Evaluation of a Chemiluminescent Immunoassay for Measurement of Equine Insulin. 2017

H B Carslake, and G L Pinchbeck, and C M McGowan
Institute of Aging and Chronic Disease, Faculty of Health and Life Sciences, University of Liverpool, Neston, Wirral, UK.

BACKGROUND Many diagnostic tests for insulin dysregulation use reference intervals established with an insulin radioimmunoassay (RIA) that is no longer available. A chemiluminescent immunoassay (CLIA) is commonly used for the measurement of serum insulin concentration in clinical practice but requires further validation, especially at clinically relevant reference intervals. OBJECTIVE To evaluate the CLIA for measurement of equine insulin and compare it to the previously validated, but now unavailable RIA. METHODS Equine serum samples (n = 78) from clinical and experimental studies. METHODS In this experimental study, performance of the CLIA was evaluated using standard variables, including comparison with the RIA. Continuous and binary outcomes were analyzed. RESULTS The CLIA showed good intra-assay (coefficient of variation [CV], 1.8-2.4%) and interassay (CV, 3-7.1%) precision. Acceptable recovery on dilution (100 ± 10%) was achieved only at dilutions <1:1. Recovery on addition was acceptable. Comparison of the CLIA and RIA showed strong positive correlation (r = 0.91-0.98), with fixed and proportional bias. At 3 diagnostic cutoffs, sensitivity of CLIA compared with RIA ranged from 67 to 100% and specificity from 96 to 100%. CONCLUSIONS The CLIA is a highly repeatable assay which is suitable for within- and between-horse comparisons. Dilution of high concentration samples should be performed with charcoal-stripped serum (CSS) and at the lowest dilution factor possible. At concentrations commonly used for diagnosis of insulin dysregulation (≤100 μIU/mL), results from the CLIA tend to be lower than from the RIA and should be interpreted accordingly. Further standardization of equine insulin assays is required.

UI MeSH Term Description Entries
D007118 Immunoassay A technique using antibodies for identifying or quantifying a substance. Usually the substance being studied serves as antigen both in antibody production and in measurement of antibody by the test substance. Immunochromatographic Assay,Assay, Immunochromatographic,Assays, Immunochromatographic,Immunoassays,Immunochromatographic Assays
D007328 Insulin A 51-amino acid pancreatic hormone that plays a major role in the regulation of glucose metabolism, directly by suppressing endogenous glucose production (GLYCOGENOLYSIS; GLUCONEOGENESIS) and indirectly by suppressing GLUCAGON secretion and LIPOLYSIS. Native insulin is a globular protein comprised of a zinc-coordinated hexamer. Each insulin monomer containing two chains, A (21 residues) and B (30 residues), linked by two disulfide bonds. Insulin is used as a drug to control insulin-dependent diabetes mellitus (DIABETES MELLITUS, TYPE 1). Iletin,Insulin A Chain,Insulin B Chain,Insulin, Regular,Novolin,Sodium Insulin,Soluble Insulin,Chain, Insulin B,Insulin, Sodium,Insulin, Soluble,Regular Insulin
D008163 Luminescent Measurements Techniques used for determining the values of photometric parameters of light resulting from LUMINESCENCE. Bioluminescence Measurements,Bioluminescent Assays,Bioluminescent Measurements,Chemiluminescence Measurements,Chemiluminescent Assays,Chemiluminescent Measurements,Chemoluminescence Measurements,Luminescence Measurements,Luminescent Assays,Luminescent Techniques,Phosphorescence Measurements,Phosphorescent Assays,Phosphorescent Measurements,Assay, Bioluminescent,Assay, Chemiluminescent,Assay, Luminescent,Assay, Phosphorescent,Assays, Bioluminescent,Assays, Chemiluminescent,Assays, Luminescent,Assays, Phosphorescent,Bioluminescence Measurement,Bioluminescent Assay,Bioluminescent Measurement,Chemiluminescence Measurement,Chemiluminescent Assay,Chemiluminescent Measurement,Chemoluminescence Measurement,Luminescence Measurement,Luminescent Assay,Luminescent Measurement,Luminescent Technique,Measurement, Bioluminescence,Measurement, Bioluminescent,Measurement, Chemiluminescence,Measurement, Chemiluminescent,Measurement, Chemoluminescence,Measurement, Luminescence,Measurement, Luminescent,Measurement, Phosphorescence,Measurement, Phosphorescent,Measurements, Bioluminescence,Measurements, Bioluminescent,Measurements, Chemiluminescence,Measurements, Chemiluminescent,Measurements, Chemoluminescence,Measurements, Luminescence,Measurements, Luminescent,Measurements, Phosphorescence,Measurements, Phosphorescent,Phosphorescence Measurement,Phosphorescent Assay,Phosphorescent Measurement,Technique, Luminescent,Techniques, Luminescent
D011863 Radioimmunoassay Classic quantitative assay for detection of antigen-antibody reactions using a radioactively labeled substance (radioligand) either directly or indirectly to measure the binding of the unlabeled substance to a specific antibody or other receptor system. Non-immunogenic substances (e.g., haptens) can be measured if coupled to larger carrier proteins (e.g., bovine gamma-globulin or human serum albumin) capable of inducing antibody formation. Radioimmunoassays
D006736 Horses Large, hoofed mammals of the family EQUIDAE. Horses are active day and night with most of the day spent seeking and consuming food. Feeding peaks occur in the early morning and late afternoon, and there are several daily periods of rest. Equus caballus,Equus przewalskii,Horse, Domestic,Domestic Horse,Domestic Horses,Horse,Horses, Domestic
D000818 Animals Unicellular or multicellular, heterotrophic organisms, that have sensation and the power of voluntary movement. Under the older five kingdom paradigm, Animalia was one of the kingdoms. Under the modern three domain model, Animalia represents one of the many groups in the domain EUKARYOTA. Animal,Metazoa,Animalia
D012680 Sensitivity and Specificity Binary classification measures to assess test results. Sensitivity or recall rate is the proportion of true positives. Specificity is the probability of correctly determining the absence of a condition. (From Last, Dictionary of Epidemiology, 2d ed) Specificity,Sensitivity,Specificity and Sensitivity
D015203 Reproducibility of Results The statistical reproducibility of measurements (often in a clinical context), including the testing of instrumentation or techniques to obtain reproducible results. The concept includes reproducibility of physiological measurements, which may be used to develop rules to assess probability or prognosis, or response to a stimulus; reproducibility of occurrence of a condition; and reproducibility of experimental results. Reliability and Validity,Reliability of Result,Reproducibility Of Result,Reproducibility of Finding,Validity of Result,Validity of Results,Face Validity,Reliability (Epidemiology),Reliability of Results,Reproducibility of Findings,Test-Retest Reliability,Validity (Epidemiology),Finding Reproducibilities,Finding Reproducibility,Of Result, Reproducibility,Of Results, Reproducibility,Reliabilities, Test-Retest,Reliability, Test-Retest,Result Reliabilities,Result Reliability,Result Validities,Result Validity,Result, Reproducibility Of,Results, Reproducibility Of,Test Retest Reliability,Validity and Reliability,Validity, Face

Related Publications

H B Carslake, and G L Pinchbeck, and C M McGowan
January 2023, Frontiers in veterinary science,
H B Carslake, and G L Pinchbeck, and C M McGowan
September 1991, Hua xi yi ke da xue xue bao = Journal of West China University of Medical Sciences = Huaxi yike daxue xuebao,
H B Carslake, and G L Pinchbeck, and C M McGowan
April 1995, Clinical biochemistry,
H B Carslake, and G L Pinchbeck, and C M McGowan
August 2011, Domestic animal endocrinology,
H B Carslake, and G L Pinchbeck, and C M McGowan
December 2016, The Journal of reproduction and development,
H B Carslake, and G L Pinchbeck, and C M McGowan
January 2007, Clinical laboratory,
H B Carslake, and G L Pinchbeck, and C M McGowan
March 2017, Clinical chemistry and laboratory medicine,
H B Carslake, and G L Pinchbeck, and C M McGowan
January 2022, Frontiers in veterinary science,
H B Carslake, and G L Pinchbeck, and C M McGowan
March 1996, The Analyst,
H B Carslake, and G L Pinchbeck, and C M McGowan
January 1990, Journal of bioluminescence and chemiluminescence,
Copied contents to your clipboard!