[Determination of kynurenine in serum using high performance liquid chromatography-fluorescence detection]. 2008

Xibo Luo, and Aiguo Tang, and Langan Pi, and Ledong Xiao, and Xiang Ao, and Yanhong Pu, and Rui Wang
Department of Clinical Laboratory, the Second Xiangya Hospital of Central South University, Changsha 4100, China. 120002721@sina.com.com

A sensitive method of high performance liquid chromatography-fluorescence detection (HPLC-FLD) for the determination of kynurenine (Kyn) in serum was developed. A 20 microL supernatant of a serum sample deproteinized by 5% perchloric acid solution was assayed and separated on a Hypersil C8 column (300 mm x 6.0 mm, 10 microm) with an isocratic elution of 0.25 mol/L zinc acetate-50 mmol/L acetate containing 3% (v/v) acetonitrile. The fluorescence excitation and emission wavelengths were 365 nm and 480 nm, respectively. The limit of detection was approximately 0.04 micromol/L (signal-to-noise ratio of 3) and the linearity of the assay was from 0.098 micromol/L to 19.6 micromol/L. The relative standard deviations of intraday and interday determinations were 3.8% and 4.6%, respectively. The results indicated tryptophan (Trp), 5-hydroxytryptamine (5-HT), kynurenic acid (KYNA), phenylalanine (Phe), tyrosine (Tyr) and creatinine (Cr) had no interfering effects to the determination of Kyn. The method is of high accuracy, easy operation, satisfactory recoveries and good reproducibility, and can be used for routine analysis.

UI MeSH Term Description Entries
D007700 Kinetics The rate dynamics in chemical or physical systems.
D007737 Kynurenine A metabolite of the essential amino acid tryptophan metabolized via the tryptophan-kynurenine pathway.
D001774 Blood Chemical Analysis An examination of chemicals in the blood. Analysis, Blood Chemical,Chemical Analysis, Blood,Analyses, Blood Chemical,Blood Chemical Analyses,Chemical Analyses, Blood
D002138 Calibration Determination, by measurement or comparison with a standard, of the correct value of each scale reading on a meter or other measuring instrument; or determination of the settings of a control device that correspond to particular values of voltage, current, frequency or other output. Calibrations
D002851 Chromatography, High Pressure Liquid Liquid chromatographic techniques which feature high inlet pressures, high sensitivity, and high speed. Chromatography, High Performance Liquid,Chromatography, High Speed Liquid,Chromatography, Liquid, High Pressure,HPLC,High Performance Liquid Chromatography,High-Performance Liquid Chromatography,UPLC,Ultra Performance Liquid Chromatography,Chromatography, High-Performance Liquid,High-Performance Liquid Chromatographies,Liquid Chromatography, High-Performance
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D013050 Spectrometry, Fluorescence Measurement of the intensity and quality of fluorescence. Fluorescence Spectrophotometry,Fluorescence Spectroscopy,Spectrofluorometry,Fluorescence Spectrometry,Spectrophotometry, Fluorescence,Spectroscopy, Fluorescence
D015203 Reproducibility of Results The statistical reproducibility of measurements (often in a clinical context), including the testing of instrumentation or techniques to obtain reproducible results. The concept includes reproducibility of physiological measurements, which may be used to develop rules to assess probability or prognosis, or response to a stimulus; reproducibility of occurrence of a condition; and reproducibility of experimental results. Reliability and Validity,Reliability of Result,Reproducibility Of Result,Reproducibility of Finding,Validity of Result,Validity of Results,Face Validity,Reliability (Epidemiology),Reliability of Results,Reproducibility of Findings,Test-Retest Reliability,Validity (Epidemiology),Finding Reproducibilities,Finding Reproducibility,Of Result, Reproducibility,Of Results, Reproducibility,Reliabilities, Test-Retest,Reliability, Test-Retest,Result Reliabilities,Result Reliability,Result Validities,Result Validity,Result, Reproducibility Of,Results, Reproducibility Of,Test Retest Reliability,Validity and Reliability,Validity, Face
D016014 Linear Models Statistical models in which the value of a parameter for a given value of a factor is assumed to be equal to a + bx, where a and b are constants. The models predict a linear regression. Linear Regression,Log-Linear Models,Models, Linear,Linear Model,Linear Regressions,Log Linear Models,Log-Linear Model,Model, Linear,Model, Log-Linear,Models, Log-Linear,Regression, Linear,Regressions, Linear
D057230 Limit of Detection Concentration or quantity that is derived from the smallest measure that can be detected with reasonable certainty for a given analytical procedure. Limits of Detection,Detection Limit,Detection Limits

Related Publications

Xibo Luo, and Aiguo Tang, and Langan Pi, and Ledong Xiao, and Xiang Ao, and Yanhong Pu, and Rui Wang
November 1985, Journal of chromatography,
Xibo Luo, and Aiguo Tang, and Langan Pi, and Ledong Xiao, and Xiang Ao, and Yanhong Pu, and Rui Wang
November 1985, Journal of chromatography,
Xibo Luo, and Aiguo Tang, and Langan Pi, and Ledong Xiao, and Xiang Ao, and Yanhong Pu, and Rui Wang
December 2002, Hunan yi ke da xue xue bao = Hunan yike daxue xuebao = Bulletin of Hunan Medical University,
Xibo Luo, and Aiguo Tang, and Langan Pi, and Ledong Xiao, and Xiang Ao, and Yanhong Pu, and Rui Wang
February 1984, Journal of chromatography,
Xibo Luo, and Aiguo Tang, and Langan Pi, and Ledong Xiao, and Xiang Ao, and Yanhong Pu, and Rui Wang
October 1990, Journal of chromatography,
Xibo Luo, and Aiguo Tang, and Langan Pi, and Ledong Xiao, and Xiang Ao, and Yanhong Pu, and Rui Wang
December 1985, The Analyst,
Xibo Luo, and Aiguo Tang, and Langan Pi, and Ledong Xiao, and Xiang Ao, and Yanhong Pu, and Rui Wang
March 2006, Se pu = Chinese journal of chromatography,
Xibo Luo, and Aiguo Tang, and Langan Pi, and Ledong Xiao, and Xiang Ao, and Yanhong Pu, and Rui Wang
March 1995, Journal of chromatography. B, Biomedical applications,
Xibo Luo, and Aiguo Tang, and Langan Pi, and Ledong Xiao, and Xiang Ao, and Yanhong Pu, and Rui Wang
August 1984, Journal of chromatography,
Xibo Luo, and Aiguo Tang, and Langan Pi, and Ledong Xiao, and Xiang Ao, and Yanhong Pu, and Rui Wang
November 1980, Journal of chromatography,
Copied contents to your clipboard!