Stability of pacing threshold, impedance, and R wave amplitude at rest and during exercise. 1990

A Schuchert, and K H Kuck, and W Bleifeld
Department of Cardiology, University Hospital Eppendorf, Hamburg, Federal Republic of Germany.

The pacing threshold of the human heart may be altered by physiological factors such as physical exercise. These changes may influence the individual programming of a pacemaker, since pacemakers can be programmed at pulse amplitudes of 2.5 volts and less. We investigated 22 patients with a multiprogrammable ventricular demand pacemaker 3 months after implantation; 16 patients had received a steroid-eluting lead and six patients had an Elgiloy lead. Parameters measured at rest and immediately after exercise were: Voltage threshold at pulse durations between 0.05 and 0.6 ms, impedance, R wave amplitude and energy consumption for the pacing threshold at 0.5 ms pulse duration. All patients performed a symptom limited supine bicycle exercise test. None of the investigated parameters showed a significant difference between rest and exercise, neither for the steroid eluting lead nor for the Elgiloy lead. The data suggest that the individual programming of a pacemaker adapted to the measurements at rest is also reliable and safe during exercise.

UI MeSH Term Description Entries
D008297 Male Males
D011674 Pulse The rhythmical expansion and contraction of an ARTERY produced by waves of pressure caused by the ejection of BLOOD from the left ventricle of the HEART as it contracts. Pulses
D012146 Rest Freedom from activity. Rests
D002304 Cardiac Pacing, Artificial Regulation of the rate of contraction of the heart muscles by an artificial pacemaker. Pacing, Cardiac, Artificial,Artificial Cardiac Pacing,Artificial Cardiac Pacings,Cardiac Pacings, Artificial,Pacing, Artificial Cardiac,Pacings, Artificial Cardiac
D004562 Electrocardiography Recording of the moment-to-moment electromotive forces of the HEART as projected onto various sites on the body's surface, delineated as a scalar function of time. The recording is monitored by a tracing on slow moving chart paper or by observing it on a cardioscope, which is a CATHODE RAY TUBE DISPLAY. 12-Lead ECG,12-Lead EKG,12-Lead Electrocardiography,Cardiography,ECG,EKG,Electrocardiogram,Electrocardiograph,12 Lead ECG,12 Lead EKG,12 Lead Electrocardiography,12-Lead ECGs,12-Lead EKGs,12-Lead Electrocardiographies,Cardiographies,ECG, 12-Lead,EKG, 12-Lead,Electrocardiograms,Electrocardiographies, 12-Lead,Electrocardiographs,Electrocardiography, 12-Lead
D005080 Exercise Test Controlled physical activity which is performed in order to allow assessment of physiological functions, particularly cardiovascular and pulmonary, but also aerobic capacity. Maximal (most intense) exercise is usually required but submaximal exercise is also used. Arm Ergometry Test,Bicycle Ergometry Test,Cardiopulmonary Exercise Testing,Exercise Testing,Step Test,Stress Test,Treadmill Test,Cardiopulmonary Exercise Test,EuroFit Tests,Eurofit Test Battery,European Fitness Testing Battery,Fitness Testing,Physical Fitness Testing,Arm Ergometry Tests,Bicycle Ergometry Tests,Cardiopulmonary Exercise Tests,Ergometry Test, Arm,Ergometry Test, Bicycle,Ergometry Tests, Arm,Ergometry Tests, Bicycle,EuroFit Test,Eurofit Test Batteries,Exercise Test, Cardiopulmonary,Exercise Testing, Cardiopulmonary,Exercise Tests,Exercise Tests, Cardiopulmonary,Fitness Testing, Physical,Fitness Testings,Step Tests,Stress Tests,Test Battery, Eurofit,Test, Arm Ergometry,Test, Bicycle Ergometry,Test, Cardiopulmonary Exercise,Test, EuroFit,Test, Exercise,Test, Step,Test, Stress,Test, Treadmill,Testing, Cardiopulmonary Exercise,Testing, Exercise,Testing, Fitness,Testing, Physical Fitness,Tests, Arm Ergometry,Tests, Bicycle Ergometry,Tests, Cardiopulmonary Exercise,Tests, EuroFit,Tests, Exercise,Tests, Step,Tests, Stress,Tests, Treadmill,Treadmill Tests
D005260 Female Females
D006339 Heart Rate The number of times the HEART VENTRICLES contract per unit of time, usually per minute. Cardiac Rate,Chronotropism, Cardiac,Heart Rate Control,Heartbeat,Pulse Rate,Cardiac Chronotropy,Cardiac Chronotropism,Cardiac Rates,Chronotropy, Cardiac,Control, Heart Rate,Heart Rates,Heartbeats,Pulse Rates,Rate Control, Heart,Rate, Cardiac,Rate, Heart,Rate, Pulse
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D000368 Aged A person 65 years of age or older. For a person older than 79 years, AGED, 80 AND OVER is available. Elderly

Related Publications

A Schuchert, and K H Kuck, and W Bleifeld
August 1981, British heart journal,
A Schuchert, and K H Kuck, and W Bleifeld
January 1983, Clinical cardiology,
A Schuchert, and K H Kuck, and W Bleifeld
April 1990, The American journal of cardiology,
A Schuchert, and K H Kuck, and W Bleifeld
September 1984, Journal of the American College of Cardiology,
A Schuchert, and K H Kuck, and W Bleifeld
January 1981, Bollettino della Societa italiana di cardiologia,
A Schuchert, and K H Kuck, and W Bleifeld
January 1983, Cor et vasa,
A Schuchert, and K H Kuck, and W Bleifeld
February 1981, The American journal of cardiology,
A Schuchert, and K H Kuck, and W Bleifeld
September 1986, Pacing and clinical electrophysiology : PACE,
A Schuchert, and K H Kuck, and W Bleifeld
February 1983, The American journal of cardiology,
A Schuchert, and K H Kuck, and W Bleifeld
March 1997, Journal of the American College of Cardiology,
Copied contents to your clipboard!