In vitro accuracy of three blood O2 saturation optic catheter systems. 1994

G Janvier, and H Guenard, and A M Lomenech
Département d'Anesthésiologie, Hôpital Pellegrin, Bordeaux, France.

OBJECTIVE The measure of the precision and accuracy of three optic SvO2 systems in vitro (SAT-2 Baxter, oximetrix 3 Abbott, HEMOPRO-2 Spectramed in a wide range of saturations (24-85%) and hematocrit (20-40%). METHODS In vitro comparison of the results given by a Co-Oximeter and the SvO2 systems. RESULTS 144 blood samples were tonometered with a variety of gas mixtures. Measurements of SvO2 were compared to those obtained by means of a Co-Oximeter, the difference (dSO2) between paired values were calculated. Precision was evaluated from confidence intervals (2SD) and accuracy was tested using dSO2 figures. As there was no difference in either precision or accuracy among catheters for a given system, data obtained from a given system were pooled. Mean precision was 14.2%, 8.6%, 9% for the Spectramed, Baxter and Abbott systems, respectively. CONCLUSIONS The rather low reliability of the optic method should be borne in mind in patients with low SvO2 and/or low hematocrit.

UI MeSH Term Description Entries
D010092 Oximetry The determination of oxygen-hemoglobin saturation of blood either by withdrawing a sample and passing it through a classical photoelectric oximeter or by electrodes attached to some translucent part of the body like finger, earlobe, or skin fold. It includes non-invasive oxygen monitoring by pulse oximetry. Pulse Oximetry,Oximetry, Pulse,Oximetries,Oximetries, Pulse,Pulse Oximetries
D010100 Oxygen An element with atomic symbol O, atomic number 8, and atomic weight [15.99903; 15.99977]. It is the most abundant element on earth and essential for respiration. Dioxygen,Oxygen-16,Oxygen 16
D005336 Fiber Optic Technology The technology of transmitting light over long distances through strands of glass or other transparent material. Fiber Optic Technologies,Optic Technologies, Fiber,Optic Technology, Fiber,Technologies, Fiber Optic,Technology, Fiber Optic
D006400 Hematocrit The volume of packed RED BLOOD CELLS in a blood specimen. The volume is measured by centrifugation in a tube with graduated markings, or with automated blood cell counters. It is an indicator of erythrocyte status in disease. For example, ANEMIA shows a low value; POLYCYTHEMIA, a high value. Erythrocyte Volume, Packed,Packed Red-Cell Volume,Erythrocyte Volumes, Packed,Hematocrits,Packed Erythrocyte Volume,Packed Erythrocyte Volumes,Packed Red Cell Volume,Packed Red-Cell Volumes,Red-Cell Volume, Packed,Red-Cell Volumes, Packed,Volume, Packed Erythrocyte,Volume, Packed Red-Cell,Volumes, Packed Erythrocyte,Volumes, Packed Red-Cell
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D000704 Analysis of Variance A statistical technique that isolates and assesses the contributions of categorical independent variables to variation in the mean of a continuous dependent variable. ANOVA,Analysis, Variance,Variance Analysis,Analyses, Variance,Variance Analyses
D015203 Reproducibility of Results The statistical reproducibility of measurements (often in a clinical context), including the testing of instrumentation or techniques to obtain reproducible results. The concept includes reproducibility of physiological measurements, which may be used to develop rules to assess probability or prognosis, or response to a stimulus; reproducibility of occurrence of a condition; and reproducibility of experimental results. Reliability and Validity,Reliability of Result,Reproducibility Of Result,Reproducibility of Finding,Validity of Result,Validity of Results,Face Validity,Reliability (Epidemiology),Reliability of Results,Reproducibility of Findings,Test-Retest Reliability,Validity (Epidemiology),Finding Reproducibilities,Finding Reproducibility,Of Result, Reproducibility,Of Results, Reproducibility,Reliabilities, Test-Retest,Reliability, Test-Retest,Result Reliabilities,Result Reliability,Result Validities,Result Validity,Result, Reproducibility Of,Results, Reproducibility Of,Test Retest Reliability,Validity and Reliability,Validity, Face
D016001 Confidence Intervals A range of values for a variable of interest, e.g., a rate, constructed so that this range has a specified probability of including the true value of the variable. Confidence Interval,Interval, Confidence,Intervals, Confidence
D066298 In Vitro Techniques Methods to study reactions or processes taking place in an artificial environment outside the living organism. In Vitro Test,In Vitro Testing,In Vitro Tests,In Vitro as Topic,In Vitro,In Vitro Technique,In Vitro Testings,Technique, In Vitro,Techniques, In Vitro,Test, In Vitro,Testing, In Vitro,Testings, In Vitro,Tests, In Vitro,Vitro Testing, In

Related Publications

G Janvier, and H Guenard, and A M Lomenech
May 1957, La Medicina del lavoro,
G Janvier, and H Guenard, and A M Lomenech
April 1992, Journal of applied physiology (Bethesda, Md. : 1985),
G Janvier, and H Guenard, and A M Lomenech
September 1977, Microvascular research,
G Janvier, and H Guenard, and A M Lomenech
January 1983, Journal of applied physiology: respiratory, environmental and exercise physiology,
G Janvier, and H Guenard, and A M Lomenech
June 1968, Orvosi hetilap,
G Janvier, and H Guenard, and A M Lomenech
May 1988, Journal of applied physiology (Bethesda, Md. : 1985),
G Janvier, and H Guenard, and A M Lomenech
September 1983, Critical care medicine,
G Janvier, and H Guenard, and A M Lomenech
January 1998, Cardiovascular and interventional radiology,
G Janvier, and H Guenard, and A M Lomenech
February 2014, Trauma monthly,
G Janvier, and H Guenard, and A M Lomenech
January 1947, Federation proceedings,
Copied contents to your clipboard!