[The Dosage and Quality of Imaging program and quality assurance in mammography]. 1991

O Rimondi, and M Gambaccini, and M Marziani, and G C Candini, and P L Indovina, and A Toti, and E De Guglielmo, and A Calicchia
Dipartimento di Fisica, Università, Ferrara.

The aim of the optimization program for mammography (DQM) in Italy is to achieve the best compromise between image quality and dose to patient. 272 centers agreed to the second phase of the program, from February 1987 to January 1990. Exposure, half-value layer (HVL) and focal spot size were measured according to the same methods employed in the first phase of the DQM program (1985), while image quality was evaluated by means of a new performance phantom. The average exposure was 0.96 R (2.48 x 10(-4) C/kg); in 173 (64.3%) centers exposure was less than 1 R (2.58 x 10(-4) C/kg), and only in 3 centers, where a direct X-ray film was employed, was exposure greater than 5 R (12.9 x 10(-4) C/kg). In every center the average whole-breast dose to a reference organ (5 cm thick, composed of 50% fat + 50% water) was calculated on the basis of entrance exposure, HVL, and focus-skin distance; in 63.2% of the centers doses less than 0.15 cGy were employed. The results allowed dose and image quality to be correlated in order to divide the centers (using a film-screen system) into groups with a different efficiency level: in 101 centers dose and image quality were good, in 64 centers too high a dose was employed, in 66 centers image quality was poor, and in 38 centers dose was too high and image quality was poor. It must be stressed that the DQM program can play its role only if each center carries out its Quality Assurance activity after the methods recommended by the World Health Organization.

UI MeSH Term Description Entries
D008327 Mammography Radiographic examination of the breast. 3D-Mammography,Digital Breast Tomosynthesis,Digital Mammography,X-ray Breast Tomosynthesis,3D Mammography,3D-Mammographies,Breast Tomosyntheses, Digital,Breast Tomosyntheses, X-ray,Breast Tomosynthesis, Digital,Breast Tomosynthesis, X-ray,Digital Breast Tomosyntheses,Digital Mammographies,Mammographies,Mammographies, Digital,Mammography, Digital,X ray Breast Tomosynthesis,X-ray Breast Tomosyntheses
D011786 Quality Control A system for verifying and maintaining a desired level of quality in a product or process by careful planning, use of proper equipment, continued inspection, and corrective action as required. (Random House Unabridged Dictionary, 2d ed) Control, Quality,Controls, Quality,Quality Controls
D011829 Radiation Dosage The amount of radiation energy that is deposited in a unit mass of material, such as tissues of plants or animal. In RADIOTHERAPY, radiation dosage is expressed in gray units (Gy). In RADIOLOGIC HEALTH, the dosage is expressed by the product of absorbed dose (Gy) and quality factor (a function of linear energy transfer), and is called radiation dose equivalent in sievert units (Sv). Sievert Units,Dosage, Radiation,Gray Units,Gy Radiation,Sv Radiation Dose Equivalent,Dosages, Radiation,Radiation Dosages,Units, Gray,Units, Sievert
D015397 Program Evaluation Studies designed to assess the efficacy of programs. They may include the evaluation of cost-effectiveness, the extent to which objectives are met, or impact. Evaluation, Program,Family Planning Program Evaluation,Program Appropriateness,Program Effectiveness,Program Sustainability,Appropriateness, Program,Effectiveness, Program,Evaluations, Program,Program Evaluations,Program Sustainabilities,Sustainabilities, Program,Sustainability, Program

Related Publications

O Rimondi, and M Gambaccini, and M Marziani, and G C Candini, and P L Indovina, and A Toti, and E De Guglielmo, and A Calicchia
September 1990, Wisconsin medical journal,
O Rimondi, and M Gambaccini, and M Marziani, and G C Candini, and P L Indovina, and A Toti, and E De Guglielmo, and A Calicchia
February 1994, Administrative radiology : AR,
O Rimondi, and M Gambaccini, and M Marziani, and G C Candini, and P L Indovina, and A Toti, and E De Guglielmo, and A Calicchia
November 2012, Deutsches Arzteblatt international,
O Rimondi, and M Gambaccini, and M Marziani, and G C Candini, and P L Indovina, and A Toti, and E De Guglielmo, and A Calicchia
November 1979, Radiology,
O Rimondi, and M Gambaccini, and M Marziani, and G C Candini, and P L Indovina, and A Toti, and E De Guglielmo, and A Calicchia
September 2001, Australasian physical & engineering sciences in medicine,
O Rimondi, and M Gambaccini, and M Marziani, and G C Candini, and P L Indovina, and A Toti, and E De Guglielmo, and A Calicchia
April 1987, Radiology,
O Rimondi, and M Gambaccini, and M Marziani, and G C Candini, and P L Indovina, and A Toti, and E De Guglielmo, and A Calicchia
April 1996, Minnesota medicine,
O Rimondi, and M Gambaccini, and M Marziani, and G C Candini, and P L Indovina, and A Toti, and E De Guglielmo, and A Calicchia
December 1991, AJR. American journal of roentgenology,
O Rimondi, and M Gambaccini, and M Marziani, and G C Candini, and P L Indovina, and A Toti, and E De Guglielmo, and A Calicchia
January 1992, Journal of the American Medical Women's Association (1972),
O Rimondi, and M Gambaccini, and M Marziani, and G C Candini, and P L Indovina, and A Toti, and E De Guglielmo, and A Calicchia
December 1989, Cancer,
Copied contents to your clipboard!