Precision and accuracy in a metabolic monitor for indirect calorimetry. 1998

J C Wells, and N J Fuller
Dunn Clinical Nutrition Centre, Cambridge.

OBJECTIVE To determine within-machine and between-machine precision (reproducibility) and accuracy, of the Deltatrac Mk 1 Metabolic Monitor. METHODS Within-machine and between-machine comparison for gas exchange (VO2 and VCO2), respiratory quotient (RQ) and energy expenditure (EE). METHODS 3 Deltatrac Mk 1 Metabolic Monitors. METHODS Within-machine and between-machine reproducibility were assessed by five successive 10 min gas infusion tests in each machine. Accuracy was assessed by measuring independently the gas content of the infusion mixture. The Deltatrac flowmeters were evaluated by further infusion tests. RESULTS Within-study reproducibility was < 1 ml/min for VCO2, < 2.5 ml/min for VO2, < 5 kcal/d for EE and < 0.01 for RQ. Between-study reproducibility was < 0.2% for RQ, < 1% for VCO2 and < 2% for VO2 and EE. Between-machine reproducibility was < 0.1% for RQ and < 1% for VO2, VCO2 and EE. Accuracy in all three machines was within 3% for VO2, VCO2 and EE, and within 0.2% for RQ. CONCLUSIONS The Deltatrac Mk 1 is a very precise metabolic monitor, and is accurate within 3% for gas exchange and EE. RQ is measured with greatest reproducibility and accuracy (within 0.2%), making the monitor particularly suitable for studies of substrate utilisation.

UI MeSH Term Description Entries
D010101 Oxygen Consumption The rate at which oxygen is used by a tissue; microliters of oxygen STPD used per milligram of tissue per hour; the rate at which oxygen enters the blood from alveolar gas, equal in the steady state to the consumption of oxygen by tissue metabolism throughout the body. (Stedman, 25th ed, p346) Consumption, Oxygen,Consumptions, Oxygen,Oxygen Consumptions
D011659 Pulmonary Gas Exchange The exchange of OXYGEN and CARBON DIOXIDE between alveolar air and pulmonary capillary blood that occurs across the BLOOD-AIR BARRIER. Exchange, Pulmonary Gas,Gas Exchange, Pulmonary
D002153 Calorimetry, Indirect Calculation of the energy expenditure in the form of heat production of the whole body or individual organs based on respiratory gas exchange. Calorimetry, Respiration,Calorimetries, Indirect,Calorimetries, Respiration,Indirect Calorimetries,Indirect Calorimetry,Respiration Calorimetries,Respiration Calorimetry
D004734 Energy Metabolism The chemical reactions involved in the production and utilization of various forms of energy in cells. Bioenergetics,Energy Expenditure,Bioenergetic,Energy Expenditures,Energy Metabolisms,Expenditure, Energy,Expenditures, Energy,Metabolism, Energy,Metabolisms, Energy
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D000328 Adult A person having attained full growth or maturity. Adults are of 19 through 44 years of age. For a person between 19 and 24 years of age, YOUNG ADULT is available. Adults
D001481 Basal Metabolism Metabolism (e.g., heat production) of an organism in an inactive, awake, fasting state. It may be determined directly by means of a calorimeter or indirectly by calculating the end products of oxidation within the organism or from the amount of oxygen utilized. Metabolism, Basal,Basal Metabolic Rate,Metabolic Rate, Basal,Resting Metabolic Rate,Basal Metabolic Rates,Metabolic Rate, Resting,Rate, Basal Metabolic,Rate, Resting Metabolic,Resting Metabolic Rates
D012680 Sensitivity and Specificity Binary classification measures to assess test results. Sensitivity or recall rate is the proportion of true positives. Specificity is the probability of correctly determining the absence of a condition. (From Last, Dictionary of Epidemiology, 2d ed) Specificity,Sensitivity,Specificity and Sensitivity
D015203 Reproducibility of Results The statistical reproducibility of measurements (often in a clinical context), including the testing of instrumentation or techniques to obtain reproducible results. The concept includes reproducibility of physiological measurements, which may be used to develop rules to assess probability or prognosis, or response to a stimulus; reproducibility of occurrence of a condition; and reproducibility of experimental results. Reliability and Validity,Reliability of Result,Reproducibility Of Result,Reproducibility of Finding,Validity of Result,Validity of Results,Face Validity,Reliability (Epidemiology),Reliability of Results,Reproducibility of Findings,Test-Retest Reliability,Validity (Epidemiology),Finding Reproducibilities,Finding Reproducibility,Of Result, Reproducibility,Of Results, Reproducibility,Reliabilities, Test-Retest,Reliability, Test-Retest,Result Reliabilities,Result Reliability,Result Validities,Result Validity,Result, Reproducibility Of,Results, Reproducibility Of,Test Retest Reliability,Validity and Reliability,Validity, Face

Related Publications

J C Wells, and N J Fuller
April 2013, The American journal of clinical nutrition,
J C Wells, and N J Fuller
January 1960, Nutritio et dieta; European review of nutrition and dietetics,
J C Wells, and N J Fuller
January 1999, Nutricion hospitalaria,
J C Wells, and N J Fuller
May 2003, AACN clinical issues,
J C Wells, and N J Fuller
January 2006, Physiological and biochemical zoology : PBZ,
J C Wells, and N J Fuller
January 1990, Biomedical instrumentation & technology,
J C Wells, and N J Fuller
January 1993, Journal of applied physiology (Bethesda, Md. : 1985),
J C Wells, and N J Fuller
January 2022, Methods in molecular biology (Clifton, N.J.),
Copied contents to your clipboard!