[Recording of ventricular pressure by conventional catheter manometer systems. I. Minimal requirements of blood pressure recording systems and estimation of frequency response characteristics]. 1976

G Hellige

Parallel recordings of pressure pulses by conventional catheter manometer systems and catheter tip manometer demonstrate severe errors in the peak velocity of pressure rise estimated by conventional systems. This fact is due to inadequate dynamic response characteristics of conventional systems in relation to the frequency content of pressure curves. During cardiac rest the error in dp/dt max is less than 10% if the frequency response of the recording system is uniform up to 10 Hz, the corresponding value under maximal cardiac stimulation is about 40 Hz. This is equal to the first 10 harmonics of heart rate. The examination of left and right ventricular pressure curves leads to similar results. The experimental determination of dynamic response characteristics of cathermanometer systems requires a test system producing suitable sinus or step functions, parallel high fidelity recording of pressure functions to be recommended. A simple test station is described. Examinations of temperature influence on catheter material and resulting changes in dynamic response characteristics were carried out. The incubation of catheters at the temperature of 37 degrees C is indispensable. A new diagram for simplifying the interpretation of results is described. Other publications are discussed in viewpoint of employed techniques and representation of results.

UI MeSH Term Description Entries
D008433 Mathematics The deductive study of shape, quantity, and dependence. (From McGraw-Hill Dictionary of Scientific and Technical Terms, 6th ed) Mathematic
D001795 Blood Pressure Determination Techniques used for measuring BLOOD PRESSURE. Blood Pressure Determinations,Determination, Blood Pressure
D006328 Cardiac Catheterization Procedures in which placement of CARDIAC CATHETERS is performed for therapeutic or diagnostic procedures. Catheterization, Cardiac,Catheterization, Heart,Heart Catheterization,Cardiac Catheterizations,Catheterizations, Cardiac,Catheterizations, Heart,Heart Catheterizations
D006339 Heart Rate The number of times the HEART VENTRICLES contract per unit of time, usually per minute. Cardiac Rate,Chronotropism, Cardiac,Heart Rate Control,Heartbeat,Pulse Rate,Cardiac Chronotropy,Cardiac Chronotropism,Cardiac Rates,Chronotropy, Cardiac,Control, Heart Rate,Heart Rates,Heartbeats,Pulse Rates,Rate Control, Heart,Rate, Cardiac,Rate, Heart,Rate, Pulse
D006352 Heart Ventricles The lower right and left chambers of the heart. The right ventricle pumps venous BLOOD into the LUNGS and the left ventricle pumps oxygenated blood into the systemic arterial circulation. Cardiac Ventricle,Cardiac Ventricles,Heart Ventricle,Left Ventricle,Right Ventricle,Left Ventricles,Right Ventricles,Ventricle, Cardiac,Ventricle, Heart,Ventricle, Left,Ventricle, Right,Ventricles, Cardiac,Ventricles, Heart,Ventricles, Left,Ventricles, Right
D013696 Temperature The property of objects that determines the direction of heat flow when they are placed in direct thermal contact. The temperature is the energy of microscopic motions (vibrational and translational) of the particles of atoms. Temperatures

Related Publications

G Hellige
May 1951, Techniques hospitalieres, medico-sociales et sanitaires,
G Hellige
October 1994, Journal of applied physiology (Bethesda, Md. : 1985),
G Hellige
September 1977, Research in experimental medicine. Zeitschrift fur die gesamte experimentelle Medizin einschliesslich experimenteller Chirurgie,
G Hellige
August 1991, Kokyu to junkan. Respiration & circulation,
G Hellige
November 1990, Anesthesiology,