Liquid disinfecting and sterilization reprocessors used for flexible endoscopes. 1994


In this issue, we evaluate three liquid disinfecting flexible endoscope reprocessors that can be used with a user-supplied liquid chemical germicide (LCG), primarily for high-level disinfection (HLD). Applying most of the same criteria and test methods, we also evaluated a fourth unit, presented separately, that is marketed as a liquid sterilizing reprocessor and must be used with manufacturer-supplied, single-use containers of LCG. Although the sterilizing unit can be used for rigid endoscopes, surgical instruments, and endoscopic accessories, we evaluated its application to the reprocessing of only flexible endoscopes. Our ratings are based on the following: (1) basic performance, such as compatibility with several different types of endoscopes and LCGs, inclusion of all essential reprocessing phases, and exposure of all endoscope surfaces to the LCG and rinse water; (2) safety, such as ensuring that essential reprocessing phases cannot be skipped or omitted, that personnel exposure to LCG vapors is minimized, and that the reprocessor is unlikely to contaminate the endoscope during reprocessing; and (3) human factors design, such as ease of use and installation. We did not perform microbiological testing to confirm HLD or sterilization by any of the evaluated reprocessors because, with a suitably clean endoscope that is in good condition, no evidence indicates that the LCGs used would not be effective on the surfaces that they contact; also, the results would be unique to the two flexible endoscopes that we used for our testing. We rated all four reprocessors Conditionally Acceptable. The units are Acceptable on the condition that users understand that they cannot be used to process the elevator cable channel of side-viewing duodenoscopes; additional conditions apply only to the liquid disinfecting units. However, if the conditions are met, all four units reduce the likelihood of using a contaminated endoscope on a patient and reduce personnel exposure to the LCG. Selection of a unit will be based on the needs and preferences of the individual healthcare facility. Our ratings also assume that these units will be used according to the manufacturers' instructions and the endoscope's operator's manual. Note that, although the evaluated liquid disinfecting units provide detergent-flushing, post-detergent water-rinse, and post-detergent water-rinse-removal phases, manual cleaning of endoscopes before automatic reprocessing is essential, regardless of which reprocessor is used.

UI MeSH Term Description Entries
D007507 Therapeutic Irrigation The washing of a body cavity or surface by flowing water or solution for therapy or diagnosis. Douching,Lavage,Douchings,Irrigation, Therapeutic,Irrigations, Therapeutic,Lavages,Therapeutic Irrigations
D011682 Purchasing, Hospital Hospital department responsible for the purchasing of supplies and equipment. Hospital Purchasing,Hospital Purchasings,Purchasings, Hospital
D001999 Bronchoscopy Endoscopic examination, therapy or surgery of the bronchi. Bronchoscopic Surgical Procedures,Surgical Procedures, Bronchoscopic,Bronchoscopic Surgery,Surgery, Bronchoscopic,Bronchoscopic Surgeries,Bronchoscopic Surgical Procedure,Bronchoscopies,Surgeries, Bronchoscopic,Surgical Procedure, Bronchoscopic
D003428 Cross Infection Any infection which a patient contracts in a health-care institution. Hospital Infections,Nosocomial Infections,Health Care Associated Infection,Health Care Associated Infections,Healthcare Associated Infections,Infection, Cross,Infections, Hospital,Infections, Nosocomial,Cross Infections,Healthcare Associated Infection,Hospital Infection,Infection, Healthcare Associated,Infection, Hospital,Infection, Nosocomial,Infections, Cross,Infections, Healthcare Associated,Nosocomial Infection
D004203 Disinfection Rendering pathogens harmless through the use of heat, antiseptics, antibacterial agents, etc.
D004865 Equipment and Supplies, Hospital Any materials used in providing care specifically in the hospital. Hospital Equipment and Supplies,Equipment, Hospital,Hospital Equipment,Hospital Supplies,Supplies, Hospital,Supply, Hospital,Hospital Supply
D004866 Equipment Contamination The presence of an infectious agent on instruments, prostheses, or other inanimate articles. Contamination, Equipment,Contaminations, Equipment,Equipment Contaminations
D004867 Equipment Design Methods and patterns of fabricating machines and related hardware. Design, Equipment,Device Design,Medical Device Design,Design, Medical Device,Designs, Medical Device,Device Design, Medical,Device Designs, Medical,Medical Device Designs,Design, Device,Designs, Device,Designs, Equipment,Device Designs,Equipment Designs
D004869 Equipment Safety Freedom of equipment from actual or potential hazards. Device Safety,Hazards, Equipment,Medical Device Safety,Safety, Equipment,Device Safety, Medical,Safety, Medical Device,Equipment Hazard,Equipment Hazards,Hazard, Equipment,Safety, Device
D005069 Evaluation Studies as Topic Works about studies that determine the effectiveness or value of processes, personnel, and equipment, or the material on conducting such studies. Critique,Evaluation Indexes,Evaluation Methodology,Evaluation Report,Evaluation Research,Methodology, Evaluation,Pre-Post Tests,Qualitative Evaluation,Quantitative Evaluation,Theoretical Effectiveness,Use-Effectiveness,Critiques,Effectiveness, Theoretical,Evaluation Methodologies,Evaluation Reports,Evaluation, Qualitative,Evaluation, Quantitative,Evaluations, Qualitative,Evaluations, Quantitative,Indexes, Evaluation,Methodologies, Evaluation,Pre Post Tests,Pre-Post Test,Qualitative Evaluations,Quantitative Evaluations,Report, Evaluation,Reports, Evaluation,Research, Evaluation,Test, Pre-Post,Tests, Pre-Post,Use Effectiveness

Related Publications

No sufficiently related or similar publications found.

Copied contents to your clipboard!