Measurement of the replacement correction factor for parallel-plate chambers in electron fields. 1991

C S Reft, and F T Kuchnir
Department of Radiation and Cellular Oncology, University of Chicago, Illinois 60637.

When parallel-plate chambers are used for dosimetry in electron fields, the AAPM dosimetry protocol recommends a value of 1.0 for the replacement correction factor, P(repl),pp,E, until further data become available. Here, P(repl),pp,E for five commercially available parallel-plate chambers was measured as a function of electron energy from a nominal value of 5.5 to 22 MeV by comparison with a cylindrical chamber whose P(repl),cyl,E was obtained from data in the protocol. Since this method is based on the concept of a constant value for Ngas,pp, the energy and modality dependence of Ngap,pp is also investigated for these chambers for Co-60, 4-, 6-, 24-MV photons and for 22-MeV electrons. It is found that for three of the chambers P(repl),pp,E is independent of energy, consistent with unity within one or two standard deviations (s.d.). For the fourth chamber P(repl),pp,E is similarly consistent with one above 10 MeV, but decreases at lower energies, while for the fifth one it shows a systematic drop with decreasing energy.

UI MeSH Term Description Entries
D011874 Radiometry The measurement of radiation by photography, as in x-ray film and film badge, by Geiger-Mueller tube, and by SCINTILLATION COUNTING. Geiger-Mueller Counters,Nuclear Track Detection,Radiation Dosimetry,Dosimetry, Radiation,Geiger Counter,Geiger-Mueller Counter Tube,Geiger-Mueller Probe,Geiger-Mueller Tube,Radiation Counter,Counter Tube, Geiger-Mueller,Counter Tubes, Geiger-Mueller,Counter, Geiger,Counter, Radiation,Counters, Geiger,Counters, Geiger-Mueller,Counters, Radiation,Detection, Nuclear Track,Dosimetries, Radiation,Geiger Counters,Geiger Mueller Counter Tube,Geiger Mueller Counters,Geiger Mueller Probe,Geiger Mueller Tube,Geiger-Mueller Counter Tubes,Geiger-Mueller Probes,Geiger-Mueller Tubes,Probe, Geiger-Mueller,Probes, Geiger-Mueller,Radiation Counters,Radiation Dosimetries,Tube, Geiger-Mueller,Tube, Geiger-Mueller Counter,Tubes, Geiger-Mueller,Tubes, Geiger-Mueller Counter
D002138 Calibration Determination, by measurement or comparison with a standard, of the correct value of each scale reading on a meter or other measuring instrument; or determination of the settings of a control device that correspond to particular values of voltage, current, frequency or other output. Calibrations
D004583 Electrons Stable elementary particles having the smallest known negative charge, present in all elements; also called negatrons. Positively charged electrons are called positrons. The numbers, energies and arrangement of electrons around atomic nuclei determine the chemical identities of elements. Beams of electrons are called CATHODE RAYS. Fast Electrons,Negatrons,Positrons,Electron,Electron, Fast,Electrons, Fast,Fast Electron,Negatron,Positron

Related Publications

C S Reft, and F T Kuchnir
February 2010, Medical physics,
C S Reft, and F T Kuchnir
January 1988, Medical physics,
C S Reft, and F T Kuchnir
November 1997, Physics in medicine and biology,
C S Reft, and F T Kuchnir
June 2006, Medical physics,
C S Reft, and F T Kuchnir
September 2014, Zeitschrift fur medizinische Physik,
C S Reft, and F T Kuchnir
June 1991, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology,
C S Reft, and F T Kuchnir
April 2012, Physics in medicine and biology,
C S Reft, and F T Kuchnir
January 1989, Medical physics,
Copied contents to your clipboard!