An evaluation of solid state detectors for the relative dosimetry of kilovoltage X-ray beams. 2022

John Daniel, and Yousif A M Yousif, and Jackson Zifodya, and Robin Hill
North West Cancer Centre, Tamworth, New South Wales, Australia.

BACKGROUND Kilovoltage (kV) X-ray beams are an essential modality in radiotherapy. Solid state detectors are widely available in radiotherapy departments, but their use for kV dosimetry has been limited to date. This study aimed to evaluate the dosimetric performance of a range of solid state detectors for kV dosimetry. METHODS Percentage depth doses (PDDs) and relative output factors (ROFs) were measured on an XStrahl 300 unit (XStrahl-Ltd., UK) using 60, 100, 150, and 300 kVp X-ray beams. The fields were defined by circular applicators with field sizes of 2, 5, 8, and 10 cm diameter and square applicators of field sizes 10 × 10 and 20 × 20 cm2 . The following Physikalisch-Technische Werkstätten (PTW) dosimeters were used for measurements: Advanced Markus, PinPoint 3D and Semiflex ionization chambers; photon, electron, and stereotactic radiosurgery (SRS) diodes plus the microDiamond detector. All PDDs were normalized at 5 mm depth, and ROFs were measured at 3 mm depth to avoid collisions with the end of the applicators. ROFs measured using chambers were corrected for polarity and ion-recombination effects. CONCLUSIONS PDD measurements for 60, 100, and 150 kVp beams exhibited good agreement between all diodes and the ionization chambers over the entire range of depths except in the first few millimeters near the surface. However, for the 300 kVp, all diode detectors exhibited an overresponding behavior compared to reference depth dose data measured with the Advanced Markus chamber. ROFs with the diodes were higher than the Advanced Markus chamber at low energy, and the magnitude of these differences is inversely proportional to the field sizes. The PTW P diode showed the highest variation of up to 15% in the output factor compared to the Advanced Markus chamber. CONCLUSIONS This study evaluated the dosimetric performance of a range of solid state detectors in kV relative dosimetry. This study showed that diode detectors are a suitable replacement for ionization chambers for the PDD measurement of low energy kV beams (60-150 kVp) except for the PDD of 60 kVp with the smaller field sizes. However, an overresponding behavior of diode detectors at 300 kVp beams shows that diode detectors are not suitable for the PDD measurement of high energy kV beams. Generally, all solid state detectors overresponded to ROF measurements, indicating that it is not suitable for ROF measurements. In general, both shielded and unshielded diodes produced a similar dosimetric response, which demonstrates that the energy dependence of solid state detectors should be considered before they are used for any kV relative dosimetric measurements.

UI MeSH Term Description Entries
D011874 Radiometry The measurement of radiation by photography, as in x-ray film and film badge, by Geiger-Mueller tube, and by SCINTILLATION COUNTING. Geiger-Mueller Counters,Nuclear Track Detection,Radiation Dosimetry,Dosimetry, Radiation,Geiger Counter,Geiger-Mueller Counter Tube,Geiger-Mueller Probe,Geiger-Mueller Tube,Radiation Counter,Counter Tube, Geiger-Mueller,Counter Tubes, Geiger-Mueller,Counter, Geiger,Counter, Radiation,Counters, Geiger,Counters, Geiger-Mueller,Counters, Radiation,Detection, Nuclear Track,Dosimetries, Radiation,Geiger Counters,Geiger Mueller Counter Tube,Geiger Mueller Counters,Geiger Mueller Probe,Geiger Mueller Tube,Geiger-Mueller Counter Tubes,Geiger-Mueller Probes,Geiger-Mueller Tubes,Probe, Geiger-Mueller,Probes, Geiger-Mueller,Radiation Counters,Radiation Dosimetries,Tube, Geiger-Mueller,Tube, Geiger-Mueller Counter,Tubes, Geiger-Mueller,Tubes, Geiger-Mueller Counter
D004583 Electrons Stable elementary particles having the smallest known negative charge, present in all elements; also called negatrons. Positively charged electrons are called positrons. The numbers, energies and arrangement of electrons around atomic nuclei determine the chemical identities of elements. Beams of electrons are called CATHODE RAYS. Fast Electrons,Negatrons,Positrons,Electron,Electron, Fast,Electrons, Fast,Fast Electron,Negatron,Positron
D000072232 Radiation Dosimeters Devices that measure exposure to IONIZING RADIATION. Dosimeters,Radiation Dosimeter,Dosimeter,Dosimeter, Radiation,Dosimeters, Radiation
D014965 X-Rays Penetrating electromagnetic radiation emitted when the inner orbital electrons of an atom are excited and release radiant energy. X-ray wavelengths range from 1 pm to 10 nm. Hard X-rays are the higher energy, shorter wavelength X-rays. Soft x-rays or Grenz rays are less energetic and longer in wavelength. The short wavelength end of the X-ray spectrum overlaps the GAMMA RAYS wavelength range. The distinction between gamma rays and X-rays is based on their radiation source. Grenz Ray,Grenz Rays,Roentgen Ray,Roentgen Rays,X Ray,X-Ray,Xray,Radiation, X,X-Radiation,Xrays,Ray, Grenz,Ray, Roentgen,Ray, X,Rays, Grenz,Rays, Roentgen,Rays, X,X Radiation,X Rays,X-Radiations
D017785 Photons Discrete concentrations of energy, apparently massless elementary particles, that move at the speed of light. They are the unit or quantum of electromagnetic radiation. Photons are emitted when electrons move from one energy state to another. (From Hawley's Condensed Chemical Dictionary, 11th ed)

Related Publications

John Daniel, and Yousif A M Yousif, and Jackson Zifodya, and Robin Hill
September 2009, Medical physics,
John Daniel, and Yousif A M Yousif, and Jackson Zifodya, and Robin Hill
June 2020, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB),
John Daniel, and Yousif A M Yousif, and Jackson Zifodya, and Robin Hill
December 1998, Medical physics,
John Daniel, and Yousif A M Yousif, and Jackson Zifodya, and Robin Hill
December 2012, Australasian physical & engineering sciences in medicine,
John Daniel, and Yousif A M Yousif, and Jackson Zifodya, and Robin Hill
April 1995, Medical physics,
John Daniel, and Yousif A M Yousif, and Jackson Zifodya, and Robin Hill
February 2000, Physics in medicine and biology,
John Daniel, and Yousif A M Yousif, and Jackson Zifodya, and Robin Hill
August 2008, Physics in medicine and biology,
John Daniel, and Yousif A M Yousif, and Jackson Zifodya, and Robin Hill
November 2005, Physics in medicine and biology,
John Daniel, and Yousif A M Yousif, and Jackson Zifodya, and Robin Hill
March 2014, Physics in medicine and biology,
John Daniel, and Yousif A M Yousif, and Jackson Zifodya, and Robin Hill
January 1993, Medical dosimetry : official journal of the American Association of Medical Dosimetrists,
Copied contents to your clipboard!