Comparative study of two laser Doppler blood flowmeters. 1990

N J Barnett, and G Dougherty, and S J Pettinger
School of Engineering, Oxford Polytechnic, Headington, UK.

A new infrared laser Doppler blood flow instrument (Moor MBF3D) was evaluated using an in vitro model allowing measurements over a range of flow velocities and concentrations. The responses correlated well (r = 0.96, p less than 0.01) with those obtained simultaneously using a Perimed PF3 laser Doppler instrument. The different processing bandwidths of the instruments were investigated and the wideband mode of operation is recommended for flow measurements where there may be fast moving red blood cells (rbcs). The infrared instrument is capable of dual-channel operation, and the two channels are shown to respond almost identically for similar changes in blood flow through the in vitro model (r = 0.999, p less than 0.01). The main advantage of the dual-channel instrument is that continuous measurements may be made simultaneously at two different skin sites allowing dynamic flow responses to be compared.

UI MeSH Term Description Entries
D004867 Equipment Design Methods and patterns of fabricating machines and related hardware. Design, Equipment,Device Design,Medical Device Design,Design, Medical Device,Designs, Medical Device,Device Design, Medical,Device Designs, Medical,Medical Device Designs,Design, Device,Designs, Device,Designs, Equipment,Device Designs,Equipment Designs
D012212 Rheology The study of the deformation and flow of matter, usually liquids or fluids, and of the plastic flow of solids. The concept covers consistency, dilatancy, liquefaction, resistance to flow, shearing, thixotrophy, and VISCOSITY. Flowmetry,Velocimetry,Velocimetries

Related Publications

N J Barnett, and G Dougherty, and S J Pettinger
January 1983, Microsurgery,
N J Barnett, and G Dougherty, and S J Pettinger
July 1986, Medical & biological engineering & computing,
N J Barnett, and G Dougherty, and S J Pettinger
July 1984, Medical & biological engineering & computing,
N J Barnett, and G Dougherty, and S J Pettinger
January 1992, Journal of medical engineering & technology,
N J Barnett, and G Dougherty, and S J Pettinger
March 1987, Medical & biological engineering & computing,
N J Barnett, and G Dougherty, and S J Pettinger
June 1995, Pediatric emergency care,
N J Barnett, and G Dougherty, and S J Pettinger
January 1965, The American journal of medical electronics,
N J Barnett, and G Dougherty, and S J Pettinger
January 1979, Biotelemetry and patient monitoring,
N J Barnett, and G Dougherty, and S J Pettinger
January 1967, The New England journal of medicine,
N J Barnett, and G Dougherty, and S J Pettinger
September 1984, Iyo denshi to seitai kogaku. Japanese journal of medical electronics and biological engineering,
Copied contents to your clipboard!