Performance of angiographic catheters. 1995

I Marshall, and R J Sellar
Department of Medical Physics and Medical Engineering, Western General Hospital, Edinburgh, UK.

Simple methods of quantifying selected physical characteristics of angiographic catheters, which are likely to influence their in vivo performance, have been developed. The torsional stiffness ('torque') of catheters was measured, and tip control ('manoeuvrability') in an angiographic phantom was studied. Both braided and unbraided catheters were investigated all having the 'headhunter' tip shape. Significant differences between measurements made at 20 and 37 degrees C were found. Torsional stiffness was not a good predictor of performance in the angiographic phantom. Ethylene oxide re-sterilization has only a limited effect on the physical characteristics studied.

UI MeSH Term Description Entries
D008422 Materials Testing The testing of materials and devices, especially those used for PROSTHESES AND IMPLANTS; SUTURES; TISSUE ADHESIVES; etc., for hardness, strength, durability, safety, efficacy, and biocompatibility. Biocompatibility Testing,Biocompatible Materials Testing,Hemocompatibility Testing,Testing, Biocompatible Materials,Testing, Hemocompatible Materials,Hemocompatibility Testings,Hemocompatible Materials Testing,Materials Testing, Biocompatible,Materials Testing, Hemocompatible,Testing, Biocompatibility,Testing, Hemocompatibility,Testing, Materials,Testings, Biocompatibility
D008961 Models, Structural A representation, generally small in scale, to show the structure, construction, or appearance of something. (From Random House Unabridged Dictionary, 2d ed) Model, Structural,Structural Model,Structural Models
D002404 Catheterization Use or insertion of a tubular device into a duct, blood vessel, hollow organ, or body cavity for injecting or withdrawing fluids for diagnostic or therapeutic purposes. It differs from INTUBATION in that the tube here is used to restore or maintain patency in obstructions. Cannulation,Cannulations,Catheterizations
D004202 Disinfectants Substances used on inanimate objects that destroy harmful microorganisms or inhibit their activity. Disinfectants are classed as complete, destroying SPORES as well as vegetative forms of microorganisms, or incomplete, destroying only vegetative forms of the organisms. They are distinguished from ANTISEPTICS, which are local anti-infective agents used on humans and other animals. (From Hawley's Condensed Chemical Dictionary, 11th ed) Biocide,Disinfectant,Biocides
D004548 Elasticity Resistance and recovery from distortion of shape.
D004867 Equipment Design Methods and patterns of fabricating machines and related hardware. Design, Equipment,Device Design,Medical Device Design,Design, Medical Device,Designs, Medical Device,Device Design, Medical,Device Designs, Medical,Medical Device Designs,Design, Device,Designs, Device,Designs, Equipment,Device Designs,Equipment Designs
D004869 Equipment Safety Freedom of equipment from actual or potential hazards. Device Safety,Hazards, Equipment,Medical Device Safety,Safety, Equipment,Device Safety, Medical,Safety, Medical Device,Equipment Hazard,Equipment Hazards,Hazard, Equipment,Safety, Device
D005027 Ethylene Oxide A colorless and flammable gas at room temperature and pressure. Ethylene oxide is a bactericidal, fungicidal, and sporicidal disinfectant. It is effective against most micro-organisms, including viruses. It is used as a fumigant for foodstuffs and textiles and as an agent for the gaseous sterilization of heat-labile pharmaceutical and surgical materials. (From Reynolds, Martindale The Extra Pharmacopoeia, 30th ed, p794) Oxirane,Oxide, Ethylene
D005544 Forecasting The prediction or projection of the nature of future problems or existing conditions based upon the extrapolation or interpretation of existing scientific data or by the application of scientific methodology. Futurology,Projections and Predictions,Future,Predictions and Projections
D000792 Angiography Radiography of blood vessels after injection of a contrast medium. Arteriography,Angiogram,Angiograms,Angiographies,Arteriographies

Related Publications

I Marshall, and R J Sellar
September 1978, Circulation research,
I Marshall, and R J Sellar
June 1969, The Medical journal of Australia,
I Marshall, and R J Sellar
October 1995, Health devices,
I Marshall, and R J Sellar
July 1984, Radiology,
I Marshall, and R J Sellar
January 1995, Journal of vascular and interventional radiology : JVIR,
I Marshall, and R J Sellar
January 1995, Journal of vascular and interventional radiology : JVIR,
I Marshall, and R J Sellar
October 1999, The International journal of artificial organs,
I Marshall, and R J Sellar
July 1986, British medical journal (Clinical research ed.),
I Marshall, and R J Sellar
January 2003, Bio-medical materials and engineering,
I Marshall, and R J Sellar
August 1969, Radiology,
Copied contents to your clipboard!