An investigation of the radiation doses to the lower legs and feet of staff undertaking interventional procedures. 2014

R E R Artschan, and D S Brettle, and K Chase, and A Fender, and P G Howells, and S Buchan
1 Department of Medical Physics and Engineering, Leeds Teaching Hospitals NHS Trust, St James's University Hospital Leeds, Leeds, UK.

OBJECTIVE Occupational radiation doses from fluoroscopic procedures are some of the highest doses of exposure amongst medical staff using radiography. Protective equipment and dose monitoring are used to minimize and control the risk from these occupational doses. Other studies have considered the effectiveness of this protection, but this study further considers whether protection is adequate for the lower leg and foot and the extent to which these doses can be reduced. METHODS Scatter air kerma profiles at toe level were measured with an ionization chamber. Thermoluminescent dosemeters and lower extremity phantoms were used to estimate the dose variation with the height of patient couch. A 7-week period of in situ toe dose monitoring of four radiologists was also undertaken. RESULTS The use of protective curtains effectively reduced the exposure to most of the lower extremities. Toe doses were found to be high and increased with increase in couch height. In situ monitoring indicated annual toe doses of 110 mSv for two of the four radiologists monitored. CONCLUSIONS Protective curtains should be used, but they might have limitations with respect to toe doses. Annual toe doses approaching the classification threshold of 150 mSv were measured for two radiologists. Caution should be exercised when there is a gap below curtains and, when possible, staff should step back from the couch. Lower legs and toes should be included in local radiation protection programmes. CONCLUSIONS Toe doses in interventional radiology may be higher than expected and may have to be included in radiation protection programmes.

UI MeSH Term Description Entries
D011829 Radiation Dosage The amount of radiation energy that is deposited in a unit mass of material, such as tissues of plants or animal. In RADIOTHERAPY, radiation dosage is expressed in gray units (Gy). In RADIOLOGIC HEALTH, the dosage is expressed by the product of absorbed dose (Gy) and quality factor (a function of linear energy transfer), and is called radiation dose equivalent in sievert units (Sv). Sievert Units,Dosage, Radiation,Gray Units,Gy Radiation,Sv Radiation Dose Equivalent,Dosages, Radiation,Radiation Dosages,Units, Gray,Units, Sievert
D011835 Radiation Protection Methods and practices adopted to protect against RADIATION. Protection, Radiation
D005471 Fluoroscopy Production of an image when x-rays strike a fluorescent screen. Fluoroscopies
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D013819 Thermoluminescent Dosimetry The use of a device composed of thermoluminescent material for measuring exposure to IONIZING RADIATION. The thermoluminescent material emits light when heated. The amount of light emitted is proportional to the amount of ionizing radiation to which the material has been exposed. Dosimetries, Thermoluminescent,Dosimetry, Thermoluminescent,Thermoluminescent Dosimetries
D015641 Radiography, Interventional Diagnostic and therapeutic procedures that are invasive or surgical in nature, and require the expertise of a specially trained radiologist. In general, they are more invasive than diagnostic imaging but less invasive than major surgery. They often involve catheterization, fluoroscopy, or computed tomography. Some examples include percutaneous transhepatic cholangiography, percutaneous transthoracic biopsy, balloon angioplasty, and arterial embolization. Interventional Radiography
D016273 Occupational Exposure The exposure to potentially harmful chemical, physical, or biological agents that occurs as a result of one's occupation. Exposure, Occupational,Exposures, Occupational,Occupational Exposures
D019047 Phantoms, Imaging Devices or objects in various imaging techniques used to visualize or enhance visualization by simulating conditions encountered in the procedure. Phantoms are used very often in procedures employing or measuring x-irradiation or radioactive material to evaluate performance. Phantoms often have properties similar to human tissue. Water demonstrates absorbing properties similar to normal tissue, hence water-filled phantoms are used to map radiation levels. Phantoms are used also as teaching aids to simulate real conditions with x-ray or ultrasonic machines. (From Iturralde, Dictionary and Handbook of Nuclear Medicine and Clinical Imaging, 1990) Phantoms, Radiographic,Phantoms, Radiologic,Radiographic Phantoms,Radiologic Phantoms,Phantom, Radiographic,Phantom, Radiologic,Radiographic Phantom,Radiologic Phantom,Imaging Phantom,Imaging Phantoms,Phantom, Imaging
D035002 Lower Extremity The region of the lower limb in animals, extending from the gluteal region to the FOOT, and including the BUTTOCKS; HIP; and LEG. Extremity, Lower,Lower Limb,Membrum inferius,Extremities, Lower,Limb, Lower,Limbs, Lower,Lower Extremities,Lower Limbs

Related Publications

R E R Artschan, and D S Brettle, and K Chase, and A Fender, and P G Howells, and S Buchan
January 2007, Cardiovascular and interventional radiology,
R E R Artschan, and D S Brettle, and K Chase, and A Fender, and P G Howells, and S Buchan
January 2008, The British journal of radiology,
R E R Artschan, and D S Brettle, and K Chase, and A Fender, and P G Howells, and S Buchan
May 2003, The British journal of radiology,
R E R Artschan, and D S Brettle, and K Chase, and A Fender, and P G Howells, and S Buchan
May 2018, Cardiovascular and interventional radiology,
R E R Artschan, and D S Brettle, and K Chase, and A Fender, and P G Howells, and S Buchan
January 2008, Radiation protection dosimetry,
R E R Artschan, and D S Brettle, and K Chase, and A Fender, and P G Howells, and S Buchan
January 2009, Current cardiology reviews,
R E R Artschan, and D S Brettle, and K Chase, and A Fender, and P G Howells, and S Buchan
September 2013, Radiation protection dosimetry,
R E R Artschan, and D S Brettle, and K Chase, and A Fender, and P G Howells, and S Buchan
March 2024, Radiography (London, England : 1995),
R E R Artschan, and D S Brettle, and K Chase, and A Fender, and P G Howells, and S Buchan
December 2019, Radiation protection dosimetry,
R E R Artschan, and D S Brettle, and K Chase, and A Fender, and P G Howells, and S Buchan
July 2015, Radiation protection dosimetry,
Copied contents to your clipboard!