Flow-injection determination of phosphate with a cadmium ion-selective electrode. 1990

D E Davey, and D E Mulcahy, and G R O'Connell
School of Chemical Technology, S.A. Institute of Technology, P.O. Box 1, Ingle Farm, South Australia 5098.

A flow-injection method is described, in which phosphate standards are introduced into a reagent stream containing Cd(2+) ,resulting in the formation of Cd(3)(PO(4))(2). The associated reduction in free metal concentration is sensed by a cadmium-selective electrode. With the exception of major interference from iodide and moderate interference from bromide and thiocyanate, the system exhibits excellent response to phosphate and selectivity over several common anions in solutions buffered at pH 8.4. A maximum sampling rate of 160/hr is possible for phosphate standards in the concentration range 10(-1)-10(-1)M with a 10(-4)M Cd(2+) reagent stream at a total flow-rate (carrier and reagent stream combined) of 8.4 ml/min.