Variability of the pullout strength of cancellous bone screws with cement augmentation. 2015

P Procter, and P Bennani, and C J Brown, and J Arnoldi, and D P Pioletti, and S Larsson
Stryker Osteosynthesis, Selzach, Switzerland; School of Engineering and Design, Brunel University, Uxbridge, UK.

BACKGROUND Orthopaedic surgeons often face clinical situations where improved screw holding power in cancellous bone is needed. Injectable calcium phosphate cements are one option to enhance fixation. METHODS Paired screw pullout tests were undertaken in which human cadaver bone was augmented with calcium phosphate cement. A finite element model was used to investigate sensitivity to screw positional placement. RESULTS Statistical analysis of the data concluded that the pullout strength was generally increased by cement augmentation in the in vitro human cadaver tests. However, when comparing the individual paired samples there were surprising results with lower strength than anticipated after augmentation, in apparent contradiction to the generally expected conclusion. Investigation using the finite element model showed that these strength reductions could be accounted for by small screw positional changes. A change of 0.5mm might result in predicted pullout force changes of up to 28%. CONCLUSIONS Small changes in screw position might lead to significant changes in pullout strength sufficient to explain the lower than expected individual pullout values in augmented cancellous bone. Consequently whilst the addition of cement at a position of low strength would increase the pullout strength at that point, it might not reach the pullout strength of the un-augmented paired test site. However, the overall effect of cement augmentation produces a significant improvement at whatever point in the bone the screw is placed. The use of polymeric bone-substitute materials for tests may not reveal the natural variation encountered in tests using real bone structures.

UI MeSH Term Description Entries
D008422 Materials Testing The testing of materials and devices, especially those used for PROSTHESES AND IMPLANTS; SUTURES; TISSUE ADHESIVES; etc., for hardness, strength, durability, safety, efficacy, and biocompatibility. Biocompatibility Testing,Biocompatible Materials Testing,Hemocompatibility Testing,Testing, Biocompatible Materials,Testing, Hemocompatible Materials,Hemocompatibility Testings,Hemocompatible Materials Testing,Materials Testing, Biocompatible,Materials Testing, Hemocompatible,Testing, Biocompatibility,Testing, Hemocompatibility,Testing, Materials,Testings, Biocompatibility
D001843 Bone Cements Adhesives used to fix prosthetic devices to bones and to cement bone to bone in difficult fractures. Synthetic resins are commonly used as cements. A mixture of monocalcium phosphate, monohydrate, alpha-tricalcium phosphate, and calcium carbonate with a sodium phosphate solution is also a useful bone paste. Bone Cement,Bone Glues,Bone Pastes,Bone Glue,Bone Paste,Cement, Bone,Cements, Bone,Glue, Bone,Glues, Bone,Paste, Bone,Pastes, Bone
D001863 Bone Screws Specialized devices used in ORTHOPEDIC SURGERY to repair bone fractures. Bone Screw,Screw, Bone,Screws, Bone
D002102 Cadaver A dead body, usually a human body. Corpse,Cadavers,Corpses
D002130 Calcium Phosphates Calcium salts of phosphoric acid. These compounds are frequently used as calcium supplements. Phosphates, Calcium
D004868 Equipment Failure Failure of equipment to perform to standard. The failure may be due to defects or improper use. Defects, Equipment,Device Failure,Failure, Equipment,Malfunction, Equipment,Medical Device Failure,Misuse, Equipment,Device Failure, Medical,Device Failures, Medical,Failure, Medical Device,Failures, Medical Device,Defect, Equipment,Device Failures,Equipment Defect,Equipment Defects,Equipment Failures,Equipment Malfunction,Equipment Malfunctions,Equipment Misuse,Equipment Misuses,Failure, Device,Failures, Device,Failures, Equipment,Malfunctions, Equipment,Misuses, Equipment
D005269 Femur The longest and largest bone of the skeleton, it is situated between the hip and the knee. Trochanter,Greater Trochanter,Lesser Trochanter,Femurs,Greater Trochanters,Lesser Trochanters,Trochanter, Greater,Trochanter, Lesser,Trochanters,Trochanters, Greater,Trochanters, Lesser
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D001696 Biomechanical Phenomena The properties, processes, and behavior of biological systems under the action of mechanical forces. Biomechanics,Kinematics,Biomechanic Phenomena,Mechanobiological Phenomena,Biomechanic,Biomechanic Phenomenas,Phenomena, Biomechanic,Phenomena, Biomechanical,Phenomena, Mechanobiological,Phenomenas, Biomechanic
D013718 Tensile Strength The maximum stress a material subjected to a stretching load can withstand without tearing. (McGraw-Hill Dictionary of Scientific and Technical Terms, 5th ed, p2001) Strength, Tensile,Strengths, Tensile,Tensile Strengths

Related Publications

P Procter, and P Bennani, and C J Brown, and J Arnoldi, and D P Pioletti, and S Larsson
August 1996, Journal of biomechanical engineering,
P Procter, and P Bennani, and C J Brown, and J Arnoldi, and D P Pioletti, and S Larsson
May 1995, Biomaterials,
P Procter, and P Bennani, and C J Brown, and J Arnoldi, and D P Pioletti, and S Larsson
December 2002, Journal of materials science. Materials in medicine,
P Procter, and P Bennani, and C J Brown, and J Arnoldi, and D P Pioletti, and S Larsson
November 2017, Veterinary surgery : VS,
P Procter, and P Bennani, and C J Brown, and J Arnoldi, and D P Pioletti, and S Larsson
February 2011, BMC musculoskeletal disorders,
P Procter, and P Bennani, and C J Brown, and J Arnoldi, and D P Pioletti, and S Larsson
August 1997, Clinical orthopaedics and related research,
P Procter, and P Bennani, and C J Brown, and J Arnoldi, and D P Pioletti, and S Larsson
October 2009, Clinical biomechanics (Bristol, Avon),
P Procter, and P Bennani, and C J Brown, and J Arnoldi, and D P Pioletti, and S Larsson
January 1996, Journal of orthopaedic trauma,
P Procter, and P Bennani, and C J Brown, and J Arnoldi, and D P Pioletti, and S Larsson
April 2012, Journal of biomechanics,
P Procter, and P Bennani, and C J Brown, and J Arnoldi, and D P Pioletti, and S Larsson
January 1993, Journal of orthopaedic trauma,
Copied contents to your clipboard!