Dosimetric effects of brass mesh bolus on skin dose and dose at depth for postmastectomy chest wall irradiation. 2018
OBJECTIVE To investigate the feasibility of using the brass mesh bolus as an alternative to tissue- equivalent bolus for post mastectomy chest wall cancer by characterizing the dosimetric effects of the 2-mm fine brass bolus on both the skin dose, the dose at depth and spatial distribution. METHODS Surface dose and percent depth dose data were acquired for a 6 MV photon beam in a solid water phantom using MOSkin™, Gafchromic EBT3 film and an Advanced Markus ionization chamber. Data were acquired for the case of: no bolus, Face-up bass bolus, Face-down brass bolus, double brass bolus, 0.5 cm and 1.0 cm of Superflab TE bolus. The exit doses were also measured via MOSkin™ dosimeter and Markus ionization chamber. Gafchromic EBT3 film strips were used to plot dose profile at surface and 10 cm depth for Face-up brass, Face-down brass, double brass, 0.5 cm and 1.0 cm of Superflab TE bolus. RESULTS The surface dose measured via MOSkin™ dosimeter increased from 19.2 ± 1.0% to 63.1 ± 2.1% under Face-up brass discs, 51.2 ± 1.2% under Face-up brass spaces, 61.5 ± 0.5% under Face-down brass discs, and 41.3 ± 2.1% under Face-down brass spaces. The percentage difference in the dose measured under brass discs between Face-up versus Face-down was less than 2% for entrance dose and 10% for exit dose, whereas the percentage difference under brass spaces was approximately 3% for entrance dose and about 5% for the exit dose. Gafchromic EBT3 film strip measurements show that the mesh bolus produced ripple beam profiles due to the mesh brass construction. CONCLUSIONS Brass bolus does not significantly change dose at depth (less than 0.5%), and the surface dose is increased similar to TE bolus. Considering this, brass mesh may be used as a substitute for TE bolus to increase superficial dose for chest wall tangent plans.