Beam commissioning and measurements validating the beam model in a new TPS that converts helical tomotherapy plans to step-and-shoot IMRT plans. 2011

Kristoffer Petersson, and Crister Ceberg, and Per Engström, and Tommy Knöös
Medical Radiation Physics, Clinical Sciences, Lund University, Lund SE-221 85, Sweden. kristoffer.petersson@skane.se

OBJECTIVE A new type of treatment planning system called SHAREPLAN has been studied, which enables the transfer of treatment plans generated for helical tomotherapy delivery to plans that can be delivered on C-arm linacs. The purpose is to ensure continuous patient treatment during periods of unscheduled downtime for the TomoTherapy unit, particularly in clinics without a backup unit. The purpose of this work was to verify that the plans generated in this novel planning system are deliverable and accurate. The work consists primarily of beam commissioning, verification of the beam model, and measurements verifying that generated plans are deliverable with sufficient accuracy. METHODS The beam commissioning process involves input of general geometric properties of the modeled linac, profiles and depth dose curves for a specific photon nominal energy (6 MV), and the automated modeling of other beam properties. Some manual tuning of the beam model is required. To evaluate its accuracy, the confidence limit concept [J. Venselaar et al., "Tolerances for the accuracy of photon beam dose calculations of treatment planning systems," Radiother. Oncol. 60, 191-201 (2001)] was used, which is a method supported by ESTRO. Measurements were conducted with a 2D diode array at the commissioned linac as a final check of the beam model and to evaluate whether the generated plans were deliverable and accurate. RESULTS The comparison and evaluation of calculated data points and measured data according to the method applied confirmed the accuracy of the beam model. The profiles had a confidence limit of 1.1% and the depth dose curves had a confidence limit of 1.7%, both of which were well below the tolerance limit of 2%. Plan specific QC measurements and evaluation verified that different plans generated in the TPS were deliverable with sufficient accuracy at the commissioned linac, as none of the 160 beams for the 20 different plans evaluated had a fraction of approved data points below 90%, the local clinical approval criterion for delivery QA measurements. CONCLUSIONS This study is a validation of the new TPS as it verifies that the generated plans are deliverable at a commissioned linac with adequate accuracy. A thorough investigation of the treatment plan quality will require a separate study. The TPS is proving to be a useful and time-saving complement, especially for clinics having a single unit for helical delivery among its conventional linacs.

UI MeSH Term Description Entries
D008962 Models, Theoretical Theoretical representations that simulate the behavior or activity of systems, processes, or phenomena. They include the use of mathematical equations, computers, and other electronic equipment. Experimental Model,Experimental Models,Mathematical Model,Model, Experimental,Models (Theoretical),Models, Experimental,Models, Theoretic,Theoretical Study,Mathematical Models,Model (Theoretical),Model, Mathematical,Model, Theoretical,Models, Mathematical,Studies, Theoretical,Study, Theoretical,Theoretical Model,Theoretical Models,Theoretical Studies
D011879 Radiotherapy Dosage The total amount of radiation absorbed by tissues as a result of radiotherapy. Dosage, Radiotherapy,Dosages, Radiotherapy,Radiotherapy Dosages
D011880 Radiotherapy Planning, Computer-Assisted Computer-assisted mathematical calculations of beam angles, intensities of radiation, and duration of irradiation in radiotherapy. Computer-Assisted Radiotherapy Planning,Dosimetry Calculations, Computer-Assisted,Planning, Computer-Assisted Radiotherapy,Calculation, Computer-Assisted Dosimetry,Calculations, Computer-Assisted Dosimetry,Computer Assisted Radiotherapy Planning,Computer-Assisted Dosimetry Calculation,Computer-Assisted Dosimetry Calculations,Dosimetry Calculation, Computer-Assisted,Dosimetry Calculations, Computer Assisted,Planning, Computer Assisted Radiotherapy,Radiotherapy Planning, Computer Assisted
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D000465 Algorithms A procedure consisting of a sequence of algebraic formulas and/or logical steps to calculate or determine a given task. Algorithm
D012984 Software Sequential operating programs and data which instruct the functioning of a digital computer. Computer Programs,Computer Software,Open Source Software,Software Engineering,Software Tools,Computer Applications Software,Computer Programs and Programming,Computer Software Applications,Application, Computer Software,Applications Software, Computer,Applications Softwares, Computer,Applications, Computer Software,Computer Applications Softwares,Computer Program,Computer Software Application,Engineering, Software,Open Source Softwares,Program, Computer,Programs, Computer,Software Application, Computer,Software Applications, Computer,Software Tool,Software, Computer,Software, Computer Applications,Software, Open Source,Softwares, Computer Applications,Softwares, Open Source,Source Software, Open,Source Softwares, Open,Tool, Software,Tools, Software
D050397 Radiotherapy, Intensity-Modulated CONFORMAL RADIOTHERAPY that combines several intensity-modulated beams to provide improved dose homogeneity and highly conformal dose distributions. Helical Tomotherapy,Intensity-Modulated Arc Therapy,Volumetric-Modulated Arc Therapy,Arc Therapies, Intensity-Modulated,Arc Therapies, Volumetric-Modulated,Arc Therapy, Intensity-Modulated,Arc Therapy, Volumetric-Modulated,Helical Tomotherapies,Intensity Modulated Arc Therapy,Intensity-Modulated Arc Therapies,Intensity-Modulated Radiotherapies,Intensity-Modulated Radiotherapy,Radiotherapies, Intensity-Modulated,Radiotherapy, Intensity Modulated,Therapies, Intensity-Modulated Arc,Therapies, Volumetric-Modulated Arc,Therapy, Intensity-Modulated Arc,Therapy, Volumetric-Modulated Arc,Tomotherapies, Helical,Tomotherapy, Helical,Volumetric Modulated Arc Therapy,Volumetric-Modulated Arc Therapies

Related Publications

Kristoffer Petersson, and Crister Ceberg, and Per Engström, and Tommy Knöös
August 2005, International journal of radiation oncology, biology, physics,
Kristoffer Petersson, and Crister Ceberg, and Per Engström, and Tommy Knöös
January 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB),
Kristoffer Petersson, and Crister Ceberg, and Per Engström, and Tommy Knöös
June 2010, Medical physics,
Kristoffer Petersson, and Crister Ceberg, and Per Engström, and Tommy Knöös
July 2012, Medical physics,
Kristoffer Petersson, and Crister Ceberg, and Per Engström, and Tommy Knöös
July 2016, Nihon Hoshasen Gijutsu Gakkai zasshi,
Kristoffer Petersson, and Crister Ceberg, and Per Engström, and Tommy Knöös
May 2013, Journal of applied clinical medical physics,
Kristoffer Petersson, and Crister Ceberg, and Per Engström, and Tommy Knöös
November 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB),
Kristoffer Petersson, and Crister Ceberg, and Per Engström, and Tommy Knöös
May 2015, Journal of radiation research,
Kristoffer Petersson, and Crister Ceberg, and Per Engström, and Tommy Knöös
June 2006, International journal of radiation oncology, biology, physics,
Kristoffer Petersson, and Crister Ceberg, and Per Engström, and Tommy Knöös
March 2015, Journal of radiation research,
Copied contents to your clipboard!