No single test is comprehensive enough to detect all of the variants of von Willebrand Disease (VWD), making determination of both concentration and function of von Willebrand Factor (VWF) important for an accurate diagnosis. The objective of the study was to validate a newly developed VWF collagen binding assay (VWF:CB) and VWF antigen enzyme-linked immunosorbent assay (ELISA) developed at the Ontario Veterinary College (OVC VWF:Ag). Linearity, sensitivity, and coefficients of variation were determined. The Asserachrom VWF:Ag ELISA was used as the reference assay for this study. Concordance correlation and Bland-Altman plots were used to evaluate agreement between both VWF:Ag assays. The VWF:CB accuracy was assessed by degree of association with the VWF:Ag assays, and the VWF:Ag to VWF:CB ratio. All assays were assessed for their ability to distinguish between VWD negative and VWD positive patients. Linearity, intra-assay coefficients of variation, and inter-assay coefficients of variation were acceptable for both the newly developed VWF:CB (R2 = 0.97, average CV = 4.4, and 15, respectively) and OVC VWF:Ag assays (R2 = 0.96, average CV = 7.9, and 5.9, respectively). Agreement between the OVC VWF:Ag assay and reference assay was excellent (rho(c) = 0.89), and although differences between assay results precluded interchangeable use of the assays, both successfully distinguished VWD positive and VWD negative dogs (P < 0.0001). The VWF:CB showed a strong association with both VWF:Ag assays (R2 = 0.86, 0.82) and VWF:Ag to VWF:CB ratios (< or = 1) were as expected. The excellent performance of both assays in this validation study confirm their reliability and potential for clinical application.