Investigation of optimum X-ray beam tube voltage and filtration for chest radiography with a computed radiography system. 2008

C S Moore, and A W Beavis, and J R Saunderson
Radiation Physics Department, Hull & East Yorkshire Hospitals, Princess Royal Hospital, Saltshouse Road, Kingston Upon Hull, UK. craig.moore@hey.nhs.uk

The purpose of this study was to determine the optimum tube voltage and amount of added copper (Cu) filtration for processed chest radiographs obtained with an Agfa 75.0 Computed Radiography (CR) system. The contrast-to-noise ratio (CNR) was measured in the lung, heart/spine and diaphragm compartments of a validated chest phantom using various tube voltages and amounts of Cu filtration. The CNR was derived as a function of air kerma at the CR plate and with the effective dose. As rib contrast can interfere with detection of nodules in chest radiography, a tissue-to-rib ratio (TRR) was derived to investigate which tube voltages suppress the contrast of rib. Although processing algorithms affect the signal and noise in a way that is hard to predict, we found that, for a given set of processing parameters, the CNR was related to the plate air kerma and effective dose in a logarithmic manner (all R(2) >or=0.97). For imaging of the lung region, a low voltage (60 kVp) produced the highest CNR, whereas a high voltage (125 kVp) produced the highest TRR. In the heart/spine region, 80-125 kVp produced the highest CNR, while in the diaphragm region 60-90 kVp produced the highest CNR. For chest radiography with this CR system, the optimal tube voltage depends upon the region of interest. Of the filters tested, a 0.1 mm Cu thickness was found to provide a statistically significant increase in the CNR in the diaphragm region with tube potentials of 60 kVp and 80 kVp, without affecting the CNR in the other anatomical compartments.

UI MeSH Term Description Entries
D008168 Lung Either of the pair of organs occupying the cavity of the thorax that effect the aeration of the blood. Lungs
D011829 Radiation Dosage The amount of radiation energy that is deposited in a unit mass of material, such as tissues of plants or animal. In RADIOTHERAPY, radiation dosage is expressed in gray units (Gy). In RADIOLOGIC HEALTH, the dosage is expressed by the product of absorbed dose (Gy) and quality factor (a function of linear energy transfer), and is called radiation dose equivalent in sievert units (Sv). Sievert Units,Dosage, Radiation,Gray Units,Gy Radiation,Sv Radiation Dose Equivalent,Dosages, Radiation,Radiation Dosages,Units, Gray,Units, Sievert
D011857 Radiographic Image Interpretation, Computer-Assisted Computer systems or networks designed to provide radiographic interpretive information. Computer Assisted Radiographic Image Interpretation,Computer-Assisted Radiographic Image Interpretation,Radiographic Image Interpretation, Computer Assisted
D005374 Filtration A process of separating particulate matter from a fluid, such as air or a liquid, by passing the fluid carrier through a medium that will not pass the particulates. (McGraw-Hill Dictionary of Scientific and Technical Terms, 4th ed) Filtrations
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D012272 Ribs A set of twelve curved bones which connect to the vertebral column posteriorly, and terminate anteriorly as costal cartilage. Together, they form a protective cage around the internal thoracic organs. Rib
D013902 Radiography, Thoracic X-ray visualization of the chest and organs of the thoracic cavity. It is not restricted to visualization of the lungs. Thoracic Radiography,Radiographies, Thoracic,Thoracic Radiographies
D014962 X-Ray Film A film base coated with an emulsion designed for use with x-rays. Film, X-Ray,Radiographic Film,Film, Radiographic,Xray Film,Film, X Ray,Film, Xray,Films, Radiographic,Films, X-Ray,Films, Xray,Radiographic Films,X Ray Film,X-Ray Films,Xray Films
D019047 Phantoms, Imaging Devices or objects in various imaging techniques used to visualize or enhance visualization by simulating conditions encountered in the procedure. Phantoms are used very often in procedures employing or measuring x-irradiation or radioactive material to evaluate performance. Phantoms often have properties similar to human tissue. Water demonstrates absorbing properties similar to normal tissue, hence water-filled phantoms are used to map radiation levels. Phantoms are used also as teaching aids to simulate real conditions with x-ray or ultrasonic machines. (From Iturralde, Dictionary and Handbook of Nuclear Medicine and Clinical Imaging, 1990) Phantoms, Radiographic,Phantoms, Radiologic,Radiographic Phantoms,Radiologic Phantoms,Phantom, Radiographic,Phantom, Radiologic,Radiographic Phantom,Radiologic Phantom,Imaging Phantom,Imaging Phantoms,Phantom, Imaging

Related Publications

C S Moore, and A W Beavis, and J R Saunderson
June 2019, Scientific reports,
C S Moore, and A W Beavis, and J R Saunderson
February 1994, Medical physics,
C S Moore, and A W Beavis, and J R Saunderson
September 2023, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine,
C S Moore, and A W Beavis, and J R Saunderson
May 2005, The British journal of radiology,
C S Moore, and A W Beavis, and J R Saunderson
July 2021, Scientific reports,
C S Moore, and A W Beavis, and J R Saunderson
March 1996, Investigative radiology,
C S Moore, and A W Beavis, and J R Saunderson
April 1996, Health physics,
C S Moore, and A W Beavis, and J R Saunderson
October 2011, Medical physics,
C S Moore, and A W Beavis, and J R Saunderson
February 1995, Journal of digital imaging,
C S Moore, and A W Beavis, and J R Saunderson
November 1998, Medical physics,
Copied contents to your clipboard!