Apparent pacemaker malfunction due to ventricular blanking. 1987

N D Berman, and C A George, and G B Duxbury

Electrocardiographic interpretation in patients with dual chamber pacemakers requires an intimate understanding of their manner of functioning. This report describes a patient whose pacemaker was suspected of malfunction because of failure to sense premature beats. This sense-failure resulted from the fortuitous concurrence of the sensed portion of the premature beat with the ventricular blanking period and was resolved by reducing the low rate limit of the pacemaker. Three principles must be recognized in interpreting suspected failure of sensing: 1. The channel of origin of the pacing artifact must be identified. 2. The surface QRS morphology does not necessarily correspond to the signal detected by the pacemaker. 3. Failure to sense does not necessarily equal pacemaker malfunction.

UI MeSH Term Description Entries
D008297 Male Males
D008875 Middle Aged An adult aged 45 - 64 years. Middle Age
D010138 Pacemaker, Artificial A device designed to stimulate, by electric impulses, contraction of the heart muscles. It may be temporary (external) or permanent (internal or internal-external). Cardiac Pacemaker, Artificial,Artificial Cardiac Pacemaker,Artificial Cardiac Pacemakers,Artificial Pacemaker,Artificial Pacemakers,Cardiac Pacemakers, Artificial,Pacemaker, Artificial Cardiac,Pacemakers, Artificial,Pacemakers, Artificial Cardiac
D004562 Electrocardiography Recording of the moment-to-moment electromotive forces of the HEART as projected onto various sites on the body's surface, delineated as a scalar function of time. The recording is monitored by a tracing on slow moving chart paper or by observing it on a cardioscope, which is a CATHODE RAY TUBE DISPLAY. 12-Lead ECG,12-Lead EKG,12-Lead Electrocardiography,Cardiography,ECG,EKG,Electrocardiogram,Electrocardiograph,12 Lead ECG,12 Lead EKG,12 Lead Electrocardiography,12-Lead ECGs,12-Lead EKGs,12-Lead Electrocardiographies,Cardiographies,ECG, 12-Lead,EKG, 12-Lead,Electrocardiograms,Electrocardiographies, 12-Lead,Electrocardiographs,Electrocardiography, 12-Lead
D004868 Equipment Failure Failure of equipment to perform to standard. The failure may be due to defects or improper use. Defects, Equipment,Device Failure,Failure, Equipment,Malfunction, Equipment,Medical Device Failure,Misuse, Equipment,Device Failure, Medical,Device Failures, Medical,Failure, Medical Device,Failures, Medical Device,Defect, Equipment,Device Failures,Equipment Defect,Equipment Defects,Equipment Failures,Equipment Malfunction,Equipment Malfunctions,Equipment Misuse,Equipment Misuses,Failure, Device,Failures, Device,Failures, Equipment,Malfunctions, Equipment,Misuses, Equipment
D005117 Cardiac Complexes, Premature A group of cardiac arrhythmias in which the cardiac contractions are not initiated at the SINOATRIAL NODE. They include both atrial and ventricular premature beats, and are also known as extra or ectopic heartbeats. Their frequency is increased in heart diseases. Ectopic Heartbeats,Extrasystole,Premature Beats,Premature Cardiac Complexes,Cardiac Complex, Premature,Extrasystoles,Premature Cardiac Complex,Beat, Premature,Beats, Premature,Complexes, Premature Cardiac,Ectopic Heartbeat,Heartbeat, Ectopic,Heartbeats, Ectopic,Premature Beat,Premature Cardiac Complices
D006352 Heart Ventricles The lower right and left chambers of the heart. The right ventricle pumps venous BLOOD into the LUNGS and the left ventricle pumps oxygenated blood into the systemic arterial circulation. Cardiac Ventricle,Cardiac Ventricles,Heart Ventricle,Left Ventricle,Right Ventricle,Left Ventricles,Right Ventricles,Ventricle, Cardiac,Ventricle, Heart,Ventricle, Left,Ventricle, Right,Ventricles, Cardiac,Ventricles, Heart,Ventricles, Left,Ventricles, Right
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man

Related Publications

N D Berman, and C A George, and G B Duxbury
February 1990, International journal of cardiology,
N D Berman, and C A George, and G B Duxbury
July 1986, Pacing and clinical electrophysiology : PACE,
N D Berman, and C A George, and G B Duxbury
January 2010, The Canadian journal of cardiology,
N D Berman, and C A George, and G B Duxbury
January 2021, Indian pacing and electrophysiology journal,
N D Berman, and C A George, and G B Duxbury
October 1992, Angiology,
N D Berman, and C A George, and G B Duxbury
February 2008, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology,
N D Berman, and C A George, and G B Duxbury
January 2009, Journal of cardiovascular electrophysiology,
N D Berman, and C A George, and G B Duxbury
November 1976, Lille medical : journal de la Faculte de medecine et de pharmacie de l'Universite de Lille,
N D Berman, and C A George, and G B Duxbury
July 2005, Nursing,
Copied contents to your clipboard!