Simulation evaluation of NIST air-kerma rate calibration standard for electronic brachytherapy. 2016

Jessica R Hiatt, and Mark J Rivard, and H Grady Hughes
Biomedical Engineering and Biotechnology, University Massachusetts Lowell, Lowell, Massachusetts 01854.

OBJECTIVE Dosimetry for the model S700 50 kV electronic brachytherapy (eBT) source (Xoft, Inc., a subsidiary of iCAD, San Jose, CA) was simulated using Monte Carlo (MC) methods by Rivard et al. ["Calculated and measured brachytherapy dosimetry parameters in water for the Xoft Axxent x-ray source: An electronic brachytherapy source," Med. Phys. 33, 4020-4032 (2006)] and recently by Hiatt et al. ["A revised dosimetric characterization of the model S700 electronic brachytherapy source containing an anode-centering plastic insert and other components not included in the 2006 model," Med. Phys. 42, 2764-2776 (2015)] with improved geometric characterization. While these studies examined the dose distribution in water, there have not previously been reports of the eBT source calibration methods beyond that recently reported by Seltzer et al. ["New national air-kerma standard for low-energy electronic brachytherapy sources," J. Res. Natl. Inst. Stand. Technol. 119, 554-574 (2014)]. Therefore, the motivation for the current study was to provide an independent determination of air-kerma rate at 50 cm in air K̇air(d=50 cm) using MC methods for the model S700 eBT source. METHODS Using CAD information provided by the vendor and disassembled sources, an MC model was created for the S700 eBT source. Simulations were run using the mcnp6 radiation transport code for the NIST Lamperti air ionization chamber according to specifications by Boutillon et al. ["Comparison of exposure standards in the 10-50 kV x-ray region," Metrologia 5, 1-11 (1969)], in air without the Lamperti chamber, and in vacuum without the Lamperti chamber. K̇air(d=50 cm) was determined using the *F4 tally with NIST values for the mass energy-absorption coefficients for air. Photon spectra were evaluated over 2 π azimuthal sampling for polar angles of 0° ≤ θ ≤ 180° every 1°. Volume averaging was averted through tight radial binning. Photon energy spectra were determined over all polar angles in both air and vacuum using the F4 tally with 0.1 keV resolution. A total of 10(11) simulated histories were run for the Lamperti chamber geometry (statistical uncertainty of 0.14%), with 10(10) histories for the in-air and in-vacuum simulations (statistical uncertainty of 0.04%). The total standard uncertainty in the calculated air-kerma rate determination amounted to 6.8%. RESULTS MC simulations determined the air-kerma rate at 50 cm from the source with the modeled Lamperti chamber to be (1.850 ± 0.126) × 10(-4) Gy/s, which was within the range of K̇air(d=50 cm) values (1.67-2.11) × 10(-4) Gy/s measured by NIST. The ratio of the photon spectra in air and in vacuum were in good agreement above 13 keV, and for θ < 150° where the influence of the Kovar sleeve and the Ag epoxy components caused increased scatter in air. Below 13 keV, the ratio of the photon spectra in air to vacuum exhibited a decrease that was attributed to increased attenuation of the photons in air. Across most of the energy range on the source transverse plane, there was good agreement between the authors' simulated spectra and that measured by NIST. Discrepancies were observed above 40 keV where the NIST spectrum had a steeper fall-off towards 50 keV. CONCLUSIONS Through MC simulations of radiation transport, this study provided an independent validation of the measured air-kerma rate at 50 cm in air at NIST for the model S700 eBT source, with mean results in agreement within 3.3%. This difference was smaller than the range (i.e., 23%) of the measured values.

UI MeSH Term Description Entries
D009010 Monte Carlo Method In statistics, a technique for numerically approximating the solution of a mathematical problem by studying the distribution of some random variable, often generated by a computer. The name alludes to the randomness characteristic of the games of chance played at the gambling casinos in Monte Carlo. (From Random House Unabridged Dictionary, 2d ed, 1993) Method, Monte Carlo
D011874 Radiometry The measurement of radiation by photography, as in x-ray film and film badge, by Geiger-Mueller tube, and by SCINTILLATION COUNTING. Geiger-Mueller Counters,Nuclear Track Detection,Radiation Dosimetry,Dosimetry, Radiation,Geiger Counter,Geiger-Mueller Counter Tube,Geiger-Mueller Probe,Geiger-Mueller Tube,Radiation Counter,Counter Tube, Geiger-Mueller,Counter Tubes, Geiger-Mueller,Counter, Geiger,Counter, Radiation,Counters, Geiger,Counters, Geiger-Mueller,Counters, Radiation,Detection, Nuclear Track,Dosimetries, Radiation,Geiger Counters,Geiger Mueller Counter Tube,Geiger Mueller Counters,Geiger Mueller Probe,Geiger Mueller Tube,Geiger-Mueller Counter Tubes,Geiger-Mueller Probes,Geiger-Mueller Tubes,Probe, Geiger-Mueller,Probes, Geiger-Mueller,Radiation Counters,Radiation Dosimetries,Tube, Geiger-Mueller,Tube, Geiger-Mueller Counter,Tubes, Geiger-Mueller,Tubes, Geiger-Mueller Counter
D012015 Reference Standards A basis of value established for the measure of quantity, weight, extent or quality, e.g. weight standards, standard solutions, methods, techniques, and procedures used in diagnosis and therapy. Standard Preparations,Standards, Reference,Preparations, Standard,Standardization,Standards,Preparation, Standard,Reference Standard,Standard Preparation,Standard, Reference
D001918 Brachytherapy A collective term for interstitial, intracavity, and surface radiotherapy. It uses small sealed or partly-sealed sources that may be placed on or near the body surface or within a natural body cavity or implanted directly into the tissues. Curietherapy,Implant Radiotherapy,Plaque Therapy, Radioisotope,Radioisotope Brachytherapy,Radiotherapy, Interstitial,Radiotherapy, Intracavity,Radiotherapy, Surface,Brachytherapy, Radioisotope,Interstitial Radiotherapy,Intracavity Radiotherapy,Radioisotope Plaque Therapy,Radiotherapy, Implant,Surface Radiotherapy,Therapy, Radioisotope Plaque
D002138 Calibration Determination, by measurement or comparison with a standard, of the correct value of each scale reading on a meter or other measuring instrument; or determination of the settings of a control device that correspond to particular values of voltage, current, frequency or other output. Calibrations
D000388 Air The mixture of gases present in the earth's atmosphere consisting of oxygen, nitrogen, carbon dioxide, and small amounts of other gases.
D014618 Vacuum A space in which the pressure is far below atmospheric pressure so that the remaining gases do not affect processes being carried on in the space. Vacuums
D035501 Uncertainty The condition in which reasonable knowledge regarding risks, benefits, or the future is not available.

Related Publications

Jessica R Hiatt, and Mark J Rivard, and H Grady Hughes
December 2021, Physical and engineering sciences in medicine,
Jessica R Hiatt, and Mark J Rivard, and H Grady Hughes
January 2014, Journal of research of the National Institute of Standards and Technology,
Jessica R Hiatt, and Mark J Rivard, and H Grady Hughes
October 2000, Medical physics,
Jessica R Hiatt, and Mark J Rivard, and H Grady Hughes
November 2011, Journal of applied clinical medical physics,
Jessica R Hiatt, and Mark J Rivard, and H Grady Hughes
June 1997, Australasian physical & engineering sciences in medicine,
Jessica R Hiatt, and Mark J Rivard, and H Grady Hughes
March 1994, International journal of radiation oncology, biology, physics,
Jessica R Hiatt, and Mark J Rivard, and H Grady Hughes
May 1995, Physics in medicine and biology,
Jessica R Hiatt, and Mark J Rivard, and H Grady Hughes
July 2011, Radiation protection dosimetry,
Jessica R Hiatt, and Mark J Rivard, and H Grady Hughes
January 1998, Journal of research of the National Institute of Standards and Technology,
Copied contents to your clipboard!