T Kearney, and D R Margulies, and D J Mackenzie, and B J Leyerle, and M M Shabot
January 2017, Handbook of clinical neurology,
J F Lemus, and T Kearney, and D R Margulies, and D J Mackenzie, and B J Leyerle, and M M Shabot
February 1998, The Canadian journal of neurological sciences. Le journal canadien des sciences neurologiques,
J F Lemus, and T Kearney, and D R Margulies, and D J Mackenzie, and B J Leyerle, and M M Shabot
August 2012, Current neurology and neuroscience reports,
J F Lemus, and T Kearney, and D R Margulies, and D J Mackenzie, and B J Leyerle, and M M Shabot
September 2004, American journal of electroneurodiagnostic technology,
Copied contents to your clipboard!