Sizing emboli in blood using pulse Doppler ultrasound--I: Verification of the EBR model. 1996

M A Moehring, and J A Ritcey
Institute of Applied Physiology and Medicine, Seattle, WA 98122, USA. moehring@nwlink.com

An experiment to verify a theory describing ultrasound backscattering from emboli in flowing blood is presented. The theoretical predictions for the backscattered power versus embolus size are based on the measurement of the embolus to blood ratio (EBR) of backscattered acoustic power. This tool is necessary for in vivo clinical application because it removes the need to characterize attenuation and reflection loss in heterogeneous tissue. The experiment presented utilizes a customized ultrasound pulse Doppler that is capable of interrogating a sample volume with two different frequencies concurrently. A flow circuit including a 3.6-mm-diameter conduit in polyacrylamide gel in which emboli are observed with the dual frequency Doppler is described. The flow within the circuit has acoustic backscatter coefficient similar to blood due to a calibrated concentration of 31.1-mu diameter polystyrene microspheres. Polystyrene microsphere "emboli" having nominal diameters of 161 and 239 mu are placed in this flow loop and time series Doppler shift signatures are recorded. These signatures are investigated and a refraction artifact hypothesis is proposed to explain systematic deviation of the signatures away from theoretically expected results. Results show that gross discrimination of embolus size is feasible.

UI MeSH Term Description Entries
D007089 Image Enhancement Improvement of the quality of a picture by various techniques, including computer processing, digital filtering, echocardiographic techniques, light and ultrastructural MICROSCOPY, fluorescence spectrometry and microscopy, scintigraphy, and in vitro image processing at the molecular level. Image Quality Enhancement,Enhancement, Image,Enhancement, Image Quality,Enhancements, Image,Enhancements, Image Quality,Image Enhancements,Image Quality Enhancements,Quality Enhancement, Image,Quality Enhancements, Image
D008955 Models, Cardiovascular Theoretical representations that simulate the behavior or activity of the cardiovascular system, processes, or phenomena; includes the use of mathematical equations, computers and other electronic equipment. Cardiovascular Model,Cardiovascular Models,Model, Cardiovascular
D002138 Calibration Determination, by measurement or comparison with a standard, of the correct value of each scale reading on a meter or other measuring instrument; or determination of the settings of a control device that correspond to particular values of voltage, current, frequency or other output. Calibrations
D004617 Embolism Blocking of a blood vessel by an embolus which can be a blood clot or other undissolved material in the blood stream. Embolus,Embolisms
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D012815 Signal Processing, Computer-Assisted Computer-assisted processing of electric, ultrasonic, or electronic signals to interpret function and activity. Digital Signal Processing,Signal Interpretation, Computer-Assisted,Signal Processing, Digital,Computer-Assisted Signal Interpretation,Computer-Assisted Signal Interpretations,Computer-Assisted Signal Processing,Interpretation, Computer-Assisted Signal,Interpretations, Computer-Assisted Signal,Signal Interpretation, Computer Assisted,Signal Interpretations, Computer-Assisted,Signal Processing, Computer Assisted
D018619 Echocardiography, Doppler, Pulsed Echocardiography applying the Doppler effect, with velocity detection combined with range discrimination. Short bursts of ultrasound are transmitted at regular intervals and the echoes are demodulated as they return. Doppler Echocardiography, Pulsed,Echocardiography, Pulsed Doppler,Doppler Pulsed Echocardiography,Echocardiography, Doppler Pulsed,Pulsed Doppler Echocardiography,Pulsed Echocardiography, Doppler
D019047 Phantoms, Imaging Devices or objects in various imaging techniques used to visualize or enhance visualization by simulating conditions encountered in the procedure. Phantoms are used very often in procedures employing or measuring x-irradiation or radioactive material to evaluate performance. Phantoms often have properties similar to human tissue. Water demonstrates absorbing properties similar to normal tissue, hence water-filled phantoms are used to map radiation levels. Phantoms are used also as teaching aids to simulate real conditions with x-ray or ultrasonic machines. (From Iturralde, Dictionary and Handbook of Nuclear Medicine and Clinical Imaging, 1990) Phantoms, Radiographic,Phantoms, Radiologic,Radiographic Phantoms,Radiologic Phantoms,Phantom, Radiographic,Phantom, Radiologic,Radiographic Phantom,Radiologic Phantom,Imaging Phantom,Imaging Phantoms,Phantom, Imaging

Related Publications

M A Moehring, and J A Ritcey
May 2012, Ultrasound in medicine & biology,
M A Moehring, and J A Ritcey
January 1994, IEEE transactions on bio-medical engineering,
M A Moehring, and J A Ritcey
January 2008, European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery,
M A Moehring, and J A Ritcey
March 1994, Journal of the neurological sciences,
M A Moehring, and J A Ritcey
February 1991, Stroke,
M A Moehring, and J A Ritcey
August 2010, Annals of vascular surgery,
M A Moehring, and J A Ritcey
January 1996, Ultrasound in medicine & biology,
M A Moehring, and J A Ritcey
March 2008, IEEE transactions on bio-medical engineering,
Copied contents to your clipboard!