On output variations of therapeutic x-ray equipment. 1979

P Zaránd

The author has measured the exposure rate of 56 orthovolt X-ray units. A comparison of consecutive measurements resulted in a 4.2% standard deviation which is mainly due to output variation of the X-ray equipment itself and due to the variation of mains supply. The repeatability of an irradiation controlled by a timer is therefore +/- 9% (on the 95% confidence level), which is much lower than recommended by the ICRU or stipulated in Hungarian National Standards. This accuracy is however sufficient in non-tumor therapy. The output variation among therapeutic X-ray units is high (sigma x approximately 22%) and correlates with the HVL measurements. The advantage of a central dosimetric service is the reduction of the number of unnecessary radiation qualities and superfluous combinations of radiation qualities, tube current and FSD.

UI MeSH Term Description Entries
D008433 Mathematics The deductive study of shape, quantity, and dependence. (From McGraw-Hill Dictionary of Scientific and Technical Terms, 6th ed) Mathematic
D011878 Radiotherapy The use of IONIZING RADIATION to treat malignant NEOPLASMS and some benign conditions. Radiotherapy, Targeted,Targeted Radiotherapy,Radiation Therapy,Radiation Therapy, Targeted,Radiation Treatment,Targeted Radiation Therapy,Radiation Therapies,Radiation Therapies, Targeted,Radiation Treatments,Radiotherapies,Radiotherapies, Targeted,Targeted Radiation Therapies,Targeted Radiotherapies,Therapies, Radiation,Therapies, Targeted Radiation,Therapy, Radiation,Therapy, Targeted Radiation,Treatment, Radiation
D011879 Radiotherapy Dosage The total amount of radiation absorbed by tissues as a result of radiotherapy. Dosage, Radiotherapy,Dosages, Radiotherapy,Radiotherapy Dosages
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man

Related Publications

P Zaránd
January 1986, Radiologia diagnostica,
P Zaránd
July 1971, British medical journal,
P Zaránd
January 1981, Applied radiology,
P Zaránd
September 1957, Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin,
P Zaránd
October 1961, The Canadian veterinary journal = La revue veterinaire canadienne,
P Zaránd
January 1977, Radiologic technology,
P Zaránd
May 1994, European journal of radiology,
Copied contents to your clipboard!