Results of the recalibration of creatinine measurement with the modular Beckman Coulter Jaffe creatinine method. 2011

Catherine Fillée, and Guido Vranken, and Myriam Othmane, and Marianne Philippe, and Jean-Michel Allaeys, and Anne Courbe, and Jean-Luc Ruelle, and Rudy Peeters
Université Catholique de Louvain, Cliniques Universitaires Saint-Luc, Brussels, Belgium.

BACKGROUND Serum creatinine is important for detecting the beginning of a decline in kidney function. The Beckman Coulter Jaffe reagents for measuring creatinine have been standardized to the internationally accepted reference method: isotope dilution mass spectrometry (IDMS). The impact of this recalibration on the Beckman Coulter modular or cup (stat) Jaffe method is studied. METHODS Recalibrated Jaffe (calibrator set points IDMS traceable) and classic National Institute of Standards and Technology (NIST) creatinine methods (traceable to NIST 914a) were compared with an enzymatic method (Sentinel, traceable to NIST SRM 967). All measurements were performed on Synchron DxC 800 systems. Imprecision of the routine methods was studied using the Clinical and Laboratory Standards Institute (CLSI) protocols and laboratory quality survey. Thirteen plasma pools, with concentrations < 354 mmol/L, were analyzed with a gas chromatography isotope dilution mass spectrometry (GC-IDMS) method and routine methods. Total error of 8.2% based on biological variability, set on the GC-IDMS values and acceptance criteria (bias < 5%, imprecision < 8% at concentrations ≥ 88.4 mmol/L and a maximum 10% increase in the relative error of estimated glomerular filtration rate (eGFR) of the National Kidney Disease Educational Program (NKDEP) were used for evaluating analytical performance of the routine methods studied. RESULTS After recalibration of the Jaffe method, median bias (mmol/L) decreased from 12.4 (95% CI: 9.1-21.2) to 7.3 (95% CI: 1.5-10.5). Imprecision of the Jaffe method is in agreement with the claim of the manufacturer, namely < 9 mmol/L or < 3% below or above 292 mmol/L. Below creatinine of 88.4 mmol/L, imprecision of the recalibrated Jaffe and enzymatic methods is between 4.1% and 6.9%, and 5.0% and 7.1%, respectively, and significantly different (p = 0.02 for both the Jaffe and enzymatic methods) from the goal, based on biological variability, of 2.7%. For the adult pools, all recalibrated Jaffe and enzymatic results fall within the total error of 8.2%. In the pediatric samples, one-third of the recalibrated Jaffe and three of the six enzymatic results fall within this total error goal. CONCLUSIONS Recalibration significantly reduced bias of the Jaffe method. For pediatric samples, recalibrated Jaffe results do not comply with either the imprecision goal or the total error based on biological variability. Adult recalibrated Jaffe results are in compliance with the goals based on biological variability and with the acceptance criteria from the NKDEP.

UI MeSH Term Description Entries
D007201 Indicator Dilution Techniques Methods for assessing flow through a system by injection of a known quantity of an indicator, such as a dye, radionuclide, or chilled liquid, into the system and monitoring its concentration over time at a specific point in the system. (From Dorland, 28th ed) Dilution Techniques,Dilution Technics,Indicator Dilution Technics,Dilution Technic,Dilution Technic, Indicator,Dilution Technics, Indicator,Dilution Technique,Dilution Technique, Indicator,Dilution Techniques, Indicator,Indicator Dilution Technic,Indicator Dilution Technique,Technic, Dilution,Technic, Indicator Dilution,Technics, Dilution,Technics, Indicator Dilution,Technique, Dilution,Technique, Indicator Dilution,Techniques, Dilution,Techniques, Indicator Dilution
D008401 Gas Chromatography-Mass Spectrometry A microanalytical technique combining mass spectrometry and gas chromatography for the qualitative as well as quantitative determinations of compounds. Chromatography, Gas-Liquid-Mass Spectrometry,Chromatography, Gas-Mass Spectrometry,GCMS,Spectrometry, Mass-Gas Chromatography,Spectrum Analysis, Mass-Gas Chromatography,Gas-Liquid Chromatography-Mass Spectrometry,Mass Spectrometry-Gas Chromatography,Chromatography, Gas Liquid Mass Spectrometry,Chromatography, Gas Mass Spectrometry,Chromatography, Mass Spectrometry-Gas,Chromatography-Mass Spectrometry, Gas,Chromatography-Mass Spectrometry, Gas-Liquid,Gas Chromatography Mass Spectrometry,Gas Liquid Chromatography Mass Spectrometry,Mass Spectrometry Gas Chromatography,Spectrometries, Mass-Gas Chromatography,Spectrometry, Gas Chromatography-Mass,Spectrometry, Gas-Liquid Chromatography-Mass,Spectrometry, Mass Gas Chromatography,Spectrometry-Gas Chromatography, Mass,Spectrum Analysis, Mass Gas Chromatography
D011933 Reagent Kits, Diagnostic Commercially prepared reagent sets, with accessory devices, containing all of the major components and literature necessary to perform one or more designated diagnostic tests or procedures. They may be for laboratory or personal use. Diagnostic Reagent Kits,Diagnostic Reagents and Test Kits,Diagnostic Test Kits,In Vitro Diagnostic Device,In Vitro Diagnostic Devices,In Vitro Diagnostic Medical Device,In Vitro Diagnostic Medical Devices,Kits, Diagnostic Reagent,Diagnostic Reagent Kit,Diagnostic Test Kit,Kit, Diagnostic Reagent,Kit, Diagnostic Test,Kits, Diagnostic Test,Reagent Kit, Diagnostic,Test Kit, Diagnostic,Test Kits, Diagnostic
D012044 Regression Analysis Procedures for finding the mathematical function which best describes the relationship between a dependent variable and one or more independent variables. In linear regression (see LINEAR MODELS) the relationship is constrained to be a straight line and LEAST-SQUARES ANALYSIS is used to determine the best fit. In logistic regression (see LOGISTIC MODELS) the dependent variable is qualitative rather than continuously variable and LIKELIHOOD FUNCTIONS are used to find the best relationship. In multiple regression, the dependent variable is considered to depend on more than a single independent variable. Regression Diagnostics,Statistical Regression,Analysis, Regression,Analyses, Regression,Diagnostics, Regression,Regression Analyses,Regression, Statistical,Regressions, Statistical,Statistical Regressions
D001774 Blood Chemical Analysis An examination of chemicals in the blood. Analysis, Blood Chemical,Chemical Analysis, Blood,Analyses, Blood Chemical,Blood Chemical Analyses,Chemical Analyses, Blood
D002138 Calibration Determination, by measurement or comparison with a standard, of the correct value of each scale reading on a meter or other measuring instrument; or determination of the settings of a control device that correspond to particular values of voltage, current, frequency or other output. Calibrations
D003404 Creatinine Creatinine Sulfate Salt,Krebiozen,Salt, Creatinine Sulfate,Sulfate Salt, Creatinine
D005919 Glomerular Filtration Rate The volume of water filtered out of plasma through glomerular capillary walls into Bowman's capsules per unit of time. It is considered to be equivalent to INULIN clearance. Filtration Rate, Glomerular,Filtration Rates, Glomerular,Glomerular Filtration Rates,Rate, Glomerular Filtration,Rates, Glomerular Filtration
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D000328 Adult A person having attained full growth or maturity. Adults are of 19 through 44 years of age. For a person between 19 and 24 years of age, YOUNG ADULT is available. Adults

Related Publications

Catherine Fillée, and Guido Vranken, and Myriam Othmane, and Marianne Philippe, and Jean-Michel Allaeys, and Anne Courbe, and Jean-Luc Ruelle, and Rudy Peeters
November 2023, Annals of clinical biochemistry,
Catherine Fillée, and Guido Vranken, and Myriam Othmane, and Marianne Philippe, and Jean-Michel Allaeys, and Anne Courbe, and Jean-Luc Ruelle, and Rudy Peeters
February 1979, Clinical chemistry,
Catherine Fillée, and Guido Vranken, and Myriam Othmane, and Marianne Philippe, and Jean-Michel Allaeys, and Anne Courbe, and Jean-Luc Ruelle, and Rudy Peeters
September 1968, Rinsho byori. The Japanese journal of clinical pathology,
Catherine Fillée, and Guido Vranken, and Myriam Othmane, and Marianne Philippe, and Jean-Michel Allaeys, and Anne Courbe, and Jean-Luc Ruelle, and Rudy Peeters
January 2008, Annals of clinical biochemistry,
Catherine Fillée, and Guido Vranken, and Myriam Othmane, and Marianne Philippe, and Jean-Michel Allaeys, and Anne Courbe, and Jean-Luc Ruelle, and Rudy Peeters
June 2021, Critical care and resuscitation : journal of the Australasian Academy of Critical Care Medicine,
Catherine Fillée, and Guido Vranken, and Myriam Othmane, and Marianne Philippe, and Jean-Michel Allaeys, and Anne Courbe, and Jean-Luc Ruelle, and Rudy Peeters
February 2012, Clinical chemistry,
Catherine Fillée, and Guido Vranken, and Myriam Othmane, and Marianne Philippe, and Jean-Michel Allaeys, and Anne Courbe, and Jean-Luc Ruelle, and Rudy Peeters
January 2015, PloS one,
Catherine Fillée, and Guido Vranken, and Myriam Othmane, and Marianne Philippe, and Jean-Michel Allaeys, and Anne Courbe, and Jean-Luc Ruelle, and Rudy Peeters
January 2015, PloS one,
Catherine Fillée, and Guido Vranken, and Myriam Othmane, and Marianne Philippe, and Jean-Michel Allaeys, and Anne Courbe, and Jean-Luc Ruelle, and Rudy Peeters
February 1979, Clinical biochemistry,
Catherine Fillée, and Guido Vranken, and Myriam Othmane, and Marianne Philippe, and Jean-Michel Allaeys, and Anne Courbe, and Jean-Luc Ruelle, and Rudy Peeters
May 2021, Annals of clinical biochemistry,
Copied contents to your clipboard!