Commissioning and clinical implementation of a mega-voltage cone beam CT system for treatment localization. 2007

Olivier Gayou, and Moyed Miften
Department of Radiation Oncology, Allegheny General Hospital, Pittsburgh, Pennsylvania 15212, USA. ogayou@wpahs.org

The improvement in conformal radiotherapy techniques with steep dose gradients has allowed for the delivery of higher doses to a tumor volume while maintaining the sparing of surrounding normal tissue. In this situation, verification of patient setup and evaluation of internal organ motion just prior to radiation delivery is a crucial step. To this end, several volumetric image-guided techniques have been developed for patient localization, such as the Siemens MVision mega-voltage cone beam CT (MV-CBCT) system. In this work, the commissioning and clinical implementation of the MVision system is presented. The geometry and gain calibration procedures for the system are described, and guidelines for quality assurance procedures are provided. Different MV-CBCT clinical protocols, ranging from daily to weekly image-guidance, which includes image acquisition, reconstruction, registration with planning CT, and treatment couch offsets corrections, were commissioned. The image quality characteristics of the MVision system were measured and assessed qualitatively and quantitatively, including the image noise and uniformity, low-contrast resolution, and spatial resolution. Furthermore, the image reconstruction and registration software was evaluated. Data show that a 2 cm large object with 1% electron density contrast can be detected with the MVision system with 10 cGy at isocenter and that the registration software is accurate within 2 mm in the anterior-posterior, left-right, and superior-inferior directions.

UI MeSH Term Description Entries
D007091 Image Processing, Computer-Assisted A technique of inputting two-dimensional or three-dimensional images into a computer and then enhancing or analyzing the imagery into a form that is more useful to the human observer. Biomedical Image Processing,Computer-Assisted Image Processing,Digital Image Processing,Image Analysis, Computer-Assisted,Image Reconstruction,Medical Image Processing,Analysis, Computer-Assisted Image,Computer-Assisted Image Analysis,Computer Assisted Image Analysis,Computer Assisted Image Processing,Computer-Assisted Image Analyses,Image Analyses, Computer-Assisted,Image Analysis, Computer Assisted,Image Processing, Biomedical,Image Processing, Computer Assisted,Image Processing, Digital,Image Processing, Medical,Image Processings, Medical,Image Reconstructions,Medical Image Processings,Processing, Biomedical Image,Processing, Digital Image,Processing, Medical Image,Processings, Digital Image,Processings, Medical Image,Reconstruction, Image,Reconstructions, Image
D009369 Neoplasms New abnormal growth of tissue. Malignant neoplasms show a greater degree of anaplasia and have the properties of invasion and metastasis, compared to benign neoplasms. Benign Neoplasm,Cancer,Malignant Neoplasm,Tumor,Tumors,Benign Neoplasms,Malignancy,Malignant Neoplasms,Neoplasia,Neoplasm,Neoplasms, Benign,Cancers,Malignancies,Neoplasias,Neoplasm, Benign,Neoplasm, Malignant,Neoplasms, Malignant
D011786 Quality Control A system for verifying and maintaining a desired level of quality in a product or process by careful planning, use of proper equipment, continued inspection, and corrective action as required. (Random House Unabridged Dictionary, 2d ed) Control, Quality,Controls, Quality,Quality Controls
D011856 Radiographic Image Enhancement Improvement in the quality of an x-ray image by use of an intensifying screen, tube, or filter and by optimum exposure techniques. Digital processing methods are often employed. Digital Radiography,Image Enhancement, Radiographic,Radiography, Digital,Enhancement, Radiographic Image,Enhancements, Radiographic Image,Image Enhancements, Radiographic,Radiographic Image Enhancements
D011879 Radiotherapy Dosage The total amount of radiation absorbed by tissues as a result of radiotherapy. Dosage, Radiotherapy,Dosages, Radiotherapy,Radiotherapy Dosages
D011880 Radiotherapy Planning, Computer-Assisted Computer-assisted mathematical calculations of beam angles, intensities of radiation, and duration of irradiation in radiotherapy. Computer-Assisted Radiotherapy Planning,Dosimetry Calculations, Computer-Assisted,Planning, Computer-Assisted Radiotherapy,Calculation, Computer-Assisted Dosimetry,Calculations, Computer-Assisted Dosimetry,Computer Assisted Radiotherapy Planning,Computer-Assisted Dosimetry Calculation,Computer-Assisted Dosimetry Calculations,Dosimetry Calculation, Computer-Assisted,Dosimetry Calculations, Computer Assisted,Planning, Computer Assisted Radiotherapy,Radiotherapy Planning, Computer Assisted
D002138 Calibration Determination, by measurement or comparison with a standard, of the correct value of each scale reading on a meter or other measuring instrument; or determination of the settings of a control device that correspond to particular values of voltage, current, frequency or other output. Calibrations
D003287 Contrast Media Substances used to allow enhanced visualization of tissues. Radiopaque Media,Contrast Agent,Contrast Agents,Contrast Material,Contrast Materials,Radiocontrast Agent,Radiocontrast Agents,Radiocontrast Media,Agent, Contrast,Agent, Radiocontrast,Agents, Contrast,Agents, Radiocontrast,Material, Contrast,Materials, Contrast,Media, Contrast,Media, Radiocontrast,Media, Radiopaque
D004583 Electrons Stable elementary particles having the smallest known negative charge, present in all elements; also called negatrons. Positively charged electrons are called positrons. The numbers, energies and arrangement of electrons around atomic nuclei determine the chemical identities of elements. Beams of electrons are called CATHODE RAYS. Fast Electrons,Negatrons,Positrons,Electron,Electron, Fast,Electrons, Fast,Fast Electron,Negatron,Positron
D004867 Equipment Design Methods and patterns of fabricating machines and related hardware. Design, Equipment,Device Design,Medical Device Design,Design, Medical Device,Designs, Medical Device,Device Design, Medical,Device Designs, Medical,Medical Device Designs,Design, Device,Designs, Device,Designs, Equipment,Device Designs,Equipment Designs

Related Publications

Olivier Gayou, and Moyed Miften
November 2012, Journal of applied clinical medical physics,
Olivier Gayou, and Moyed Miften
July 2021, Radiation oncology (London, England),
Olivier Gayou, and Moyed Miften
February 2012, Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al],
Olivier Gayou, and Moyed Miften
July 2020, Physics in medicine and biology,
Olivier Gayou, and Moyed Miften
January 2013, Medical dosimetry : official journal of the American Association of Medical Dosimetrists,
Olivier Gayou, and Moyed Miften
February 2013, Physics in medicine and biology,
Olivier Gayou, and Moyed Miften
August 2013, IEEE transactions on medical imaging,
Olivier Gayou, and Moyed Miften
July 2013, Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al],
Olivier Gayou, and Moyed Miften
October 2009, Cancer radiotherapie : journal de la Societe francaise de radiotherapie oncologique,
Copied contents to your clipboard!