The effect of compression and tensile forces on different types of maxillofacial miniplates was measured, in both as supplied and after bending states. Tests showed all plates showed values of tensile force at failure in excess of that found in most clinical situations; twisting through 90 degrees about the long axis had little effect. In flatwise bending tests there was a wide scatter of values of bending stiffness and ultimate load, indicating that some plates may be stronger than necessary. Repeatedly bending the plates reduced their bending stiffness, particularly in the continuous holed plates, and increased their ultimate load to failure. Values of bending stiffness and ultimate load were much larger when bending in the edgewise direction, with large differences between manufacturers. The data generated in the current study may assist the surgeon in deciding which type of plate to select for a particular clinical situation.

UI MeSH Term Description Entries
D008422 Materials Testing The testing of materials and devices, especially those used for PROSTHESES AND IMPLANTS; SUTURES; TISSUE ADHESIVES; etc., for hardness, strength, durability, safety, efficacy, and biocompatibility. Biocompatibility Testing,Biocompatible Materials Testing,Hemocompatibility Testing,Testing, Biocompatible Materials,Testing, Hemocompatible Materials,Hemocompatibility Testings,Hemocompatible Materials Testing,Materials Testing, Biocompatible,Materials Testing, Hemocompatible,Testing, Biocompatibility,Testing, Hemocompatibility,Testing, Materials,Testings, Biocompatibility
D001860 Bone Plates Implantable fracture fixation devices attached to bone fragments with screws to bridge the fracture gap and shield the fracture site from stress as bone heals. (UMDNS, 1999) Bone Plate,Plate, Bone,Plates, Bone
D004867 Equipment Design Methods and patterns of fabricating machines and related hardware. Design, Equipment,Device Design,Medical Device Design,Design, Medical Device,Designs, Medical Device,Device Design, Medical,Device Designs, Medical,Medical Device Designs,Design, Device,Designs, Device,Designs, Equipment,Device Designs,Equipment Designs
D004868 Equipment Failure Failure of equipment to perform to standard. The failure may be due to defects or improper use. Defects, Equipment,Device Failure,Failure, Equipment,Malfunction, Equipment,Medical Device Failure,Misuse, Equipment,Device Failure, Medical,Device Failures, Medical,Failure, Medical Device,Failures, Medical Device,Defect, Equipment,Device Failures,Equipment Defect,Equipment Defects,Equipment Failures,Equipment Malfunction,Equipment Malfunctions,Equipment Misuse,Equipment Misuses,Failure, Device,Failures, Device,Failures, Equipment,Malfunctions, Equipment,Misuses, Equipment
D005147 Facial Bones The facial skeleton, consisting of bones situated between the cranial base and the mandibular region. While some consider the facial bones to comprise the hyoid (HYOID BONE), palatine (HARD PALATE), and zygomatic (ZYGOMA) bones, MANDIBLE, and MAXILLA, others include also the lacrimal and nasal bones, inferior nasal concha, and vomer but exclude the hyoid bone. (Jablonski, Dictionary of Dentistry, 1992, p113) Bone, Facial,Bones, Facial,Facial Bone
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D012399 Rotation Motion of an object in which either one or more points on a line are fixed. It is also the motion of a particle about a fixed point. (From McGraw-Hill Dictionary of Scientific and Technical Terms, 4th ed) Clinorotation,Clinorotations,Rotations
D012886 Skull The SKELETON of the HEAD including the FACIAL BONES and the bones enclosing the BRAIN. Calvaria,Cranium,Calvarium,Skulls
D013193 Stainless Steel Stainless steel. A steel containing Ni, Cr, or both. It does not tarnish on exposure and is used in corrosive environments. (Grant & Hack's Chemical Dictionary, 5th ed) Stainless Steels,Steel, Stainless,Steels, Stainless
D013314 Stress, Mechanical A purely physical condition which exists within any material because of strain or deformation by external forces or by non-uniform thermal expansion; expressed quantitatively in units of force per unit area. Mechanical Stress,Mechanical Stresses,Stresses, Mechanical

Related Publications

R A Loukota, and J C Shelton
April 1993, International journal of oral and maxillofacial surgery,
R A Loukota, and J C Shelton
March 1997, The Journal of craniofacial surgery,
R A Loukota, and J C Shelton
May 2003, Journal of oral and maxillofacial surgery : official journal of the American Association of Oral and Maxillofacial Surgeons,
R A Loukota, and J C Shelton
June 2005, Journal of oral and maxillofacial surgery : official journal of the American Association of Oral and Maxillofacial Surgeons,
R A Loukota, and J C Shelton
November 1997, The Journal of prosthetic dentistry,
R A Loukota, and J C Shelton
January 1988, Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery,
R A Loukota, and J C Shelton
January 2020, Annals of maxillofacial surgery,
R A Loukota, and J C Shelton
January 1990, Revista ADM (Asociacion Dental Mexicana : 1988),
R A Loukota, and J C Shelton
January 2020, The Journal of craniofacial surgery,
R A Loukota, and J C Shelton
June 2001, International journal of oral and maxillofacial surgery,
Copied contents to your clipboard!