A three-axis hip joint simulator for wear and friction studies on total hip prostheses. 1996

V O Saikko
Laboratory of Machine Design, Helsinki University of Technology, Finland.

A three-axial, single-station hip joint simulator was designed and built for wear and friction studies on total hip prostheses. The design of the apparatus is described in detail. Continuous level walking is simulated. All three motion components, flexion-extension, abduction-adduction and internal-external rotation, are included. The motions are implemented electromechanically and the uniaxial load pneumatically. The load is measured continuously. For accurate measurement of wear, the apparatus has a loaded control joint, which also renders both the test and control joints self-centering, as they are loaded in series. The frictional torque of the test joint can be measured continuously throughout the wear test, which is an exceptional feature. Four tests of five million cycles each were completed using 32 mm diameter Co-Cr-Mo femoral heads and 5.6 mm thick, metal-backed, ultra-high molecular weight polyethylene acetabular cups as test specimens. Their wear and friction behaviour is described and discussed in relation to previous simulator studies and clinical observations. The lubricant was distilled water, maintained at body temperature. The wear of the cups was measured gravimetrically at intervals. The average wear rate was 3.9 mg/one million cycles, corresponding to 0.03 mm/year, and the average coefficient of friction was 0.01.

UI MeSH Term Description Entries
D008153 Lubrication The application of LUBRICANTS to diminish FRICTION between two surfaces. Lubrications
D008422 Materials Testing The testing of materials and devices, especially those used for PROSTHESES AND IMPLANTS; SUTURES; TISSUE ADHESIVES; etc., for hardness, strength, durability, safety, efficacy, and biocompatibility. Biocompatibility Testing,Biocompatible Materials Testing,Hemocompatibility Testing,Testing, Biocompatible Materials,Testing, Hemocompatible Materials,Hemocompatibility Testings,Hemocompatible Materials Testing,Materials Testing, Biocompatible,Materials Testing, Hemocompatible,Testing, Biocompatibility,Testing, Hemocompatibility,Testing, Materials,Testings, Biocompatibility
D008954 Models, Biological Theoretical representations that simulate the behavior or activity of biological processes or diseases. For disease models in living animals, DISEASE MODELS, ANIMAL is available. Biological models include the use of mathematical equations, computers, and other electronic equipment. Biological Model,Biological Models,Model, Biological,Models, Biologic,Biologic Model,Biologic Models,Model, Biologic
D009068 Movement The act, process, or result of passing from one place or position to another. It differs from LOCOMOTION in that locomotion is restricted to the passing of the whole body from one place to another, while movement encompasses both locomotion but also a change of the position of the whole body or any of its parts. Movement may be used with reference to humans, vertebrate and invertebrate animals, and microorganisms. Differentiate also from MOTOR ACTIVITY, movement associated with behavior. Movements
D010316 Particle Size Relating to the size of solids. Particle Sizes,Size, Particle,Sizes, Particle
D011095 Polyethylenes Synthetic thermoplastics that are tough, flexible, inert, and resistant to chemicals and electrical current. They are often used as biocompatible materials for prostheses and implants. Ethylene Polymers,Ethene Homopolymers,Homopolymers, Ethene,Polymers, Ethylene
D011474 Prosthesis Design The plan and delineation of prostheses in general or a specific prosthesis. Design, Prosthesis,Designs, Prosthesis,Prosthesis Designs
D011475 Prosthesis Failure Malfunction of implantation shunts, valves, etc., and prosthesis loosening, migration, and breaking. Prosthesis Loosening,Prosthesis Durability,Prosthesis Migration,Prosthesis Survival,Durabilities, Prosthesis,Durability, Prosthesis,Failure, Prosthesis,Failures, Prosthesis,Loosening, Prosthesis,Loosenings, Prosthesis,Migration, Prosthesis,Migrations, Prosthesis,Prosthesis Durabilities,Prosthesis Failures,Prosthesis Loosenings,Prosthesis Migrations,Prosthesis Survivals,Survival, Prosthesis,Survivals, Prosthesis
D006621 Hip Joint The joint that is formed by the articulation of the head of FEMUR and the ACETABULUM of the PELVIS. Acetabulofemoral Joint,Acetabulofemoral Joints,Hip Joints,Joint, Acetabulofemoral,Joint, Hip,Joints, Acetabulofemoral,Joints, Hip
D006622 Hip Prosthesis Replacement for a hip joint. Femoral Head Prosthesis,Femoral Head Prostheses,Hip Prostheses,Prostheses, Femoral Head,Prostheses, Hip,Prosthesis, Femoral Head,Prosthesis, Hip

Related Publications

V O Saikko
January 1997, Chirurgie; memoires de l'Academie de chirurgie,
V O Saikko
January 2001, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine,
V O Saikko
August 1972, Acta orthopaedica Belgica,
V O Saikko
January 2007, Journal of biomechanics,
V O Saikko
September 1969, Annals of the rheumatic diseases,
V O Saikko
January 1979, Bulletin of the Hospital for Joint Diseases,
V O Saikko
January 1973, Archiv fur orthopadische und Unfall-Chirurgie,
V O Saikko
January 1974, The Journal of bone and joint surgery. American volume,
Copied contents to your clipboard!