Multilayer piezoelectric ceramics for two-dimensional array transducers. 1994

R L Goldberg, and S W Smith
Dept. of Biomed. Eng., Duke Univ., Durham, NC.

In medical ultrasound imaging, 2-D array transducers have become essential to implement dynamic focusing and phase-correction in the elevation dimension as well as real-time volumetric scanning. Unfortunately, the small size of a 2-D array element results in a small clamped capacitance and a large electrical impedance near resonance. These elements have poor sensitivity because their impedance is much higher than the electrical impedance of the transmit and receive circuitry. Sensitivity can be improved by using an N layer structure of PZT ceramic with the layers connected acoustically in series and electrically in parallel. For the multilayer ceramic (MLC), the damped capacitance is multiplied by a factor of N(2) and the electrical impedance by 1/N(2) compared to a single layer element of the same dimensions. A 3x43 phased-array transducer has been fabricated using 3 layer PZT-5H material. Each element had a thickness of 0.66 mm and an area of 0.37x3.5 mm. The MLC was manufactured using thick film technology with plated-through vias to electrically interconnect the electrode layers. The completed transducer was compared to a single layer control array of similar dimensions. With a light epoxy backing and a lambda/4 matching layer, the MLC array elements had an impedance of 100 Omega at series resonance of 2.25 MHz, compared to 800 Omega for the control elements. The lower impedance of the MLC elements resulted in a minimum round-trip insertion loss of 24.0 dB, compared to an 34.1 dB for the control array elements. These results were consistent with KLM modeling. B-scan images were made of cysts in a tissue-mimicking phantom and of the left kidney in vivo. The images clearly showed a higher signal-to-noise ratio for the MLC array compared to the control. As a result, 2-D arrays made of multilayer ceramics can be used to form images at a higher frequency and greater range than single layer arrays.

UI MeSH Term Description Entries

Related Publications

R L Goldberg, and S W Smith
October 2005, IEEE transactions on ultrasonics, ferroelectrics, and frequency control,
R L Goldberg, and S W Smith
May 2002, Ultrasonics,
R L Goldberg, and S W Smith
January 2000, IEEE transactions on ultrasonics, ferroelectrics, and frequency control,
R L Goldberg, and S W Smith
January 1993, IEEE transactions on ultrasonics, ferroelectrics, and frequency control,
R L Goldberg, and S W Smith
January 1994, IEEE transactions on ultrasonics, ferroelectrics, and frequency control,
R L Goldberg, and S W Smith
July 2012, Applied physics letters,
Copied contents to your clipboard!