Performance evaluation of Siemens ADVIA Centaur and Roche MODULAR Analytics E170 Total 25-OH Vitamin D assays. 2012

Yu Chen, and Lois Kinney, and Andrea Božović, and Hilary Smith, and Heather Tarr, and Eleftherios P Diamandis, and Adrien LeBlanc
Division of Clinical Biochemistry, Department of Laboratory Medicine, Dr. Everett Chalmers Regional Hospital, Horizon Health Network, Fredericton, New Brunswick, Canada E3B 5N5. yu.chen@horizonNB.ca

OBJECTIVE To evaluate the newly developed Roche MODULAR Analytics E170 Total Vitamin D and the Siemens ADVIA Centaur Vitamin D Total assays. METHODS Assays were evaluated using the Clinical and Laboratory Standards Institute protocols. Split patient samples were compared with LC-MS/MS and DiaSorin LIAISON assays (n=79 including 15 specimens with detectable endogenous 25-OH vitamin D(2)). Assay accuracy was also evaluated using the Vitamin D External Quality Assessment Scheme (DEQAS) samples. RESULTS The ADVIA Centaur and E170 assays demonstrated maximum total CVs of 14.1% and 5.9%, respectively. Both showed excellent linearity (R(2)>0.99). The ADVIA Centaur assay demonstrated interference with bilirubin at 800 μmol/L, hemolysis at 1.25 g/L, and triglycerides at 2.8 mmol/L. Compared to LC-MS/MS, the ADVIA Centaur assay demonstrated a R(2) value of 0.893, average bias of -8.8%; the E170 assay an R(2) value of 0.872, average bias of 14.3% with underestimation of 25-OH vitamin D(2). Compared to the LIAISON assay, the ADVIA Centaur assay demonstrated an R(2) value of 0.781, average bias of -17.3%; the E170 assay an R(2) value of 0.823, average bias of 11.4%. The ADVIA Centaur and E170 assays demonstrated a biases of <20% in 10/10 and 8/10 DEQAS samples, respectively. CONCLUSIONS The ADVIA Centaur and E170 vitamin D assays demonstrated acceptable linearity, imprecision, and accuracy. The E170 assay demonstrated consistent underestimation of 25-OH vitamin D(2) levels. Compared with LC-MS/MS, the ADVIA Centaur assay demonstrated a higher R(2) value and a smaller average bias than the E170 assay.

UI MeSH Term Description Entries
D012015 Reference Standards A basis of value established for the measure of quantity, weight, extent or quality, e.g. weight standards, standard solutions, methods, techniques, and procedures used in diagnosis and therapy. Standard Preparations,Standards, Reference,Preparations, Standard,Standardization,Standards,Preparation, Standard,Reference Standard,Standard Preparation,Standard, Reference
D001774 Blood Chemical Analysis An examination of chemicals in the blood. Analysis, Blood Chemical,Chemical Analysis, Blood,Analyses, Blood Chemical,Blood Chemical Analyses,Chemical Analyses, Blood
D002762 Cholecalciferol Derivative of 7-dehydroxycholesterol formed by ULTRAVIOLET RAYS breaking of the C9-C10 bond. It differs from ERGOCALCIFEROL in having a single bond between C22 and C23 and lacking a methyl group at C24. Vitamin D 3,(3 beta,5Z,7E)-9,10-Secocholesta-5,7,10(19)-trien-3-ol,Calciol,Cholecalciferols,Vitamin D3
D004872 Ergocalciferols Derivatives of ERGOSTEROL formed by ULTRAVIOLET RAYS breaking of the C9-C10 bond. They differ from CHOLECALCIFEROL in having a double bond between C22 and C23 and a methyl group at C24. Calciferols,Vitamin D 2,Ergocalciferol,Vitamin D2,D2, Vitamin
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D016014 Linear Models Statistical models in which the value of a parameter for a given value of a factor is assumed to be equal to a + bx, where a and b are constants. The models predict a linear regression. Linear Regression,Log-Linear Models,Models, Linear,Linear Model,Linear Regressions,Log Linear Models,Log-Linear Model,Model, Linear,Model, Log-Linear,Models, Log-Linear,Regression, Linear,Regressions, Linear
D053719 Tandem Mass Spectrometry A mass spectrometry technique using two (MS/MS) or more mass analyzers. With two in tandem, the precursor ions are mass-selected by a first mass analyzer, and focused into a collision region where they are then fragmented into product ions which are then characterized by a second mass analyzer. A variety of techniques are used to separate the compounds, ionize them, and introduce them to the first mass analyzer. For example, for in GC-MS/MS, GAS CHROMATOGRAPHY-MASS SPECTROMETRY is involved in separating relatively small compounds by GAS CHROMATOGRAPHY prior to injecting them into an ionization chamber for the mass selection. Mass Spectrometry-Mass Spectrometry,Mass Spectrometry Mass Spectrometry,Mass Spectrometry, Tandem
D057230 Limit of Detection Concentration or quantity that is derived from the smallest measure that can be detected with reasonable certainty for a given analytical procedure. Limits of Detection,Detection Limit,Detection Limits
D059629 Signal-To-Noise Ratio The comparison of the quantity of meaningful data to the irrelevant or incorrect data. Ratio, Signal-To-Noise,Ratios, Signal-To-Noise,Signal To Noise Ratio,Signal-To-Noise Ratios

Related Publications

Yu Chen, and Lois Kinney, and Andrea Božović, and Hilary Smith, and Heather Tarr, and Eleftherios P Diamandis, and Adrien LeBlanc
December 2008, Clinical chemistry,
Yu Chen, and Lois Kinney, and Andrea Božović, and Hilary Smith, and Heather Tarr, and Eleftherios P Diamandis, and Adrien LeBlanc
January 2012, Clinical laboratory,
Yu Chen, and Lois Kinney, and Andrea Božović, and Hilary Smith, and Heather Tarr, and Eleftherios P Diamandis, and Adrien LeBlanc
January 2004, Clinical chemistry and laboratory medicine,
Yu Chen, and Lois Kinney, and Andrea Božović, and Hilary Smith, and Heather Tarr, and Eleftherios P Diamandis, and Adrien LeBlanc
November 2012, Clinical chemistry and laboratory medicine,
Yu Chen, and Lois Kinney, and Andrea Božović, and Hilary Smith, and Heather Tarr, and Eleftherios P Diamandis, and Adrien LeBlanc
October 2008, Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology,
Yu Chen, and Lois Kinney, and Andrea Božović, and Hilary Smith, and Heather Tarr, and Eleftherios P Diamandis, and Adrien LeBlanc
May 2005, The Journal of clinical endocrinology and metabolism,
Yu Chen, and Lois Kinney, and Andrea Božović, and Hilary Smith, and Heather Tarr, and Eleftherios P Diamandis, and Adrien LeBlanc
May 2005, The Journal of clinical endocrinology and metabolism,
Yu Chen, and Lois Kinney, and Andrea Božović, and Hilary Smith, and Heather Tarr, and Eleftherios P Diamandis, and Adrien LeBlanc
January 2014, International journal of endocrinology,
Yu Chen, and Lois Kinney, and Andrea Božović, and Hilary Smith, and Heather Tarr, and Eleftherios P Diamandis, and Adrien LeBlanc
January 2014, International journal of endocrinology,
Yu Chen, and Lois Kinney, and Andrea Božović, and Hilary Smith, and Heather Tarr, and Eleftherios P Diamandis, and Adrien LeBlanc
November 2005, The Journal of clinical endocrinology and metabolism,
Copied contents to your clipboard!