Spencer-Attix water/medium stopping-power ratios for the dosimetry of proton pencil beams. 2013

C Gomà, and P Andreo, and J Sempau
Centre for Proton Therapy, Paul Scherrer Institute, Villigen PSI, Switzerland. carles.goma@psi.ch

This paper uses Monte Carlo simulations to calculate the Spencer-Attix water/medium stopping-power ratios (sw, med) for the dosimetry of scanned proton pencil beams. It includes proton energies from 30 to 350 MeV and typical detection materials such as air (ionization chambers), radiochromic film, gadolinium oxysulfide (scintillating screens), silicon and lithium fluoride. Track-ends and particles heavier than protons were found to have a negligible effect on the water/air stopping-power ratios (sw, air), whereas the mean excitation energy values were found to carry the largest source of uncertainty. The initial energy spread of the beam was found to have a minor influence on the sw, air values in depth. The water/medium stopping-power ratios as a function of depth in water were found to be quite constant for air and radiochromic film-within 2.5%. Also, the sw, med values were found to have no clinically relevant dependence on the radial distance-except for the case of gadolinium oxysulfide and proton radiography beams. In conclusion, the most suitable detection materials for depth-dose measurements in water were found to be air and radiochromic film active layer, although a small correction is still needed to compensate for the different sw, med values between the plateau and the Bragg peak region. Also, all the detection materials studied in this work-except for gadolinium oxysulfide-were found to be suitable for lateral dose profiles and field-specific dose distribution measurements in water.

UI MeSH Term Description Entries
D011874 Radiometry The measurement of radiation by photography, as in x-ray film and film badge, by Geiger-Mueller tube, and by SCINTILLATION COUNTING. Geiger-Mueller Counters,Nuclear Track Detection,Radiation Dosimetry,Dosimetry, Radiation,Geiger Counter,Geiger-Mueller Counter Tube,Geiger-Mueller Probe,Geiger-Mueller Tube,Radiation Counter,Counter Tube, Geiger-Mueller,Counter Tubes, Geiger-Mueller,Counter, Geiger,Counter, Radiation,Counters, Geiger,Counters, Geiger-Mueller,Counters, Radiation,Detection, Nuclear Track,Dosimetries, Radiation,Geiger Counters,Geiger Mueller Counter Tube,Geiger Mueller Counters,Geiger Mueller Probe,Geiger Mueller Tube,Geiger-Mueller Counter Tubes,Geiger-Mueller Probes,Geiger-Mueller Tubes,Probe, Geiger-Mueller,Probes, Geiger-Mueller,Radiation Counters,Radiation Dosimetries,Tube, Geiger-Mueller,Tube, Geiger-Mueller Counter,Tubes, Geiger-Mueller,Tubes, Geiger-Mueller Counter
D011879 Radiotherapy Dosage The total amount of radiation absorbed by tissues as a result of radiotherapy. Dosage, Radiotherapy,Dosages, Radiotherapy,Radiotherapy Dosages
D000388 Air The mixture of gases present in the earth's atmosphere consisting of oxygen, nitrogen, carbon dioxide, and small amounts of other gases.
D014867 Water A clear, odorless, tasteless liquid that is essential for most animal and plant life and is an excellent solvent for many substances. The chemical formula is hydrogen oxide (H2O). (McGraw-Hill Dictionary of Scientific and Technical Terms, 4th ed) Hydrogen Oxide
D061766 Proton Therapy The use of an external beam of PROTONS as radiotherapy. Proton Beam Radiation Therapy,Proton Beam Therapy,Proton Beam Therapies,Proton Therapies,Therapies, Proton,Therapies, Proton Beam,Therapy, Proton,Therapy, Proton Beam

Related Publications

C Gomà, and P Andreo, and J Sempau
February 2016, Radiation protection dosimetry,
C Gomà, and P Andreo, and J Sempau
January 1997, Physics in medicine and biology,
C Gomà, and P Andreo, and J Sempau
January 1978, Physics in medicine and biology,
C Gomà, and P Andreo, and J Sempau
May 2006, Physics in medicine and biology,
C Gomà, and P Andreo, and J Sempau
August 2011, Acta oncologica (Stockholm, Sweden),
C Gomà, and P Andreo, and J Sempau
November 2007, Physics in medicine and biology,
C Gomà, and P Andreo, and J Sempau
April 2011, Physics in medicine and biology,
C Gomà, and P Andreo, and J Sempau
May 1995, Medical physics,
C Gomà, and P Andreo, and J Sempau
March 2010, Medical physics,
C Gomà, and P Andreo, and J Sempau
March 2016, Zeitschrift fur medizinische Physik,
Copied contents to your clipboard!