Improved electron collimation system design for Elekta linear accelerators. 2017

Garrett M Pitcher, and Kenneth R Hogstrom, and Robert L Carver
Department of Physics and Astronomy, Louisiana State University, Baton Rouge, LA, USA.

Prototype 10 × 10 and 20 × 20-cm2 electron collimators were designed for the Elekta Infinity accelerator (MLCi2 treatment head), with the goal of reducing the trimmer weight of excessively heavy current applicators while maintaining acceptable beam flatness (±3% major axes, ±4% diagonals) and IEC leakage dose. Prototype applicators were designed initially using tungsten trimmers of constant thickness (1% electron transmission) and cross-sections with inner and outer edges positioned at 95% and 2% off-axis ratios (OARs), respectively, cast by the upstream collimating component. Despite redefining applicator size at isocenter (not 5 cm upstream) and reducing the energy range from 4-22 to 6-20 MeV, the designed 10 × 10 and 20 × 20-cm2 applicator trimmers weighed 6.87 and 10.49 kg, respectively, exceeding that of the current applicators (5.52 and 8.36 kg, respectively). Subsequently, five design modifications using analytical and/or Monte Carlo (MC) calculations were applied, reducing trimmer weight while maintaining acceptable in-field flatness and mean leakage dose. Design Modification 1 beveled the outer trimmer edges, taking advantage of only low-energy beams scattering primary electrons sufficiently to reach the outer trimmer edge. Design Modification 2 optimized the upper and middle trimmer distances from isocenter for minimal trimmer weights. Design Modification 3 moved inner trimmer edges inward, reducing trimmer weight. Design Modification 4 determined optimal X-ray jaw positions for each energy. Design Modification 5 adjusted middle and lower trimmer shapes and reduced upper trimmer thickness by 50%. Design Modifications 1→5 reduced trimmer weights from 6.87→5.86→5.52→5.87→5.43→3.73 kg for the 10 × 10-cm2 applicator and 10.49→9.04→8.62→7.73→7.35→5.09 kg for the 20 × 20-cm2 applicator. MC simulations confirmed these final designs produced acceptable in-field flatness and met IEC-specified leakage dose at 7, 13, and 20 MeV. These results allowed collimation system design for 6 × 6-25 × 25-cm2 applicators. Reducing trimmer weights by as much as 4 kg (25 × 25-cm2 applicator) should result in easier applicator handling by the radiotherapy team.

UI MeSH Term Description Entries
D009010 Monte Carlo Method In statistics, a technique for numerically approximating the solution of a mathematical problem by studying the distribution of some random variable, often generated by a computer. The name alludes to the randomness characteristic of the games of chance played at the gambling casinos in Monte Carlo. (From Random House Unabridged Dictionary, 2d ed, 1993) Method, Monte Carlo
D010315 Particle Accelerators Devices which accelerate electrically charged atomic or subatomic particles, such as electrons, protons or ions, to high velocities so they have high kinetic energy. Betatrons,Linear Accelerators,Accelerator, Linear,Accelerator, Particle,Accelerators, Linear,Accelerators, Particle,Betatron,Linear Accelerator,Particle Accelerator
D011879 Radiotherapy Dosage The total amount of radiation absorbed by tissues as a result of radiotherapy. Dosage, Radiotherapy,Dosages, Radiotherapy,Radiotherapy Dosages
D004583 Electrons Stable elementary particles having the smallest known negative charge, present in all elements; also called negatrons. Positively charged electrons are called positrons. The numbers, energies and arrangement of electrons around atomic nuclei determine the chemical identities of elements. Beams of electrons are called CATHODE RAYS. Fast Electrons,Negatrons,Positrons,Electron,Electron, Fast,Electrons, Fast,Fast Electron,Negatron,Positron
D004867 Equipment Design Methods and patterns of fabricating machines and related hardware. Design, Equipment,Device Design,Medical Device Design,Design, Medical Device,Designs, Medical Device,Device Design, Medical,Device Designs, Medical,Medical Device Designs,Design, Device,Designs, Device,Designs, Equipment,Device Designs,Equipment Designs
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man

Related Publications

Garrett M Pitcher, and Kenneth R Hogstrom, and Robert L Carver
January 2001, Journal of medical engineering & technology,
Garrett M Pitcher, and Kenneth R Hogstrom, and Robert L Carver
January 1995, Medical dosimetry : official journal of the American Association of Medical Dosimetrists,
Garrett M Pitcher, and Kenneth R Hogstrom, and Robert L Carver
February 2022, Journal of applied clinical medical physics,
Garrett M Pitcher, and Kenneth R Hogstrom, and Robert L Carver
June 2002, Medical physics,
Garrett M Pitcher, and Kenneth R Hogstrom, and Robert L Carver
July 2016, Journal of applied clinical medical physics,
Garrett M Pitcher, and Kenneth R Hogstrom, and Robert L Carver
January 1976, Medical physics,
Garrett M Pitcher, and Kenneth R Hogstrom, and Robert L Carver
January 2003, Journal of applied clinical medical physics,
Garrett M Pitcher, and Kenneth R Hogstrom, and Robert L Carver
May 2014, Health physics,
Garrett M Pitcher, and Kenneth R Hogstrom, and Robert L Carver
February 1985, International journal of radiation oncology, biology, physics,
Garrett M Pitcher, and Kenneth R Hogstrom, and Robert L Carver
May 2016, Journal of applied clinical medical physics,
Copied contents to your clipboard!