Flammability of silicone rubber: an unrecognized hazard. 1984

M M Whitacre, and R A Sargent

A patient sustained a burn when silicone rubber tubing was ignited by a pencil cautery unit. Unlike polyethylene, silicone rubber does not melt prior to reaching its ignition temperature. The application of heat near or to silicone rubber is to be avoided.

UI MeSH Term Description Entries
D007440 Intubation Introduction of a tube into a hollow organ to restore or maintain patency if obstructed. It is differentiated from CATHETERIZATION in that the insertion of a catheter is usually performed for the introducing or withdrawing of fluids from the body. Intubations
D002056 Burns Injuries to tissues caused by contact with heat, steam, chemicals (BURNS, CHEMICAL), electricity (BURNS, ELECTRIC), or the like. Burn
D002425 Cautery The application of a caustic substance, a hot instrument, an electric current, or other agent to control bleeding while removing or destroying tissue. Cauterization,Cauterizations
D002675 Child, Preschool A child between the ages of 2 and 5. Children, Preschool,Preschool Child,Preschool Children
D005260 Female Females
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D012826 Silicone Elastomers Polymers of silicone that are formed by crosslinking and treatment with amorphous silica to increase strength. They have properties similar to vulcanized natural rubber, in that they stretch under tension, retract rapidly, and fully recover to their original dimensions upon release. They are used in the encapsulation of surgical membranes and implants. Elastomers, Silicone,Rubber Silicone,Silicone Rubber,Elastosil,Microfil,SE-30,Elastosils,Microfils,SE 30,SE30,Silicone Elastomer

Related Publications

M M Whitacre, and R A Sargent
September 2020, Clinical anatomy (New York, N.Y.),
M M Whitacre, and R A Sargent
June 1992, The American journal of forensic medicine and pathology,
M M Whitacre, and R A Sargent
September 1984, Diseases of the colon and rectum,
M M Whitacre, and R A Sargent
July 1980, The Journal of reproductive medicine,
M M Whitacre, and R A Sargent
December 1965, Archives of ophthalmology (Chicago, Ill. : 1960),
M M Whitacre, and R A Sargent
May 1974, Nederlands tijdschrift voor tandheelkunde,
M M Whitacre, and R A Sargent
July 1980, Medical & biological engineering & computing,
M M Whitacre, and R A Sargent
February 1962, American journal of surgery,
M M Whitacre, and R A Sargent
April 2015, Medical & biological engineering & computing,
Copied contents to your clipboard!