Clinical assessment of three-dimensional ultrasound prostate localization for external beam radiotherapy. 2006

Nigel P Orton, and Hazim A Jaradat, and Wolfgang A Tomé
Department of Human Oncology, University of Wisconsin, Madison, Wisconsin 53792, USA.

Three-dimensional ultrasound localization has been performed for external beam prostate treatments at our institution since September 2001. This article presents data from the daily shifts for 221 patients and 5005 fractions, and the results of tests performed to assess the system's performance under clinical conditions. Three tests are presented: (1) To measure the accuracy of the shifts, eight patients treated on a helical tomotherapy machine were localized daily using both ultrasound (US) and a megavoltage computed tomography (MVCT) scan. Comparison of the shifts showed that US localization improved alignment for six of the eight patients when compared to alignment using skin marks alone. The mean US-MVCT vector for these six patients was 3.1+/-1.3 mm, compared to 5.1+/-2.1 mm between the MVCT and the skin marks. The other two patients were identified as poor candidates for US prior to their first treatment fraction. (2) To assess the extent of intrafraction motion, US localization was repeated after treatment for six patients and a total of 29 fractions. The mean intrafraction prostate shift was 1.9+/-1.0 mm, and the shift was within the 3 mm localization uncertainty [Tomé et al., Med. Phys. 29, 1781-1788 (2002); in New Technologies in Radiation Oncology, edited by W. Schlegel, T. Bortfelde, and A. Grosu (Springer, Berlin, 2005)] of the system for 25 of 29 fractions. (3) To assess the interuser variation in shifts, four experienced operators independently localized five patients for five consecutive fractions. The standard deviation of the users' shifts was found to be approximately the same as the system's localization uncertainty. For shifts larger than the system localization uncertainty, the standard deviation of the users' shifts was nearly always much smaller than the mean shift. Taken together with the results of the US-MVCT comparison, this indicates that the shifts improved patient localization despite differences between users.

UI MeSH Term Description Entries
D008297 Male Males
D011471 Prostatic Neoplasms Tumors or cancer of the PROSTATE. Cancer of Prostate,Prostate Cancer,Cancer of the Prostate,Neoplasms, Prostate,Neoplasms, Prostatic,Prostate Neoplasms,Prostatic Cancer,Cancer, Prostate,Cancer, Prostatic,Cancers, Prostate,Cancers, Prostatic,Neoplasm, Prostate,Neoplasm, Prostatic,Prostate Cancers,Prostate Neoplasm,Prostatic Cancers,Prostatic Neoplasm
D011878 Radiotherapy The use of IONIZING RADIATION to treat malignant NEOPLASMS and some benign conditions. Radiotherapy, Targeted,Targeted Radiotherapy,Radiation Therapy,Radiation Therapy, Targeted,Radiation Treatment,Targeted Radiation Therapy,Radiation Therapies,Radiation Therapies, Targeted,Radiation Treatments,Radiotherapies,Radiotherapies, Targeted,Targeted Radiation Therapies,Targeted Radiotherapies,Therapies, Radiation,Therapies, Targeted Radiation,Therapy, Radiation,Therapy, Targeted Radiation,Treatment, Radiation
D011880 Radiotherapy Planning, Computer-Assisted Computer-assisted mathematical calculations of beam angles, intensities of radiation, and duration of irradiation in radiotherapy. Computer-Assisted Radiotherapy Planning,Dosimetry Calculations, Computer-Assisted,Planning, Computer-Assisted Radiotherapy,Calculation, Computer-Assisted Dosimetry,Calculations, Computer-Assisted Dosimetry,Computer Assisted Radiotherapy Planning,Computer-Assisted Dosimetry Calculation,Computer-Assisted Dosimetry Calculations,Dosimetry Calculation, Computer-Assisted,Dosimetry Calculations, Computer Assisted,Planning, Computer Assisted Radiotherapy,Radiotherapy Planning, Computer Assisted
D005069 Evaluation Studies as Topic Works about studies that determine the effectiveness or value of processes, personnel, and equipment, or the material on conducting such studies. Critique,Evaluation Indexes,Evaluation Methodology,Evaluation Report,Evaluation Research,Methodology, Evaluation,Pre-Post Tests,Qualitative Evaluation,Quantitative Evaluation,Theoretical Effectiveness,Use-Effectiveness,Critiques,Effectiveness, Theoretical,Evaluation Methodologies,Evaluation Reports,Evaluation, Qualitative,Evaluation, Quantitative,Evaluations, Qualitative,Evaluations, Quantitative,Indexes, Evaluation,Methodologies, Evaluation,Pre Post Tests,Pre-Post Test,Qualitative Evaluations,Quantitative Evaluations,Report, Evaluation,Reports, Evaluation,Research, Evaluation,Test, Pre-Post,Tests, Pre-Post,Use Effectiveness
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D000465 Algorithms A procedure consisting of a sequence of algebraic formulas and/or logical steps to calculate or determine a given task. Algorithm
D014057 Tomography, X-Ray Computed Tomography using x-ray transmission and a computer algorithm to reconstruct the image. CAT Scan, X-Ray,CT Scan, X-Ray,Cine-CT,Computerized Tomography, X-Ray,Electron Beam Computed Tomography,Tomodensitometry,Tomography, Transmission Computed,X-Ray Tomography, Computed,CAT Scan, X Ray,CT X Ray,Computed Tomography, X-Ray,Computed X Ray Tomography,Computerized Tomography, X Ray,Electron Beam Tomography,Tomography, X Ray Computed,Tomography, X-Ray Computer Assisted,Tomography, X-Ray Computerized,Tomography, X-Ray Computerized Axial,Tomography, Xray Computed,X Ray Computerized Tomography,X Ray Tomography, Computed,X-Ray Computer Assisted Tomography,X-Ray Computerized Axial Tomography,Beam Tomography, Electron,CAT Scans, X-Ray,CT Scan, X Ray,CT Scans, X-Ray,CT X Rays,Cine CT,Computed Tomography, Transmission,Computed Tomography, X Ray,Computed Tomography, Xray,Computed X-Ray Tomography,Scan, X-Ray CAT,Scan, X-Ray CT,Scans, X-Ray CAT,Scans, X-Ray CT,Tomographies, Computed X-Ray,Tomography, Computed X-Ray,Tomography, Electron Beam,Tomography, X Ray Computer Assisted,Tomography, X Ray Computerized,Tomography, X Ray Computerized Axial,Transmission Computed Tomography,X Ray Computer Assisted Tomography,X Ray Computerized Axial Tomography,X Ray, CT,X Rays, CT,X-Ray CAT Scan,X-Ray CAT Scans,X-Ray CT Scan,X-Ray CT Scans,X-Ray Computed Tomography,X-Ray Computerized Tomography,Xray Computed Tomography
D014463 Ultrasonography The visualization of deep structures of the body by recording the reflections or echoes of ultrasonic pulses directed into the tissues. Use of ultrasound for imaging or diagnostic purposes employs frequencies ranging from 1.6 to 10 megahertz. Echography,Echotomography,Echotomography, Computer,Sonography, Medical,Tomography, Ultrasonic,Ultrasonic Diagnosis,Ultrasonic Imaging,Ultrasonographic Imaging,Computer Echotomography,Diagnosis, Ultrasonic,Diagnostic Ultrasound,Ultrasonic Tomography,Ultrasound Imaging,Diagnoses, Ultrasonic,Diagnostic Ultrasounds,Imaging, Ultrasonic,Imaging, Ultrasonographic,Imaging, Ultrasound,Imagings, Ultrasonographic,Imagings, Ultrasound,Medical Sonography,Ultrasonic Diagnoses,Ultrasonographic Imagings,Ultrasound, Diagnostic,Ultrasounds, Diagnostic
D015203 Reproducibility of Results The statistical reproducibility of measurements (often in a clinical context), including the testing of instrumentation or techniques to obtain reproducible results. The concept includes reproducibility of physiological measurements, which may be used to develop rules to assess probability or prognosis, or response to a stimulus; reproducibility of occurrence of a condition; and reproducibility of experimental results. Reliability and Validity,Reliability of Result,Reproducibility Of Result,Reproducibility of Finding,Validity of Result,Validity of Results,Face Validity,Reliability (Epidemiology),Reliability of Results,Reproducibility of Findings,Test-Retest Reliability,Validity (Epidemiology),Finding Reproducibilities,Finding Reproducibility,Of Result, Reproducibility,Of Results, Reproducibility,Reliabilities, Test-Retest,Reliability, Test-Retest,Result Reliabilities,Result Reliability,Result Validities,Result Validity,Result, Reproducibility Of,Results, Reproducibility Of,Test Retest Reliability,Validity and Reliability,Validity, Face

Related Publications

Nigel P Orton, and Hazim A Jaradat, and Wolfgang A Tomé
January 1978, International journal of radiation oncology, biology, physics,
Nigel P Orton, and Hazim A Jaradat, and Wolfgang A Tomé
August 2002, International journal of radiation oncology, biology, physics,
Nigel P Orton, and Hazim A Jaradat, and Wolfgang A Tomé
October 2011, Cancer,
Nigel P Orton, and Hazim A Jaradat, and Wolfgang A Tomé
November 2004, International journal of radiation oncology, biology, physics,
Nigel P Orton, and Hazim A Jaradat, and Wolfgang A Tomé
July 1996, International journal of radiation oncology, biology, physics,
Nigel P Orton, and Hazim A Jaradat, and Wolfgang A Tomé
May 2013, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology,
Nigel P Orton, and Hazim A Jaradat, and Wolfgang A Tomé
September 2010, Panminerva medica,
Nigel P Orton, and Hazim A Jaradat, and Wolfgang A Tomé
May 2010, Journal of endourology,
Copied contents to your clipboard!