A self-calibrating telemetry system for measurement of ventricular pressure-volume relations in conscious, freely moving rats. 2004

Kazunori Uemura, and Toru Kawada, and Masaru Sugimachi, and Can Zheng, and Koji Kashihara, and Takayuki Sato, and Kenji Sunagawa
Dept. of Cardiovascular Dynamics, National Cardiovascular Center Research Institute, 5-7-1, Fujishirodai, Suita 565-8565, Japan. kuemura@ri.ncvc.go.jp

Using Bluetooth wireless technology, we developed an implantable telemetry system for measurement of the left ventricular pressure-volume relation in conscious, freely moving rats. The telemetry system consisted of a pressure-conductance catheter (1.8-Fr) connected to a small (14-g) fully implantable signal transmitter. To make the system fully telemetric, calibrations such as blood resistivity and parallel conductance were also conducted telemetrically. To estimate blood resistivity, we used four electrodes arranged 0.2 mm apart on the pressure-conductance catheter. To estimate parallel conductance, we used a dual-frequency method. We examined the accuracy of calibrations, stroke volume (SV) measurements, and the reproducibility of the telemetry. The blood resistivity estimated telemetrically agreed with that measured using an ex vivo cuvette method (y=1.09x - 11.9, r2= 0.88, n=10). Parallel conductance estimated by the dual-frequency (2 and 20 kHz) method correlated well with that measured by a conventional saline injection method (y=1.59x - 1.77, r2= 0.87, n=13). The telemetric SV closely correlated with the flowmetric SV during inferior vena cava occlusions (y=0.96x + 7.5, r2=0.96, n=4). In six conscious rats, differences between the repeated telemetries on different days (3 days apart on average) were reasonably small: 13% for end-diastolic volume, 20% for end-systolic volume, 28% for end-diastolic pressure, and 6% for end-systolic pressure. We conclude that the developed telemetry system enables us to estimate the pressure-volume relation with reasonable accuracy and reproducibility in conscious, untethered rats.

UI MeSH Term Description Entries
D009043 Motor Activity Body movements of a human or an animal as a behavioral phenomenon. Activities, Motor,Activity, Motor,Motor Activities
D010827 Physiology The biological science concerned with the life-supporting properties, functions, and processes of living organisms or their parts.
D002138 Calibration Determination, by measurement or comparison with a standard, of the correct value of each scale reading on a meter or other measuring instrument; or determination of the settings of a control device that correspond to particular values of voltage, current, frequency or other output. Calibrations
D002306 Cardiac Volume The volume of the HEART, usually relating to the volume of BLOOD contained within it at various periods of the cardiac cycle. The amount of blood ejected from a ventricle at each beat is STROKE VOLUME. Heart Volume,Cardiac Volumes,Heart Volumes,Volume, Cardiac,Volume, Heart,Volumes, Cardiac,Volumes, Heart
D003243 Consciousness Sense of awareness of self and of the environment. Consciousnesses
D000818 Animals Unicellular or multicellular, heterotrophic organisms, that have sensation and the power of voluntary movement. Under the older five kingdom paradigm, Animalia was one of the kingdoms. Under the modern three domain model, Animalia represents one of the many groups in the domain EUKARYOTA. Animal,Metazoa,Animalia
D013686 Telemetry Transmission of the readings of instruments to a remote location by means of wires, radio waves, or other means. (McGraw-Hill Dictionary of Scientific and Technical Terms, 4th ed) Telemetries
D015203 Reproducibility of Results The statistical reproducibility of measurements (often in a clinical context), including the testing of instrumentation or techniques to obtain reproducible results. The concept includes reproducibility of physiological measurements, which may be used to develop rules to assess probability or prognosis, or response to a stimulus; reproducibility of occurrence of a condition; and reproducibility of experimental results. Reliability and Validity,Reliability of Result,Reproducibility Of Result,Reproducibility of Finding,Validity of Result,Validity of Results,Face Validity,Reliability (Epidemiology),Reliability of Results,Reproducibility of Findings,Test-Retest Reliability,Validity (Epidemiology),Finding Reproducibilities,Finding Reproducibility,Of Result, Reproducibility,Of Results, Reproducibility,Reliabilities, Test-Retest,Reliability, Test-Retest,Result Reliabilities,Result Reliability,Result Validities,Result Validity,Result, Reproducibility Of,Results, Reproducibility Of,Test Retest Reliability,Validity and Reliability,Validity, Face
D017725 Ventricular Pressure The pressure within a CARDIAC VENTRICLE. Ventricular pressure waveforms can be measured in the beating heart by catheterization or estimated using imaging techniques (e.g., DOPPLER ECHOCARDIOGRAPHY). The information is useful in evaluating the function of the MYOCARDIUM; CARDIAC VALVES; and PERICARDIUM, particularly with simultaneous measurement of other (e.g., aortic or atrial) pressures. Intraventricular Pressure,Intraventricular Pressures,Pressure, Intraventricular,Pressure, Ventricular,Pressures, Intraventricular,Pressures, Ventricular,Ventricular Pressures
D051381 Rats The common name for the genus Rattus. Rattus,Rats, Laboratory,Rats, Norway,Rattus norvegicus,Laboratory Rat,Laboratory Rats,Norway Rat,Norway Rats,Rat,Rat, Laboratory,Rat, Norway,norvegicus, Rattus

Related Publications

Kazunori Uemura, and Toru Kawada, and Masaru Sugimachi, and Can Zheng, and Koji Kashihara, and Takayuki Sato, and Kenji Sunagawa
January 2016, Journal of pharmacological and toxicological methods,
Kazunori Uemura, and Toru Kawada, and Masaru Sugimachi, and Can Zheng, and Koji Kashihara, and Takayuki Sato, and Kenji Sunagawa
July 2013, Laboratory animals,
Kazunori Uemura, and Toru Kawada, and Masaru Sugimachi, and Can Zheng, and Koji Kashihara, and Takayuki Sato, and Kenji Sunagawa
December 2013, Journal of applied physiology (Bethesda, Md. : 1985),
Kazunori Uemura, and Toru Kawada, and Masaru Sugimachi, and Can Zheng, and Koji Kashihara, and Takayuki Sato, and Kenji Sunagawa
August 1996, Journal of applied physiology (Bethesda, Md. : 1985),
Kazunori Uemura, and Toru Kawada, and Masaru Sugimachi, and Can Zheng, and Koji Kashihara, and Takayuki Sato, and Kenji Sunagawa
May 2016, Journal of visualized experiments : JoVE,
Kazunori Uemura, and Toru Kawada, and Masaru Sugimachi, and Can Zheng, and Koji Kashihara, and Takayuki Sato, and Kenji Sunagawa
November 1995, The Journal of physiology,
Kazunori Uemura, and Toru Kawada, and Masaru Sugimachi, and Can Zheng, and Koji Kashihara, and Takayuki Sato, and Kenji Sunagawa
January 2000, Journal of pharmacological and toxicological methods,
Kazunori Uemura, and Toru Kawada, and Masaru Sugimachi, and Can Zheng, and Koji Kashihara, and Takayuki Sato, and Kenji Sunagawa
January 2018, Methods in molecular biology (Clifton, N.J.),
Kazunori Uemura, and Toru Kawada, and Masaru Sugimachi, and Can Zheng, and Koji Kashihara, and Takayuki Sato, and Kenji Sunagawa
January 2002, Hippocampus,
Kazunori Uemura, and Toru Kawada, and Masaru Sugimachi, and Can Zheng, and Koji Kashihara, and Takayuki Sato, and Kenji Sunagawa
March 2012, Journal of the American Association for Laboratory Animal Science : JAALAS,
Copied contents to your clipboard!