Effect of electrode polarity on internal defibrillation with monophasic and biphasic waveforms using an endocardial lead system. 1997

J Huang, and B H KenKnight, and G P Walcott, and R G Walker, and W M Smith, and R E Ideker
Department of Medicine, University of Alabama at Birmingham, USA.

BACKGROUND To test the hypothesis that the effect of shock polarity on defibrillation depends on waveform duration, this study determined strength-duration defibrillation curves of monophasic and biphasic truncated exponential waveforms for both polarities. RESULTS Defibrillation thresholds (DFTs) were obtained in 32 pigs for catheter electrodes in the right ventricle (RV) and superior vena cava (SVC) using a modified Purdue technique. Both electrode polarities were tested in five different protocols. In part 1, DFTs were determined with 1- to 14-msec monophasic waveforms. In parts 2, 3, and 4, DFTs were determined with two different sizes of SVC electrodes for biphasic waveforms with a phase 1 of 4 or 6 msec and a phase 2 ranging from 1 to 10 msec. In part 5, DFTs were tested for monophasic waveforms ranging from 2 to 11 msec and for biphasic waveforms with a phase 1 duration corresponding to each monophasic waveform and a phase 2 held constant at 1 msec. Mean DFTs for monophasic waveforms were significantly lower when the RV electrode was an anode than when it was a cathode for waveform durations > or = 3 msec. For biphasic waveforms in which phase 2 was < or = phase 1 in duration, no significant difference in mean DFT was observed when polarity was reversed. Even a phase 2 as short as 1 msec could eliminate the DFT difference between polarities observed with monophasic shocks. When phase 2 was > or = 2 msec longer than phase 1, polarity did affect the DFT of biphasic waveforms; it affected the DFT similarly to a monophasic waveform of the same polarity as phase 2. Phase 1 duration and electrode size also affected the difference in DFT produced by changing the electrode polarity. CONCLUSIONS For phase durations most commonly used clinically because of their low DFTs, reversing polarity changed defibrillation efficacy for monophasic but not biphasic shocks. For inefficient biphasic waveforms with phase 2 > or = 2 msec longer than phase 1, the DFT was lower when the RV electrode was an anode during phase 2, similar to the polarity difference for monophasic waveforms, suggesting that a long second phase of biphasic waveforms defibrillates in a similar fashion to monophasic waveforms.

UI MeSH Term Description Entries
D004554 Electric Countershock An electrical current applied to the HEART to terminate a CARDIAC ARRHYTHMIA. Cardiac Electroversion,Cardioversion,Defibrillation, Electric,Electroversion, Cardiac,Electrical Cardioversion,Electroversion Therapy,Therapy, Electroversion,Cardiac Electroversions,Cardioversion, Electrical,Cardioversions,Cardioversions, Electrical,Countershock, Electric,Countershocks, Electric,Defibrillations, Electric,Electric Countershocks,Electric Defibrillation,Electric Defibrillations,Electrical Cardioversions,Electroversion Therapies,Electroversions, Cardiac,Therapies, Electroversion
D004566 Electrodes Electric conductors through which electric currents enter or leave a medium, whether it be an electrolytic solution, solid, molten mass, gas, or vacuum. Anode,Anode Materials,Cathode,Cathode Materials,Anode Material,Anodes,Cathode Material,Cathodes,Electrode,Material, Anode,Material, Cathode
D004597 Electroshock Induction of a stress reaction in experimental subjects by means of an electrical shock; applies to either convulsive or non-convulsive states. Electroconvulsive Shock,Electroconvulsive Shocks,Electroshocks,Shock, Electroconvulsive,Shocks, Electroconvulsive
D004699 Endocardium The innermost layer of the heart, comprised of endothelial cells. Endocardiums
D000818 Animals Unicellular or multicellular, heterotrophic organisms, that have sensation and the power of voluntary movement. Under the older five kingdom paradigm, Animalia was one of the kingdoms. Under the modern three domain model, Animalia represents one of the many groups in the domain EUKARYOTA. Animal,Metazoa,Animalia
D013552 Swine Any of various animals that constitute the family Suidae and comprise stout-bodied, short-legged omnivorous mammals with thick skin, usually covered with coarse bristles, a rather long mobile snout, and small tail. Included are the genera Babyrousa, Phacochoerus (wart hogs), and Sus, the latter containing the domestic pig (see SUS SCROFA). Phacochoerus,Pigs,Suidae,Warthogs,Wart Hogs,Hog, Wart,Hogs, Wart,Wart Hog
D014693 Ventricular Fibrillation A potentially lethal cardiac arrhythmia that is characterized by uncoordinated extremely rapid firing of electrical impulses (400-600/min) in HEART VENTRICLES. Such asynchronous ventricular quivering or fibrillation prevents any effective cardiac output and results in unconsciousness (SYNCOPE). It is one of the major electrocardiographic patterns seen with CARDIAC ARREST. Fibrillation, Ventricular,Fibrillations, Ventricular,Ventricular Fibrillations

Related Publications

J Huang, and B H KenKnight, and G P Walcott, and R G Walker, and W M Smith, and R E Ideker
January 1996, Pacing and clinical electrophysiology : PACE,
J Huang, and B H KenKnight, and G P Walcott, and R G Walker, and W M Smith, and R E Ideker
July 1996, The American journal of cardiology,
J Huang, and B H KenKnight, and G P Walcott, and R G Walker, and W M Smith, and R E Ideker
March 1996, American heart journal,
J Huang, and B H KenKnight, and G P Walcott, and R G Walker, and W M Smith, and R E Ideker
December 1990, American heart journal,
J Huang, and B H KenKnight, and G P Walcott, and R G Walker, and W M Smith, and R E Ideker
April 1998, Journal of cardiovascular electrophysiology,
J Huang, and B H KenKnight, and G P Walcott, and R G Walker, and W M Smith, and R E Ideker
September 1995, Circulation,
J Huang, and B H KenKnight, and G P Walcott, and R G Walker, and W M Smith, and R E Ideker
March 1994, Pacing and clinical electrophysiology : PACE,
J Huang, and B H KenKnight, and G P Walcott, and R G Walker, and W M Smith, and R E Ideker
July 1989, Journal of the American College of Cardiology,
J Huang, and B H KenKnight, and G P Walcott, and R G Walker, and W M Smith, and R E Ideker
November 2001, Resuscitation,
J Huang, and B H KenKnight, and G P Walcott, and R G Walker, and W M Smith, and R E Ideker
July 1994, Journal of the American College of Cardiology,
Copied contents to your clipboard!