The effect of temperature-guided radiofrequency ablation of ventricular myocardium. 1993

E Kongsgaard, and A Foerster, and H Aass, and J P Amlie
Department of Pathology, Rikshospitalet, Oslo, Norway.

The safety and feasibility of temperature-guided radiofrequency (RF) ablation of ventricular myocardium were assessed in an open-chest animal model. RF ablation was performed in the ventricles of 19 pigs using ablation catheters with a tip electrode of 2 or 4 mm length. The energy was delivered in a bipolar (2 mm tip electrode only) or unipolar configuration. Set temperature ranged from 60 to 90 degrees C. Pulse duration was 30 s. Histological examination showed coagulation necrosis with a haemorrhagic zone. However, lesions with a deep cleavage were found after five unipolar (2 mm tip electrode) energy applications. No such lesions were found after unipolar energy applications with a 4 mm tip electrode. During or shortly after ablation, premature ventricular beats and non-sustained VT were frequently observed and in some instances ventricular fibrillation. In the bipolar mode we found a positive correlation between lesion area/volume and peak temperature with a correlation coefficient of 0.48 and 0.56, respectively (P < 0.05). However no correlation was found between lesion size and applied energy. In the unipolar configuration there was no correlation between lesion size and these parameters. Median depth and area were: bipolar: 4.0 mm/23.5 mm2, unipolar (2 mm tip electrode): 3.5 mm/12.2 mm2, unipolar (4 mm tip electrode): 4.0 mm/15.7 mm2. We conclude that in a beating heart it is difficult to predict lesion size from temperature or energy.

UI MeSH Term Description Entries
D009206 Myocardium The muscle tissue of the HEART. It is composed of striated, involuntary muscle cells (MYOCYTES, CARDIAC) connected to form the contractile pump to generate blood flow. Muscle, Cardiac,Muscle, Heart,Cardiac Muscle,Myocardia,Cardiac Muscles,Heart Muscle,Heart Muscles,Muscles, Cardiac,Muscles, Heart
D004566 Electrodes Electric conductors through which electric currents enter or leave a medium, whether it be an electrolytic solution, solid, molten mass, gas, or vacuum. Anode,Anode Materials,Cathode,Cathode Materials,Anode Material,Anodes,Cathode Material,Cathodes,Electrode,Material, Anode,Material, Cathode
D006339 Heart Rate The number of times the HEART VENTRICLES contract per unit of time, usually per minute. Cardiac Rate,Chronotropism, Cardiac,Heart Rate Control,Heartbeat,Pulse Rate,Cardiac Chronotropy,Cardiac Chronotropism,Cardiac Rates,Chronotropy, Cardiac,Control, Heart Rate,Heart Rates,Heartbeats,Pulse Rates,Rate Control, Heart,Rate, Cardiac,Rate, Heart,Rate, Pulse
D006352 Heart Ventricles The lower right and left chambers of the heart. The right ventricle pumps venous BLOOD into the LUNGS and the left ventricle pumps oxygenated blood into the systemic arterial circulation. Cardiac Ventricle,Cardiac Ventricles,Heart Ventricle,Left Ventricle,Right Ventricle,Left Ventricles,Right Ventricles,Ventricle, Cardiac,Ventricle, Heart,Ventricle, Left,Ventricle, Right,Ventricles, Cardiac,Ventricles, Heart,Ventricles, Left,Ventricles, Right
D006439 Hemodynamics The movement and the forces involved in the movement of the blood through the CARDIOVASCULAR SYSTEM. Hemodynamic
D000818 Animals Unicellular or multicellular, heterotrophic organisms, that have sensation and the power of voluntary movement. Under the older five kingdom paradigm, Animalia was one of the kingdoms. Under the modern three domain model, Animalia represents one of the many groups in the domain EUKARYOTA. Animal,Metazoa,Animalia
D017115 Catheter Ablation Removal of tissue with electrical current delivered via electrodes positioned at the distal end of a catheter. Energy sources are commonly direct current (DC-shock) or alternating current at radiofrequencies (usually 750 kHz). The technique is used most often to ablate the AV junction and/or accessory pathways in order to interrupt AV conduction and produce AV block in the treatment of various tachyarrhythmias. Ablation, Transvenous Electric,Catheter Ablation, Electric,Catheter Ablation, Percutaneous,Catheter Ablation, Radiofrequency,Catheter Ablation, Transvenous,Ablation, Catheter,Ablation, Transvenous Electrical,Catheter Ablation, Electrical,Electric Catheter Ablation,Electrical Catheter Ablation,Percutaneous Catheter Ablation,Radiofrequency Catheter Ablation,Transvenous Catheter Ablation,Ablation, Electric Catheter,Ablation, Electrical Catheter,Ablation, Percutaneous Catheter,Ablation, Radiofrequency Catheter,Ablation, Transvenous Catheter,Electric Ablation, Transvenous,Electrical Ablation, Transvenous,Transvenous Electric Ablation,Transvenous Electrical Ablation

Related Publications

E Kongsgaard, and A Foerster, and H Aass, and J P Amlie
May 2010, International heart journal,
E Kongsgaard, and A Foerster, and H Aass, and J P Amlie
June 1994, American heart journal,
E Kongsgaard, and A Foerster, and H Aass, and J P Amlie
September 2001, Basic research in cardiology,
E Kongsgaard, and A Foerster, and H Aass, and J P Amlie
June 1997, The Korean journal of internal medicine,
E Kongsgaard, and A Foerster, and H Aass, and J P Amlie
May 1997, Pacing and clinical electrophysiology : PACE,
E Kongsgaard, and A Foerster, and H Aass, and J P Amlie
September 1998, Journal of cardiovascular electrophysiology,
E Kongsgaard, and A Foerster, and H Aass, and J P Amlie
February 2010, Sheng wu yi xue gong cheng xue za zhi = Journal of biomedical engineering = Shengwu yixue gongchengxue zazhi,
E Kongsgaard, and A Foerster, and H Aass, and J P Amlie
March 1998, Pacing and clinical electrophysiology : PACE,
E Kongsgaard, and A Foerster, and H Aass, and J P Amlie
September 2018, Pacing and clinical electrophysiology : PACE,
Copied contents to your clipboard!