Doppler echocardiographic characteristics of normal and dysfunctioning prosthetic valves in the tricuspid and mitral position. 1990

M Pye, and N Weerasana, and W H Bain, and I Hutton, and S M Cobbe
Department of Medical Cardiology, Royal Infirmary, Glasgow.

The Doppler echocardiographic characteristics of 70 prosthetic valves in 35 patients are reported. Twenty nine patients had a Björk-Shiley prosthesis in both mitral and tricuspid positions and six had Carpentier-Edwards valves in both sites. Five of the patients had abnormal tricuspid prostheses on the basis of clinical and echocardiographic criteria. Pulsed wave Doppler echocardiography was used in all examinations. The pressure half times for the normal tricuspid prosthetic valves, 105 (40) ms (Björk-Shiley) and 97 (53) ms (Carpentier-Edwards), were significantly longer than those of normal mitral prosthetic valves, 75 (18) ms (Björk-Shiley) and 83 (15) ms (Carpentier-Edwards). The range of pressure half times for the abnormal tricuspid valves (237-530 ms) was distinct from that of the apparently normal tricuspid prosthetic valves (38-197 ms). There was an increase in the peak velocity of the obstructed tricuspid prosthetic valves (1.69 (0.12) m/s) in comparison with normal prostheses (1.06 (0.26) m/s). The normal range of pressure half times for the Björk-Shiley and Carpentier-Edwards valves in the mitral position is not applicable to the same valves in the tricuspid position. The valve appears to function well with very long pressure half times but a pressure half time of greater than 200 ms coupled with a peak velocity of greater than 1.60 ms without significant valve regurgitation indicates tricuspid obstruction of the tricuspid prosthetic valve.

UI MeSH Term Description Entries
D008297 Male Males
D008875 Middle Aged An adult aged 45 - 64 years. Middle Age
D008943 Mitral Valve The valve between the left atrium and left ventricle of the heart. Bicuspid Valve,Bicuspid Valves,Mitral Valves,Valve, Bicuspid,Valve, Mitral,Valves, Bicuspid,Valves, Mitral
D011475 Prosthesis Failure Malfunction of implantation shunts, valves, etc., and prosthesis loosening, migration, and breaking. Prosthesis Loosening,Prosthesis Durability,Prosthesis Migration,Prosthesis Survival,Durabilities, Prosthesis,Durability, Prosthesis,Failure, Prosthesis,Failures, Prosthesis,Loosening, Prosthesis,Loosenings, Prosthesis,Migration, Prosthesis,Migrations, Prosthesis,Prosthesis Durabilities,Prosthesis Failures,Prosthesis Loosenings,Prosthesis Migrations,Prosthesis Survivals,Survival, Prosthesis,Survivals, Prosthesis
D001783 Blood Flow Velocity A value equal to the total volume flow divided by the cross-sectional area of the vascular bed. Blood Flow Velocities,Flow Velocities, Blood,Flow Velocity, Blood,Velocities, Blood Flow,Velocity, Blood Flow
D001794 Blood Pressure PRESSURE of the BLOOD on the ARTERIES and other BLOOD VESSELS. Systolic Pressure,Diastolic Pressure,Pulse Pressure,Pressure, Blood,Pressure, Diastolic,Pressure, Pulse,Pressure, Systolic,Pressures, Systolic
D005260 Female Females
D006350 Heart Valve Prosthesis A device that substitutes for a heart valve. It may be composed of biological material (BIOPROSTHESIS) and/or synthetic material. Prosthesis, Heart Valve,Cardiac Valve Prosthesis,Cardiac Valve Prostheses,Heart Valve Prostheses,Prostheses, Cardiac Valve,Prostheses, Heart Valve,Prosthesis, Cardiac Valve,Valve Prostheses, Cardiac,Valve Prostheses, Heart,Valve Prosthesis, Cardiac,Valve Prosthesis, Heart
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D000328 Adult A person having attained full growth or maturity. Adults are of 19 through 44 years of age. For a person between 19 and 24 years of age, YOUNG ADULT is available. Adults

Related Publications

M Pye, and N Weerasana, and W H Bain, and I Hutton, and S M Cobbe
September 2003, Asian cardiovascular & thoracic annals,
M Pye, and N Weerasana, and W H Bain, and I Hutton, and S M Cobbe
October 1987, Journal of the American College of Cardiology,
M Pye, and N Weerasana, and W H Bain, and I Hutton, and S M Cobbe
September 1988, International journal of cardiology,
M Pye, and N Weerasana, and W H Bain, and I Hutton, and S M Cobbe
January 1990, International journal of cardiac imaging,
M Pye, and N Weerasana, and W H Bain, and I Hutton, and S M Cobbe
June 1996, Journal of the American College of Cardiology,
M Pye, and N Weerasana, and W H Bain, and I Hutton, and S M Cobbe
May 1993, The Journal of heart valve disease,
M Pye, and N Weerasana, and W H Bain, and I Hutton, and S M Cobbe
April 1991, American heart journal,
M Pye, and N Weerasana, and W H Bain, and I Hutton, and S M Cobbe
January 1994, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography,
M Pye, and N Weerasana, and W H Bain, and I Hutton, and S M Cobbe
June 2001, The Journal of cardiovascular surgery,
M Pye, and N Weerasana, and W H Bain, and I Hutton, and S M Cobbe
January 1991, Acta cardiologica,
Copied contents to your clipboard!