Reflectance pulse oximetry from core body in neonates and infants: comparison to arterial blood oxygen saturation and to transmission pulse oximetry. 2004

Amir Kugelman, and Yoram Wasserman, and Frida Mor, and Leonid Goldinov, and Yoav Geller, and David Bader
Department of Neonatology, Bnai Zion Medical Center, B Rappaport Faculty of Medicine, Technion-IIT, 47 Golomb Street, Haifa 31048, Israel.

OBJECTIVE To compare pulse oximetry oxygen saturation (SpO(2)) measured by a novel reflectance method from core body to arterial oxygen saturation (SaO(2)) in neonates and infants. Transmission pulse oximetry (TPO) was measured for comparison. METHODS We monitored 18 infants by the two pulse oximeters simultaneously. The reflectance pulse oximetry (RPO) (PRO2, ConMed, Utica, NY) was measured on the upper back or chest, while the TPO (N395-Nellcor, Pleasanton, CA) was measured from the finger of the infant on the left hand or feet. Data from the two methods were compared to functional SaO(2) derived from blood sample drawn from arterial line for patient care and measured by a Co-oximeter (Ilex, Instrument Lab. Lexington, MA). The potential advantage of the RPO is demonstrated in a case of a premature infant with hypovolemic shock, where SaO(2) or TPO could not be obtained but oximetry was available from the RPO. RESULTS We used for analysis 56 RPO and 32 TPO measurements. SpO(2) obtained from the RPO was 88.3+/-9.8%, from the TPO 84.2+/-10.1%, and functional SaO(2) was 88.2+/-11.7%, with correlation coefficient of 0.93 and 0.88, respectively (p<0.0001). The mean difference (bias) and standard deviation of the differences (precision) between the RPO and the TPO compared to functional SaO(2) were -0.09+/-4.5% and 1.26+/-5.9% and the absolute errors were 3.2+/-3.1%, and 4.4+/-4.0%, respectively. The accuracy of both RPO and TPO was diminished when SaO(2) was <85%, but only the RPO remained correlated with the functional SaO(2). CONCLUSIONS Reflectance pulse oximetry measured from core body of neonates and infants is accurate and reliable and is comparable to the transmission SpO(2) when compared to functional SaO(2). We speculate that the reflectance method might be advantageous in cases of poor peripheral perfusion in neonates and infants.

UI MeSH Term Description Entries
D007223 Infant A child between 1 and 23 months of age. Infants
D007231 Infant, Newborn An infant during the first 28 days after birth. Neonate,Newborns,Infants, Newborn,Neonates,Newborn,Newborn Infant,Newborn Infants
D008297 Male Males
D010092 Oximetry The determination of oxygen-hemoglobin saturation of blood either by withdrawing a sample and passing it through a classical photoelectric oximeter or by electrodes attached to some translucent part of the body like finger, earlobe, or skin fold. It includes non-invasive oxygen monitoring by pulse oximetry. Pulse Oximetry,Oximetry, Pulse,Oximetries,Oximetries, Pulse,Pulse Oximetries
D010100 Oxygen An element with atomic symbol O, atomic number 8, and atomic weight [15.99903; 15.99977]. It is the most abundant element on earth and essential for respiration. Dioxygen,Oxygen-16,Oxygen 16
D001785 Blood Gas Monitoring, Transcutaneous The noninvasive measurement or determination of the partial pressure (tension) of oxygen and/or carbon dioxide locally in the capillaries of a tissue by the application to the skin of a special set of electrodes. These electrodes contain photoelectric sensors capable of picking up the specific wavelengths of radiation emitted by oxygenated versus reduced hemoglobin. Carbon Dioxide Partial Pressure Determination, Transcutaneous,Cutaneous Oximetry,Oximetry, Transcutaneous,Oxygen Partial Pressure Determination, Transcutaneous,Transcutaneous Blood Gas Monitoring,Transcutaneous Capnometry,Transcutaneous Oximetry,PtcO2,TcPCO2,Capnometries, Transcutaneous,Capnometry, Transcutaneous,Cutaneous Oximetries,Oximetries, Cutaneous,Oximetries, Transcutaneous,Oximetry, Cutaneous,Transcutaneous Capnometries,Transcutaneous Oximetries
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man

Related Publications

Amir Kugelman, and Yoram Wasserman, and Frida Mor, and Leonid Goldinov, and Yoav Geller, and David Bader
August 1995, European journal of obstetrics, gynecology, and reproductive biology,
Amir Kugelman, and Yoram Wasserman, and Frida Mor, and Leonid Goldinov, and Yoav Geller, and David Bader
April 1987, AORN journal,
Amir Kugelman, and Yoram Wasserman, and Frida Mor, and Leonid Goldinov, and Yoav Geller, and David Bader
January 2012, Neonatology,
Amir Kugelman, and Yoram Wasserman, and Frida Mor, and Leonid Goldinov, and Yoav Geller, and David Bader
March 1992, Lancet (London, England),
Amir Kugelman, and Yoram Wasserman, and Frida Mor, and Leonid Goldinov, and Yoav Geller, and David Bader
April 2012, Journal of wildlife diseases,
Amir Kugelman, and Yoram Wasserman, and Frida Mor, and Leonid Goldinov, and Yoav Geller, and David Bader
April 1995, Laboratory animal science,
Amir Kugelman, and Yoram Wasserman, and Frida Mor, and Leonid Goldinov, and Yoav Geller, and David Bader
December 2018, Aerospace medicine and human performance,
Amir Kugelman, and Yoram Wasserman, and Frida Mor, and Leonid Goldinov, and Yoav Geller, and David Bader
October 2018, Medicine and science in sports and exercise,
Amir Kugelman, and Yoram Wasserman, and Frida Mor, and Leonid Goldinov, and Yoav Geller, and David Bader
December 1987, Critical care medicine,
Amir Kugelman, and Yoram Wasserman, and Frida Mor, and Leonid Goldinov, and Yoav Geller, and David Bader
January 1992, Biomedical instrumentation & technology,
Copied contents to your clipboard!