Characterization of a multi-leaf collimator system. 1993

J M Galvin, and A R Smith, and B Lally
New York University Medical Center, Tisch Hospital, Division of Radiation Oncology, NY 10016.

Commissioning measurements for a multi-leaf collimator installed on a dual energy accelerator with 6 and 15 MV photons are described. Detailed dosimetric characterization of the multi-leaf collimator is a requirement for modeling the collimator with treatment planning software. Measurements include a determination of the penumbra width, leaf transmission, between-leaf leakage, and localization of the leaf ends and sides. Standard radiographic film was used for the penumbra measurements, and separate experiments using radiochromic film and thermoluminescent dosimeters were performed to verify that distortions of the dose distribution at an edge due to changing energy sensitivity of silver bromide film are negligible. Films were analyzed with a scanning laser densitometer with a 210 micron spot. Little change in the penumbra edge distribution was noted for different positions of a leaf in the field. Experiments localizing the physical end of the leaves showed less than 1 mm deviation from the 50% decrement line. This small difference is attributed to the shaped end on the leaves. One side of a single leaf corresponded to the 50% decrement line, but the opposite face was aligned with a lower value. This difference is due to the tongue and groove used to decrease between-leaf leakage. For both energies, approximately 2% of photons incident on the multi-leaf collimator are transmitted and an additional 0.5% leakage occurs between the leaves. Alignment of the leaves to form a straight edge results in a penumbra profile which compares favorably with the standard technique of using alloy blocks. When the edge is stepped, the isodose lines follow the leaf pattern and the boundary is poorly defined compared to divergent blocks.

UI MeSH Term Description Entries
D011879 Radiotherapy Dosage The total amount of radiation absorbed by tissues as a result of radiotherapy. Dosage, Radiotherapy,Dosages, Radiotherapy,Radiotherapy Dosages
D011880 Radiotherapy Planning, Computer-Assisted Computer-assisted mathematical calculations of beam angles, intensities of radiation, and duration of irradiation in radiotherapy. Computer-Assisted Radiotherapy Planning,Dosimetry Calculations, Computer-Assisted,Planning, Computer-Assisted Radiotherapy,Calculation, Computer-Assisted Dosimetry,Calculations, Computer-Assisted Dosimetry,Computer Assisted Radiotherapy Planning,Computer-Assisted Dosimetry Calculation,Computer-Assisted Dosimetry Calculations,Dosimetry Calculation, Computer-Assisted,Dosimetry Calculations, Computer Assisted,Planning, Computer Assisted Radiotherapy,Radiotherapy Planning, Computer Assisted
D011882 Radiotherapy, High-Energy Radiotherapy using high-energy (megavolt or higher) ionizing radiation. Types of radiation include gamma rays, produced by a radioisotope within a teletherapy unit; x-rays, electrons, protons, alpha particles (helium ions) and heavy charged ions, produced by particle acceleration; and neutrons and pi-mesons (pions), produced as secondary particles following bombardment of a target with a primary particle. Megavolt Radiotherapy,High-Energy Radiotherapy,Radiotherapy, Megavolt,High Energy Radiotherapy,Radiotherapy, High Energy
D004867 Equipment Design Methods and patterns of fabricating machines and related hardware. Design, Equipment,Device Design,Medical Device Design,Design, Medical Device,Designs, Medical Device,Device Design, Medical,Device Designs, Medical,Medical Device Designs,Design, Device,Designs, Device,Designs, Equipment,Device Designs,Equipment Designs
D005069 Evaluation Studies as Topic Works about studies that determine the effectiveness or value of processes, personnel, and equipment, or the material on conducting such studies. Critique,Evaluation Indexes,Evaluation Methodology,Evaluation Report,Evaluation Research,Methodology, Evaluation,Pre-Post Tests,Qualitative Evaluation,Quantitative Evaluation,Theoretical Effectiveness,Use-Effectiveness,Critiques,Effectiveness, Theoretical,Evaluation Methodologies,Evaluation Reports,Evaluation, Qualitative,Evaluation, Quantitative,Evaluations, Qualitative,Evaluations, Quantitative,Indexes, Evaluation,Methodologies, Evaluation,Pre Post Tests,Pre-Post Test,Qualitative Evaluations,Quantitative Evaluations,Report, Evaluation,Reports, Evaluation,Research, Evaluation,Test, Pre-Post,Tests, Pre-Post,Use Effectiveness
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man

Related Publications

J M Galvin, and A R Smith, and B Lally
January 2012, Nihon Hoshasen Gijutsu Gakkai zasshi,
J M Galvin, and A R Smith, and B Lally
July 2012, Physics in medicine and biology,
J M Galvin, and A R Smith, and B Lally
December 2017, Medical physics,
J M Galvin, and A R Smith, and B Lally
December 2011, Physics in medicine and biology,
J M Galvin, and A R Smith, and B Lally
January 2011, Nihon Hoshasen Gijutsu Gakkai zasshi,
J M Galvin, and A R Smith, and B Lally
April 2006, Physics in medicine and biology,
J M Galvin, and A R Smith, and B Lally
March 1999, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology,
J M Galvin, and A R Smith, and B Lally
June 2002, Medical physics,
Copied contents to your clipboard!