Electrode pad size, transthoracic impedance and success of external ventricular defibrillation. 1989

G W Dalzell, and S R Cunningham, and J Anderson, and A A Adgey
Regional Medical Cardiology Center, Royal Victoria Hospital, Belfast, Northern Ireland.

Electrode pad size is an important determinant of transthoracic current flow during external countershock. Self-adhesive, dual function electrocardiogram/defibrillator pads were used to assess the effect of electrode pad size on defibrillation success with low energy (200 J) shocks. The study analyzed 123 cardiac arrests due to primary ventricular fibrillation (VF) in 105 patients (74 men, 31 women) ages 40 to 84 years (mean 64). Transthoracic impedance was measured before defibrillation using a low amplitude 30-kHz current passed through the chest by way of the electrocardiogram/defibrillator pads applied anteroanteriorly. Pad diameters were small (8/8 cm) in 26 cardiac arrests, intermediate (8/12 cm) in 63 arrests and large (12/12 cm) in 34 cardiac arrests. Transthoracic impedance decreased with increasing pad size (112 +/- 17 vs 92 +/- 22 vs 72 +/- 14 omega, respectively, p = 0.0001). Only the first episode of primary VF during a cardiac arrest was analyzed. A single shock of 200 J (delivered energy) was successful in 8 of 26 (31%) arrests using small pads, in 40 of 63 (63%) with intermediate pads and in 28 of 34 (82%) with large pads (p = 0.0003). A second 200-J shock increased the cumulative defibrillation rates to 12 of 26 (46%), 50 of 63 (79%) and 33 of 34 (97%), respectively (p less than 0.0001). In primary VF, larger self-adhesive electrocardiogram/defibrillator pads are associated with a lower transthoracic impedance and improved defibrillation success rates with low energy shocks.

UI MeSH Term Description Entries
D008297 Male Males
D008875 Middle Aged An adult aged 45 - 64 years. Middle Age
D002307 Cardiography, Impedance A type of impedance plethysmography in which bioelectrical impedance is measured between electrodes positioned around the neck and around the lower thorax. It is used principally to calculate stroke volume and cardiac volume, but it is also related to myocardial contractility, thoracic fluid content, and circulation to the extremities. Impedance, Transthoracic,Plethysmography, Impedance, Transthoracic,Impedance Cardiography,Impedance Plethysmography, Transthoracic,Plethysmography, Transthoracic Impedance,Transthoracic Impedance Plethysmography,Cardiographies, Impedance,Impedance Cardiographies,Impedance Plethysmographies, Transthoracic,Impedances, Transthoracic,Plethysmographies, Transthoracic Impedance,Transthoracic Impedance,Transthoracic Impedance Plethysmographies,Transthoracic Impedances
D004554 Electric Countershock An electrical current applied to the HEART to terminate a CARDIAC ARRHYTHMIA. Cardiac Electroversion,Cardioversion,Defibrillation, Electric,Electroversion, Cardiac,Electrical Cardioversion,Electroversion Therapy,Therapy, Electroversion,Cardiac Electroversions,Cardioversion, Electrical,Cardioversions,Cardioversions, Electrical,Countershock, Electric,Countershocks, Electric,Defibrillations, Electric,Electric Countershocks,Electric Defibrillation,Electric Defibrillations,Electrical Cardioversions,Electroversion Therapies,Electroversions, Cardiac,Therapies, Electroversion
D004562 Electrocardiography Recording of the moment-to-moment electromotive forces of the HEART as projected onto various sites on the body's surface, delineated as a scalar function of time. The recording is monitored by a tracing on slow moving chart paper or by observing it on a cardioscope, which is a CATHODE RAY TUBE DISPLAY. 12-Lead ECG,12-Lead EKG,12-Lead Electrocardiography,Cardiography,ECG,EKG,Electrocardiogram,Electrocardiograph,12 Lead ECG,12 Lead EKG,12 Lead Electrocardiography,12-Lead ECGs,12-Lead EKGs,12-Lead Electrocardiographies,Cardiographies,ECG, 12-Lead,EKG, 12-Lead,Electrocardiograms,Electrocardiographies, 12-Lead,Electrocardiographs,Electrocardiography, 12-Lead
D004566 Electrodes Electric conductors through which electric currents enter or leave a medium, whether it be an electrolytic solution, solid, molten mass, gas, or vacuum. Anode,Anode Materials,Cathode,Cathode Materials,Anode Material,Anodes,Cathode Material,Cathodes,Electrode,Material, Anode,Material, Cathode
D005260 Female Females
D006323 Heart Arrest Cessation of heart beat or MYOCARDIAL CONTRACTION. If it is treated within a few minutes, heart arrest can be reversed in most cases to normal cardiac rhythm and effective circulation. Asystole,Cardiac Arrest,Cardiopulmonary Arrest,Arrest, Cardiac,Arrest, Cardiopulmonary,Arrest, Heart,Asystoles
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D000269 Adhesives Substances that cause the adherence of two surfaces. They include glues (properly collagen-derived adhesives), mucilages, sticky pastes, gums, resins, or latex. Glues,Mucilage Adhesive,Mucilage Adhesives,Mucilages, Adhesive,Adhesive,Adhesive Mucilage,Adhesive Mucilages,Adhesive, Mucilage,Adhesives, Mucilage,Glue,Mucilage, Adhesive

Related Publications

G W Dalzell, and S R Cunningham, and J Anderson, and A A Adgey
June 1998, Resuscitation,
G W Dalzell, and S R Cunningham, and J Anderson, and A A Adgey
February 1997, Pacing and clinical electrophysiology : PACE,
G W Dalzell, and S R Cunningham, and J Anderson, and A A Adgey
January 2016, Prehospital emergency care,
G W Dalzell, and S R Cunningham, and J Anderson, and A A Adgey
March 2005, Medical & biological engineering & computing,
G W Dalzell, and S R Cunningham, and J Anderson, and A A Adgey
February 2004, Resuscitation,
G W Dalzell, and S R Cunningham, and J Anderson, and A A Adgey
December 1988, Pediatrics,
G W Dalzell, and S R Cunningham, and J Anderson, and A A Adgey
February 2000, Physiological measurement,
G W Dalzell, and S R Cunningham, and J Anderson, and A A Adgey
January 1975, Medical instrumentation,
G W Dalzell, and S R Cunningham, and J Anderson, and A A Adgey
January 1973, Journal of electrocardiology,
Copied contents to your clipboard!