Transtelephone monitoring for pacemaker follow-up 1981-1994. 1996

S Platt, and S Furman, and J N Gross, and C Andrews, and M Benedek
Department of Medicine, Montefiore Medical Center, Bronx, NY 10467-2409, USA.

Transtelephone monitoring (TTM) is capable of detecting pacemaker pulse generator malfunction, battery depletion, and lead failure. The accuracy of TTM was analyzed by a review of Montefiore Medical Center records between October 1981 and March 1994. Each group of transmissions from a single patient, starting with implant and ending with a pacemaker operation, was defined as a closed cycle (CLOSE), if undergoing continuing follow-up at the time of analysis, as a continuing cycle (CONT), and if a cycle had ended with death or loss to follow-up, an open cycle (OPEN). TTM records of 2,632 patients were analyzed, providing 3,291 cycles. There were 731 CONT, 433 CLOSE, and 2,127 OPEN cycles; 331 procedures were indicated by TTM, of which 279 were impending depletion, 30 sudden depletion, and 22 lead malfunctions. Of the 102 procedures not indicated by TTM, 85 were for nonurgent reasons (recall: 41; DDD upgrade: 16; patient/MD request: 28) and 17 for urgent reasons. In patients followed by TTM who had a lead problem, 22 were detected by TTM before clinical manifestations and 16 were not. There were no cases in which TTM follow-up did not detect battery depletion. The total number of TTM contacts, available for 3,094 cycles, was 88,654 (range, 1-163, median 19), of which 0.4% yielded a procedure. During the same period, 75% of all secondary interventions during the first 2 years occurred during the first 2 months after implant because of lead malfunction, with a subsequent SI rate of 0.005 per month for the third through the twenty-fourth months.

UI MeSH Term Description Entries
D008991 Monitoring, Physiologic The continuous measurement of physiological processes, blood pressure, heart rate, renal output, reflexes, respiration, etc., in a patient or experimental animal; includes pharmacologic monitoring, the measurement of administered drugs or their metabolites in the blood, tissues, or urine. Patient Monitoring,Monitoring, Physiological,Physiologic Monitoring,Monitoring, Patient,Physiological Monitoring
D010138 Pacemaker, Artificial A device designed to stimulate, by electric impulses, contraction of the heart muscles. It may be temporary (external) or permanent (internal or internal-external). Cardiac Pacemaker, Artificial,Artificial Cardiac Pacemaker,Artificial Cardiac Pacemakers,Artificial Pacemaker,Artificial Pacemakers,Cardiac Pacemakers, Artificial,Pacemaker, Artificial Cardiac,Pacemakers, Artificial,Pacemakers, Artificial Cardiac
D004562 Electrocardiography Recording of the moment-to-moment electromotive forces of the HEART as projected onto various sites on the body's surface, delineated as a scalar function of time. The recording is monitored by a tracing on slow moving chart paper or by observing it on a cardioscope, which is a CATHODE RAY TUBE DISPLAY. 12-Lead ECG,12-Lead EKG,12-Lead Electrocardiography,Cardiography,ECG,EKG,Electrocardiogram,Electrocardiograph,12 Lead ECG,12 Lead EKG,12 Lead Electrocardiography,12-Lead ECGs,12-Lead EKGs,12-Lead Electrocardiographies,Cardiographies,ECG, 12-Lead,EKG, 12-Lead,Electrocardiograms,Electrocardiographies, 12-Lead,Electrocardiographs,Electrocardiography, 12-Lead
D004868 Equipment Failure Failure of equipment to perform to standard. The failure may be due to defects or improper use. Defects, Equipment,Device Failure,Failure, Equipment,Malfunction, Equipment,Medical Device Failure,Misuse, Equipment,Device Failure, Medical,Device Failures, Medical,Failure, Medical Device,Failures, Medical Device,Defect, Equipment,Device Failures,Equipment Defect,Equipment Defects,Equipment Failures,Equipment Malfunction,Equipment Malfunctions,Equipment Misuse,Equipment Misuses,Failure, Device,Failures, Device,Failures, Equipment,Malfunctions, Equipment,Misuses, Equipment
D005500 Follow-Up Studies Studies in which individuals or populations are followed to assess the outcome of exposures, procedures, or effects of a characteristic, e.g., occurrence of disease. Followup Studies,Follow Up Studies,Follow-Up Study,Followup Study,Studies, Follow-Up,Studies, Followup,Study, Follow-Up,Study, Followup
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D013689 Telephone An instrument for reproducing sounds especially articulate speech at a distance. (Webster, 3rd ed) Switchboard Service,Service, Switchboard,Services, Switchboard,Switchboard Services,Telephones

Related Publications

S Platt, and S Furman, and J N Gross, and C Andrews, and M Benedek
January 1984, IEEE engineering in medicine and biology magazine : the quarterly magazine of the Engineering in Medicine & Biology Society,
S Platt, and S Furman, and J N Gross, and C Andrews, and M Benedek
September 1975, The Annals of thoracic surgery,
S Platt, and S Furman, and J N Gross, and C Andrews, and M Benedek
January 1976, Journal of electrocardiology,
S Platt, and S Furman, and J N Gross, and C Andrews, and M Benedek
May 1971, The Journal of thoracic and cardiovascular surgery,
S Platt, and S Furman, and J N Gross, and C Andrews, and M Benedek
August 1971, New York state journal of medicine,
S Platt, and S Furman, and J N Gross, and C Andrews, and M Benedek
January 1978, Giornale italiano di cardiologia,
S Platt, and S Furman, and J N Gross, and C Andrews, and M Benedek
November 1984, Harefuah,
S Platt, and S Furman, and J N Gross, and C Andrews, and M Benedek
April 2007, The Journal of the Association of Physicians of India,
S Platt, and S Furman, and J N Gross, and C Andrews, and M Benedek
March 2019, Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing,
S Platt, and S Furman, and J N Gross, and C Andrews, and M Benedek
March 1972, Progress in cardiovascular diseases,
Copied contents to your clipboard!