[Recording of ventricular pressure by conventional catheter manometer systems. Efficiency of several combinations of conventional catheters, modern transducers and catheter-flush systems (author's transl)]. 1976

G Hellige

The experimentally in vitro determined dynamic response characteristics of 38 catheter manometer systems were uniform in the worst case to 5 c.p.s. and optimally to 26 c.p.s. Accordingly, some systems are only satisfactory for ordinary pressure recording in cardiac rest, while better systems record dp/dt correct up to moderate inotropic stimulation of the heart. In the frequency range of uniform response (amplitude error less +/- 5%) the phase distortion is also negligible. In clinical application the investigator is often restricted to special type of cardiac catheter. In this case a low compliant transducer yields superior results. In all examined systems the combination with MSD 10 transducers is best, whereas the combination with P 23 Db transducers leads to minimal results. An inadequate system for recording ventricular pressure pulses leads in most cases to overestimations of dp/dtmax. The use of low frequency pass filters to attenuate higher frequency artefacts is, under clinical conditions, not suitable for extending the range of uniform frequency response. The dynamic response of 14 catheter manometer systems with two types of continuous self flush units was determined. The use of the P 37 flush unit in combination with small internal diameter catheters leads to serious error in ordinary pressure recording, due to amplitude distortion of the lower harmonics. The frequency response characteristics of the combination of an Intraflow flush system and MSD 10 transducer was similar to the non-flushing P 23 Db transducer feature.

UI MeSH Term Description Entries
D008365 Manometry Measurement of the pressure or tension of liquids or gases with a manometer. Tonometry,Manometries
D008433 Mathematics The deductive study of shape, quantity, and dependence. (From McGraw-Hill Dictionary of Scientific and Technical Terms, 6th ed) Mathematic
D001795 Blood Pressure Determination Techniques used for measuring BLOOD PRESSURE. Blood Pressure Determinations,Determination, Blood Pressure
D006328 Cardiac Catheterization Procedures in which placement of CARDIAC CATHETERS is performed for therapeutic or diagnostic procedures. Catheterization, Cardiac,Catheterization, Heart,Heart Catheterization,Cardiac Catheterizations,Catheterizations, Cardiac,Catheterizations, Heart,Heart Catheterizations
D016276 Ventricular Function The hemodynamic and electrophysiological action of the HEART VENTRICLES. Function, Ventricular,Functions, Ventricular,Ventricular Functions

Related Publications

G Hellige
May 1969, The Journal of thoracic and cardiovascular surgery,
G Hellige
August 1991, Clinical physics and physiological measurement : an official journal of the Hospital Physicists' Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics,
G Hellige
January 1982, Acta medica Scandinavica. Supplementum,
G Hellige
August 1981, Nihon Ika Daigaku zasshi,
G Hellige
January 1957, Circulation research,
Copied contents to your clipboard!