Multicentre evaluation of the Boehringer Mannheim/Hitachi 747 analysis system. 1992

P Bonini, and F Ceriotti, and F Keller, and P Brauer, and H Stolz, and C Pascual, and L García Beltrán, and D J Vonderschmitt, and P Pei
Ospedale San Raffaele, Milan, Italy.

Analytical performance and practicability of the new Boehringer Mannheim/Hitachi 747 analysis system were assessed in a multicentre evaluation involving four laboratories. The analytical performance was evaluated according to a protocol similar to the ECCLS guidelines and comprised 13 analytes including enzymes, substrates and electrolytes. About 65,000 results were obtained within three months. The evaluation was planned and supported by a program system called "Computer Aided Evaluation". Acceptance criteria have been established for judging the results. The median of the within-run coefficients of variation (CVs) in control sera of all methods was below 1%, being far below the acceptance limit of 2%. The median of CVs of between-days imprecision was below 2% (acceptance criterion 3%). The high degree of precision prompted us to set up a biometrical model suitable for the differentiation between deviant points, outliers and measurements that can still be explained by the system performance. No relevant drift effects were observed during eight hours. The methods were linear over a wide range, avoiding rerun analysis in most cases. No sample-related carry-over was found. Reagent-dependent carry-over outside the acceptance limits was measured from uric acid to phosphorus to a slight extent, and from triacylglycerols to lipase, as well as from total protein to bilirubin to a perceptible degree. It can be avoided by separating these reagent combinations in the channel arrangement. Taking a systematic deviation of more than 10% as unacceptable, four of the 13 analytes suffered from interference by haemoglobin, one by bilirubin and one by turbidity. The Boehringer Mannheim/Hitachi 747 analysis system is capable of determining serum indices which in combination with the interferogram allow an assessment of the interference. With the exception of chloride the recovery of the assigned values for all control sera showed values between 95 and 105%. Out of 40 method comparison studies for enzymes and substrates, 31 yielded regression equations with less than 5% proportional errors and less than 5% constant errors. Deviations exceeding these acceptance criteria can be explained by differences in the reagent formulation, in the method employed or in calibration. The agreement of the ISE method comparisons was within a +/- 5% deviation over a wide analytical range. Practicability of the Boehringer Mannheim/Hitachi 747 analysis system was assessed with the help of a questionnaire, in which properties of the instrument were quantified, thus permitting a relatively objective rating. The 190 questions were placed in 14 groups, each dealing with an attribute of the instrument.(ABSTRACT TRUNCATED AT 400 WORDS)

UI MeSH Term Description Entries
D002138 Calibration Determination, by measurement or comparison with a standard, of the correct value of each scale reading on a meter or other measuring instrument; or determination of the settings of a control device that correspond to particular values of voltage, current, frequency or other output. Calibrations
D002624 Chemistry, Clinical The specialty of ANALYTIC CHEMISTRY applied to assays of physiologically important substances found in blood, urine, tissues, and other biological fluids for the purpose of aiding the physician in making a diagnosis or following therapy. Clinical Chemistry
D004573 Electrolytes Substances that dissociate into two or more ions, to some extent, in water. Solutions of electrolytes thus conduct an electric current and can be decomposed by it (ELECTROLYSIS). (Grant & Hackh's Chemical Dictionary, 5th ed) Electrolyte
D004798 Enzymes Biological molecules that possess catalytic activity. They may occur naturally or be synthetically created. Enzymes are usually proteins, however CATALYTIC RNA and CATALYTIC DNA molecules have also been identified. Biocatalyst,Enzyme,Biocatalysts
D005069 Evaluation Studies as Topic Works about studies that determine the effectiveness or value of processes, personnel, and equipment, or the material on conducting such studies. Critique,Evaluation Indexes,Evaluation Methodology,Evaluation Report,Evaluation Research,Methodology, Evaluation,Pre-Post Tests,Qualitative Evaluation,Quantitative Evaluation,Theoretical Effectiveness,Use-Effectiveness,Critiques,Effectiveness, Theoretical,Evaluation Methodologies,Evaluation Reports,Evaluation, Qualitative,Evaluation, Quantitative,Evaluations, Qualitative,Evaluations, Quantitative,Indexes, Evaluation,Methodologies, Evaluation,Pre Post Tests,Pre-Post Test,Qualitative Evaluations,Quantitative Evaluations,Report, Evaluation,Reports, Evaluation,Research, Evaluation,Test, Pre-Post,Tests, Pre-Post,Use Effectiveness
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D001322 Autoanalysis Method of analyzing chemicals using automation. Autoanalyses
D013056 Spectrophotometry, Ultraviolet Determination of the spectra of ultraviolet absorption by specific molecules in gases or liquids, for example Cl2, SO2, NO2, CS2, ozone, mercury vapor, and various unsaturated compounds. (McGraw-Hill Dictionary of Scientific and Technical Terms, 4th ed) Ultraviolet Spectrophotometry
D015203 Reproducibility of Results The statistical reproducibility of measurements (often in a clinical context), including the testing of instrumentation or techniques to obtain reproducible results. The concept includes reproducibility of physiological measurements, which may be used to develop rules to assess probability or prognosis, or response to a stimulus; reproducibility of occurrence of a condition; and reproducibility of experimental results. Reliability and Validity,Reliability of Result,Reproducibility Of Result,Reproducibility of Finding,Validity of Result,Validity of Results,Face Validity,Reliability (Epidemiology),Reliability of Results,Reproducibility of Findings,Test-Retest Reliability,Validity (Epidemiology),Finding Reproducibilities,Finding Reproducibility,Of Result, Reproducibility,Of Results, Reproducibility,Reliabilities, Test-Retest,Reliability, Test-Retest,Result Reliabilities,Result Reliability,Result Validities,Result Validity,Result, Reproducibility Of,Results, Reproducibility Of,Test Retest Reliability,Validity and Reliability,Validity, Face

Related Publications

P Bonini, and F Ceriotti, and F Keller, and P Brauer, and H Stolz, and C Pascual, and L García Beltrán, and D J Vonderschmitt, and P Pei
April 1990, Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie,
P Bonini, and F Ceriotti, and F Keller, and P Brauer, and H Stolz, and C Pascual, and L García Beltrán, and D J Vonderschmitt, and P Pei
January 1993, The Journal of automatic chemistry,
P Bonini, and F Ceriotti, and F Keller, and P Brauer, and H Stolz, and C Pascual, and L García Beltrán, and D J Vonderschmitt, and P Pei
December 1987, Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie,
P Bonini, and F Ceriotti, and F Keller, and P Brauer, and H Stolz, and C Pascual, and L García Beltrán, and D J Vonderschmitt, and P Pei
June 1994, European journal of clinical chemistry and clinical biochemistry : journal of the Forum of European Clinical Chemistry Societies,
P Bonini, and F Ceriotti, and F Keller, and P Brauer, and H Stolz, and C Pascual, and L García Beltrán, and D J Vonderschmitt, and P Pei
June 1989, Clinical chemistry,
P Bonini, and F Ceriotti, and F Keller, and P Brauer, and H Stolz, and C Pascual, and L García Beltrán, and D J Vonderschmitt, and P Pei
December 1990, Clinical biochemistry,
P Bonini, and F Ceriotti, and F Keller, and P Brauer, and H Stolz, and C Pascual, and L García Beltrán, and D J Vonderschmitt, and P Pei
April 1983, Clinical chemistry,
P Bonini, and F Ceriotti, and F Keller, and P Brauer, and H Stolz, and C Pascual, and L García Beltrán, and D J Vonderschmitt, and P Pei
January 1992, Wiener klinische Wochenschrift. Supplementum,
P Bonini, and F Ceriotti, and F Keller, and P Brauer, and H Stolz, and C Pascual, and L García Beltrán, and D J Vonderschmitt, and P Pei
December 1996, European journal of clinical chemistry and clinical biochemistry : journal of the Forum of European Clinical Chemistry Societies,
P Bonini, and F Ceriotti, and F Keller, and P Brauer, and H Stolz, and C Pascual, and L García Beltrán, and D J Vonderschmitt, and P Pei
January 1998, Wiener klinische Wochenschrift,
Copied contents to your clipboard!