Pulsatile flow simulator for comparison of cardiac output measurements by electromagnetic flow meter and thermodilution. 1986

P J Jebson, and W S Karkow

This study examined a pulsatile flow simulator for the purpose of evaluating two measurement devices, an extracorporeal flow probe with an electromagnetic flow meter and several thermodilution catheters. We measured the performance of these devices in a range of low to high flows. Using either saline or blood as a perfusate, we obtained different results with these fluids (p less than 0.001). Each catheter behaved in a linear manner, although variation occurred among the catheters with both saline (minimum slope 1.090, maximum slope 1.190) and blood (minimum slope 1.107, maximum slope 1.154). An increase in rate and stroke volumes of the simulator did not demonstrate an identifiable trend in error. The thermodilution catheters were most accurate at 5.0 L/min irrespective of rate, stroke volume, or perfusate used. In contrast, the electromagnetic flow meter accurately represented flows across the wide range of outputs examined (2.4 to 10.7 L/min). (Slope with saline 1.091, slope with blood 1.080) Throughout the range of flow, the flow meter gave a calibration line 5% higher with blood than with saline. The results indicate that accurate measurement of pulsatile blood flow can be achieved in vitro with an electromagnetic flow meter using saline as a perfusate, provided a correction factor is determined and applied to convert values for saline to accurate values for blood.

UI MeSH Term Description Entries
D008955 Models, Cardiovascular Theoretical representations that simulate the behavior or activity of the cardiovascular system, processes, or phenomena; includes the use of mathematical equations, computers and other electronic equipment. Cardiovascular Model,Cardiovascular Models,Model, Cardiovascular
D001783 Blood Flow Velocity A value equal to the total volume flow divided by the cross-sectional area of the vascular bed. Blood Flow Velocities,Flow Velocities, Blood,Flow Velocity, Blood,Velocities, Blood Flow,Velocity, Blood Flow
D002302 Cardiac Output The volume of BLOOD passing through the HEART per unit of time. It is usually expressed as liters (volume) per minute so as not to be confused with STROKE VOLUME (volume per beat). Cardiac Outputs,Output, Cardiac,Outputs, Cardiac
D004574 Electromagnetic Fields Fields representing the joint interplay of electric and magnetic forces. Electromagnetic Field,Field, Electromagnetic,Fields, Electromagnetic
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D012212 Rheology The study of the deformation and flow of matter, usually liquids or fluids, and of the plastic flow of solids. The concept covers consistency, dilatancy, liquefaction, resistance to flow, shearing, thixotrophy, and VISCOSITY. Flowmetry,Velocimetry,Velocimetries
D013815 Thermodilution Measurement of blood flow based on induction at one point of the circulation of a known change in the intravascular heat content of flowing blood and detection of the resultant change in temperature at a point downstream. Thermodilutions

Related Publications

P J Jebson, and W S Karkow
January 1974, Surgical forum,
P J Jebson, and W S Karkow
May 1980, The Journal of thoracic and cardiovascular surgery,
P J Jebson, and W S Karkow
July 1984, Critical care medicine,
P J Jebson, and W S Karkow
November 1986, Canadian Anaesthetists' Society journal,
P J Jebson, and W S Karkow
November 1986, Heart & lung : the journal of critical care,
P J Jebson, and W S Karkow
November 1995, American journal of critical care : an official publication, American Association of Critical-Care Nurses,
P J Jebson, and W S Karkow
February 1980, Journal of pharmacological methods,
P J Jebson, and W S Karkow
May 1990, Canadian journal of anaesthesia = Journal canadien d'anesthesie,
P J Jebson, and W S Karkow
September 2002, Intensive care medicine,
Copied contents to your clipboard!